Manual Programming—NO Guardrails Suite MX Protection (Continued)
Infusion (Continued)
2.
Enter secondary infusion rate or duration, as follows:
To enter secondary infusion rate, press
and use numeric data entry keys.
To enter duration, press
numeric data entry keys.
3.
To enter secondary volume to be infused, press
key and use numeric data entry keys.
4.
Open clamp on secondary administration set.
5.
Verify correct infusion parameters and press
key.
6.
Verify that drops are flowing from the secondary container
drip chamber.
Change Primary Infusion Parameter
1.
Press
CHANNEL SELECT
2.
Press
PRIMARY
3.
To change primary infusion parameter, press applicable
soft key (
or
RATE
4.
Verify correct primary infusion parameters and press
soft key.
SECONDARY
Secondary setup screen is displayed.
5.
To resume secondary infusion, press
6.
Verify that drops are flowing from the secondary container
drip chamber.

Stop Secondary and Return to Primary

See Manual Programming—Secondary Infusion (Pump
Module) on page 2-51.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
DURATION
key.
soft key.
) and use numeric data entry keys.
VTBI
soft key
RATE
soft key and use
soft
VTBI
soft
START
soft key.
START
Programming—NO Guardrails Suite MX Protection
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
2-95