1.
Press
OPTIONS
2.
Press Prime Set with Syringe soft key.
3.
Press and hold
of syringe administration set is complete.
Volume used during priming is displayed but is not
added to
VTBI
Fluid is delivered during priming only while
key is pressed. Each press of
up to 2 mL at the maximum rate of the selected
syringe. (see Rate Restriction by Syringe Size: on
page 3-59) of priming/fluid per continuous press. To
deliver additional amounts, press
again.
4.
When priming is complete, release
5.
To return to main screen, press
Guardrails Drug Setup screen is displayed.
6.
Select infusion mode
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Prepare Infusion (Continued)
Prime (Continued)
key.
soft key until fluid flows and priming
PRIME
.
PRIME
PRIME
EXIT
.
soft
PRIME
soft key delivers
soft key
PRIME
soft key.