CareFusion Alaris 8015 User Manual page 116

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Manual Programming—Primary Infusion (Continued)
Fluid Infusion (Continued)
Volume/Duration Infusion (Continued)
NOTES:
The infusion can be paused by pressing the PAUSE soft key.
See "Pause, Change, Restart Infusion," "Pause and Restart
Infusion" procedure.
Time Left
To view infusion
press CHANNEL SELECT key. To return to previous screen,
press START soft key.
NOTE:
It is possible to program an infusion with a rate that is
displayed with two decimal places (one-hundredth of a mL per
hour) on the PC unit for the Pump module. However, due to
space limitations on the Pump module rate display, the rate
is displayed to the nearest one-tenth of a mL per hour on the
Pump module. This value is only used for display purposes
and the Pump module is actually infusing at the more precise
rate noted on the PC unit.
Manual Programming with Guardrails Suite MX
2-50
during a volume/duration infusion,
Time Left:
>Press START
PAUSE
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Guardrails Fluid Setup
TPN
4.17
mL/h
RATE
ALL
VTBI
00 45
h
min
VOLUME
START
DURATION