Volume/Duration Enabled (Continued)
9.
To see detail screen during or between infusions, press
CHANNEL SELECT
During infusion, Volume Remaining is displayed.
Between infusions:
Number of doses completed and when next dose
starts display.
Yellow Standby Status Indicator illuminates.
NOTES:
If the current time has been previously confirmed, the
Day
screen does not display.
RATE is calculated with each keystroke for DURATION.
See "Delay Options" procedure to program an infusion delay.
When delaying an infusion, a multidose cannot be delayed
for more than 8 hours, and all doses in the multidose program
must be completed within a 24-hour program.
Syringe module: If NEOI is enabled, the Near End of infusion
message appears near the end of the last dose.

Volume/Duration Disabled

1.
To enter rate, use numeric data entry keys.
2.
To enter volume to be infused for each dose, press
soft key and use numeric data entry keys.
VOLUME/DOSE
3.
To enter time interval (1 to 24 hours) between doses, press
DOSE INTERVAL
4.
To enter number of doses, press
use numeric data entry keys.
If Delay Options is enabled, DELAY OPTIONS soft key
appears.
5.
To begin multidose infusion, press
Main Display shows remaining VTBI for that dose.
At completion of a multidose program,
appears on Main Display.
COMPLETE
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Multidose Mode (Continued)
key.
soft key and use numeric data entry keys.
# OF DOSES
START
Time of
soft key and
soft key.
MULTIDOSE
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
General Programming