Set Bolus Dose (Continued)
2.
Press
BOLUS DOSE
3.
Set key to
PROGRAM
code and press
4.
To enter dose value, use numeric data entry keys.
5.
Press
CONFIRM
If the programmed Bolus Dose is outside the Soft Limit
for that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
If Yes soft key is pressed, programming continues;
if No soft key is pressed, infusion needs to be
reprogrammed.
If the programmed Bolus Dose is outside the Hard
Limit for that care area, an audio alert sounds and a
visual prompt appears before programming can
continue. Infusion must be reprogrammed.
If there is a potential for a programmed "_ _ / _ _ mL"
parameter to result in an excessive volume or dose
being delivered, following prompt appears:
"Cannot proceed due to incorrect concentration or
dosing parameters. Remove syringe, verify
concentration, and reprogram."
The prompt can be (a) the result of an incorrect drug
amount and/or diluent volume entry, or (b) can occur if
hospital-established Guardrails limits are very wide.
Be sure to enter either a drug amount per
drug amount per total volume—for example, a 30 mL
syringe with a concentration of 1 mg/1 mL can be
entered in one of two ways:
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Infusion Modes (Continued)
soft key.
position or enter
soft key.
CONFIRM
soft key.
-digit authorization
4
mL or total
1
A
Summary
Morphine
2
Cont. Dose
mg/h
30
Max Limit:
mg/4 h
[Conc]: 1 mg/mL
>Press START to Close
Summary
BOLUS
PRO-
START
GRAM
DOSE
Programming
3-23