Alarm Message
ATTACH HANDSET
ACCUMULATED
AIR-IN-LINE ALARM
AIR-IN-LINE
ANESTHESIA MODE
AC POWER
CONNECTED
ANESTHESIA MODE
AC POWER
DISCONNECTED
ATTACH
MONITORING
MODULE
A-12
Alarms and Alerts (continued)
Alarms
Device
Priority
PCA
Low
Pump
High
Pump
High
PC unit
Informational
PC unit
Informational
PC unit
Informational
Alarms
Meaning
The Dose Request Cord
detached from the device during
an active infusion. The Dose
Request Cord is required only
for PCA and PCA + Continuous
Infusion modes.
A large number of air bubbles
smaller than the current
air-in-line limit has recently
passed the air-in-line detector.
Air has been detected in the
administration set during an
infusion. The infusion stops on
affected channel.
The AC power cord is connected
The AC power cord is
disconnected
A Clinical Advisory is displayed
for a PCA Infusion that contains
PCA PAUSE PROTOCOL.
Alaris System User Manual – with v9.33 Model 8015
Response
Reattach the Dose Request Cord and
press the RESTART key.
Clear air from the administration set.
Press the RESTART key, or press the
CHANNEL SELECT key and then the
START soft key.
Ensure that the tubing is properly
installed in the Air-in-Line detector. If
air is present, clear the air from
administration set. Press the
RESTART key, or press the
CHANNEL SELECT key and then the
START soft key.
Select Yes or No to continue in the
anesthesia mode
The anesthesia mode was
discontinued when the AC power cord
was disconnected.
Attach the EtCO
or SpO
module.
2
2
Appendix Section