Definitions (continued)
Terms
Maintenance Reminder
Medium Priority Alarm
NEOI Snooze (Reminder signal)
Monitoring Alarm
Presilence
Prompt
Technical Alarm
All Alaris System Alerts are prioritized by type/source and the
condition of the alert.
ALERT Priority
HIGH Priority
MEDIUM Priority
LOW Priority
INFORMATIONAL
Reminder Signal
Troubleshooting and Maintenance
A-8
Alarms and Alerts (continued)
A visual message that when enabled appears at startup when a scheduled preventive
maintenance is due/overdue for component of Alaris System (PC unit or attached module).
A medium-priority alarm is initiated by any alarm condition requiring a prompt user response.
Configurable alert, if enabled, provides an audio only tone when a Near End of Infusion alarm
had been previously silenced.
An alarm arising from a monitoring module (SpO2 or EtCO2).
Allows the user to silence a monitoring module for 2 minutes. Also known as Audio Pause, the
duration of Presilence (Audio Pause) is not configurable..
An audio signal and/or a visual message appearing on bottom line of Main Display or in
Message Display. Audio signal can be silenced for 12 seconds pressing the SILENCE key.
An alarm arising from a hardware or software failure condition.
Alert Prioritization
ALERT Type/Source
Technical Software Fatal
Uncontrolled Infusion
Technical Malfunction
Monitoring
Infusion/PC unit
Monitoring
Infusion/PC unit
Infusion/PC unit
Informational
Reminder Signal
Definitions
Color Theme
Silenceable
N/A
NO
Flashing Red
NO
Flashing Red
YES
Flashing Red
YES
Flashing Red
YES
Flashing Yellow
YES
Flashing Yellow
YES
Steady Yellow
YES
N/A
N/A
N/A
NO
Alaris System User Manual – with v9.33 Model 8015