Programming with Interoperability—Initial Primary Infusion (Continued)
Continuous Infusion (Continued)
5.
For the Syringe module, unclamp tubing and attach
administration set to patient.
NOTE:
Unclamping tubing and starting infusion before attaching
administration set to patient minimizes any potential bolus that can
be released from pressure built up in set due to normal syringe
loading and priming.
Programming with Interoperability and Guardrails Suite MX
2-68
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015