CareFusion Alaris 8015 User Manual page 146

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Programming with Interoperability—Subsequent Primary Infusion
Hospital policy should determine the preferred method for
programming subsequent infusions.
Subsequent infusion for the Syringe module refers to new
syringes used with the Syringe module.
An automated programming request for a subsequent infusion
can be sent to a Syringe module when it is infusing or in an idle
state.
NOTE:
If an infusion is in Delay when a subsequent automated
programming request is sent, the Delay will be canceled and the
subsequent automated programming request infusion parameters
will be pre-populated on the PC unit. To add a delay manually, see
Delay Options on page 2-102
Pre-population of infusion parameters for a subsequent syringe
is the same workflow as an Initial Primary Infusion., see
Programming with Interoperability—Initial Primary Infusion on
page 2-62.
Programming with Interoperability and Guardrails Suite MX
2-80
(Syringe Module)
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015