Prepare Syringe And Administration Set (Syringe Module) - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Prepare Administration Set (Pump Module) (Continued)
5.
If container requires venting, open vent cap on
administration set spike.
6.
To prime tubing and clear air from injection sites and
tubing fitments, slowly open roller clamp.
7.
When priming is complete, close roller clamp.
8.
Verify no fluid flow.

Prepare Syringe and Administration Set (Syringe Module)

To decrease potential start-up delays, delivery inaccuracies,
and delayed generation of occlusion alarms, each time a new
syringe is loaded:
Use smallest syringe size possible (for example, if infusing
2.3 mL of fluid, use a 3 mL syringe).
Use compatible components which have the smallest
internal volume or "deadspace." For example:
◦ Tubing internal diameter: Smallbore or microbore tubing
is recommended when infusing at low rates
◦ Tubing length: Tubing length should be minimized, when
possible
◦ Filters: Internal volume (deadspace) of in-line filters
should be minimized
◦ Connection sites: The number of connection sites such
as stopcocks and y-sites should be limited, and high risk
or life-sustaining solutions should be connected as close
to the vascular access site as possible
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Prime (Continued)
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Getting Started
2-7