Wireless Connection Soft Key
Send the PC unit to Biomed to resolve wireless connectivity
issues.
The Wireless Connection soft key is inactive (grayed out) for
one of the following reasons:
The System Maintenance software was used to disable the

wireless connection.

The CF card flashing process was done without the
programming of the proper AppConfig file (v9.12 or later).
A valid network configuration has not been transferred.
To enable the wireless connection, consult the v9 or later
System Maintenance software.
Wireless Connection Scenarios
At various locations, the PC unit may or may not be able to
complete a wireless connection. When connected, the
connection icon lights. If the PC unit cannot connect, the
connection icon does not light, and device data is stored until a
connection can be completed.
NOTE:
If there is no wireless connection, PC unit data (such as log
information) is stored until a connection is re-established. System
history and CQI logs are stored in non-volatile compact flash
memory and are retained until memory is full and older items are
purged.
The following table explains how the pump functions under
various wireless scenarios.
Full Connectivity
Connectivity
icon lighted
Near real-time
full data transfer
A-26
Wireless Connection
No Connectivity
Connectivity
icon not lighted
Data stored
until connection
is reestablished
Marginal Connectivity
NA
Data stored and
transferred as PC
unit reestablishes
connection
Alaris System User Manual – with v9.33 Model 8015