A subsequent primary infusion may be scanned with the Pump
module running if the following criteria match between the
scanned label and the running infusion:
•
Drug or Fluid name (Alias or NDC)
•
Drug Amount
•
Drug Amount Units
•
Diluent Volume
To start a secondary infusion while a primary infusion is in
progress:
1.
To scan barcode on IV container, press
key on Alaris Auto-ID module or scan trigger on handheld
scanner.
The PC unit displays the scanned medication/fluid
information.
2.
Press
CHANNEL SELECT
Primary infusion parameters display.
3.
Press
SECONDARY
4.
Program secondary infusion (see Pump module section of
this User Manual).
Auto-ID Module Section
Alaris System User Manual – with v9.33 Model 8015
Subsequent Primary Infusion
Secondary Infusion
SCAN
key on appropriate module.
soft key.
/
CANCEL
CAUTIONS
•
CLASS 1 LED PRODUCT
stare into the beam or allow beam
to strike patient's face.
• Always verify that information
displayed on the PC unit matches
scanned data.
: Do not
Programming
6-9