Manual Programming—Primary Infusion (Continued)
4.
Press
CHANNEL SELECT
5.
Syringe module: Select syringe type and size, as
follows; otherwise, proceed to step
syringe is loaded correctly, but not recognized, check
FU
for the following:
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
a. If a label is between the syringe barrel and the barrel
clamp, make sure that the label does not erroneously
enlarge the barrel size of the syringe.
b. If a needle-free valve or other component is added to
the syringe, ensure that it is no larger than the diameter
of the syringe barrel.
NOTE:
Thick labeling or adding a component to the syringe that is larger
than the diameter of the syringe may prevent the device from
correctly recognizing the installed syringe. If the issue continues
despite the above troubleshooting, send the device to your
facility's biomedical engineering department for servicing.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
key.
.
If the installed
6
Pump Module
WARNING
Ensure that the displayed syringe
manufacturer and syringe size
correctly identify the installed
syringe. Mismatches might
cause an under-infusion or
over-infusion to the patient that
could result in serious injury and/or
death. For a list of compatible
syringes, refer to Compatible
Syringes (Syringe Module) on page
2-136 and the Alaris
Module Set Compatibility Card.
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Syringe Module
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
TM
Syringe
2-25