Prime - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Select Syringe Type and Size (Continued)
2.
Press soft key next to installed syringe type and size. If
installed syringe is not listed, press
key and select syringe from list.
Selection is highlighted.
soft key is activated.
CONFIRM
3.
To accept, press
Drug Library screen is displayed.
The Priming option can be enabled at the time the
Alaris System is configured for use. The Priming selection
(
soft key) is available only after the syringe type and

PRIME

medication selection (prior to infusion mode selection).
Programming
3-8
Prepare Infusion (Continued)
ALL SYRINGES
soft key.
CONFIRM
Prime
soft
WARNING
When priming:
• Ensure that patient is not
connected.
• Ensure that air is expelled from
line prior to beginning infusion
(unexpelled air in line could have
serious consequences).
Failure to prime correctly can delay
infusion delivery and cause the total
volume to be infused to read higher
than the actual total delivered to the
patient.
CAUTION
During priming, the pressure limit
alarms are temporarily increased to
their maximum level.
Alaris System User Manual – with v9.33 Model 8015
A
Syringe Selection
IMS Pre-fill 30 mL
>Select Syringe Size
ALL
CONFIRM
SYRINGES
PCA Module Section

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents