Syringe Module (Continued)
Bolus Volume, Maximum after
Occlusion: (Continued)
Critical Volume:
Dimensions:
Environmental Conditions:
Equipment Orientation:
General Information
2-156
Specifications (Continued)
Maximum Bolus Volume specifications are based on following standard
operating conditions:
Atmospheric Pressure:
Disposable Type:
No Pressure Disc:
With Pressure Disc:
Humidity:
Rate:
Syringe Type:
Temperature:
Volume Collection Time:
Maximum over-infusion which can occur in the event of a single-fault
condition will not exceed 2% of nominal syringe fill volume during loading
and 1% of maximum syringe travel after syringe loading.
x
4.5" W
15.0" H
Symbol
Atmospheric
Pressure
Relative Humidity
(Avoid prolonged
exposure to
relative humidity
>85%)
Temperature
Range
To ensure proper operation, Alaris System must remain in an upright position.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
645 - 795 mmHg
#30914
#30920
20 - 90%
5 mL/h
BD 50/60 mL
68 ±4°F
approximately 2 minutes
x
7.5" D
Meaning
525 - 4560 mmHg
(700 - 6080 hPa)
Noncondensing
Alaris System User Manual – with v9.33 Model 8015
Operating
Storage/Transport
375 - 760 mmHg
(500 - 1013 hPa)
20 - 90%
Noncondensing
41 - 104°F
(5 - 40°C)
(-20 - 60°C)
Pump and Syringe Modules Section
5 - 85%
-4 - 140°F

Hide quick links: