Prepare Syringe and Administration Set (Syringe Module) (Continued)
10. Insert pressure sensing disc (if used), as follows:
a. Orient pressure sensing disc, as follows:
fluid side up (patient side down)
cavity forward (membrane toward instrument)
b. Gently slide pressure sensing disc up into slot in
pressure sensing disc housing.
c. Apply firm upward pressure on pressure sensing disc
(not tubing) until disc snaps into place.
Getting Started
2-14
Load (Continued)
WARNING
When the pressure sensing disc
is not being used and an occlusion
occurs, there is a risk of infusing
pressurized buildup of infusates
upon correction of the occlusion. To
avoid an inadvertent bolus, relieve
the pressure before restarting the
infusion.
Slot in Pressure
Sensing Disc
Housing
Pressure
Sensing Disc
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Syringe Module
Fluid Side
Patient Side
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Sensing Disc