Set Up PCA Dose and Continuous Infusion (Continued)
10. Verify correct parameters on second nurse summary
screen and press
During PCA dose + Continuous Infusion:
•
Green Infusing Status Indicator illuminates.
•
DELIVERING PCA
when initiated. Continuous and PCA drug name scrolls
in Channel Message Display between PCA doses.
•
Volume infused for continuous dose is displayed in
Rate Display.
•
Main Display alternates between volume remaining
and infusion mode with PCA drug name.
•
When PCA dose is complete,
in Channel Message Display and resumes continuous
dose.
Set Loading Dose Only
The following procedures should be used when setting a
LOADING DOSE ONLY
1.
Perform steps in "Prepare Infusion."
2.
Press
LOADING DOSE ONLY
screen.
3.
To enter dose value, use numeric data entry keys.
Programming
3-20
Infusion Modes (Continued)
soft key.
START
scrolls in Channel Message Display
PCA COMPLETE
using the Drug Library.
soft key from Infusion Mode
scrolls
Alaris System User Manual – with v9.33 Model 8015
A
Second Nurse Summary
Morphine
1
PCA Dose:
mg
Lockout
10
minutes
Interval:
2
Cont. Dose:
mg/h
8
Max Limit:
mg/4 h
[Conc]: 1 mg/mL
>Press START
PRO-
START
PAUSE
GRAM
PCA Module Section