Waveform Analysis

The EtCO
module provides the option to display EtCO
readings as a waveform. The following graph
2
2
is an example of a normal waveform (normal ventilation, 35 - 45 mmHg). In the event the EtCO
value
2
is above the waveform display range, the top of the waveform will be clipped. Numerical EtCO
values
2
continue to be displayed on both the EtCO
module and PC unit.
2
A - B: baseline period of no CO
; end of inhalation
2
B - C: rapid rise in CO
2
D D D D
C - D: alveolar plateau
D:
end of expiration; end tidal CO
(EtCO
)
2
2
D - E: inhalation
EtCO
Module Section
General Information
2
Alaris System User Manual – with v9.33 Model 8015
5-27