PCA syringe sizes and the correlating 35% of the syringe
Syringe Size
60mL
35mL
30mL
25mL
20mL

Stop a Loading, PCA, or Bolus Dose

1.
Press
CHANNEL SELECT
2.
Press
STOP LOAD, STOP PCA
applicable.
Available soft key and stop confirmation screen are
dependent on type of dose currently infusing and current
infusion mode.
3.
To stop dose and resume current program, press Yes soft
key.
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Infusion Modes (Continued)
capacity
35% of Syringe Capacity
21mL
12.25mL
10.5mL
8.75mL
7mL
key.
, or
STOP BOLUS
soft key as
A
Summary
Morphine
1
Loading
mg
Dose:
[Conc]: 1 mg/mL
>Press START to Close
Summary
STOP
START
LOAD
Programming