Syringe Module (Continued)
Flow Rate Programming:
Fluid Ingress Protection:
Infusion Pressure, Maximum:
KVO (Keep Vein Open) Rate:
KVO Rate Range:
Occlusion Alarm Thresholds:
Operating Principle:
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Specifications (Continued)
Flow rate range is from 0.01 to 999 mL/h and can be selected as follows:
Flow Rates (mL)
0.01 - 9.99
10 - 99.9
100 - 999
Rate Restriction by Syringe Size:
Syringe Size (mL)
50/60
30
20
10
5
3
1
IPX
, Drip Proof
1
Without Pressure Sensing Disc: approximately 800 mmHg (actual
occlusion pressure varies based on syringe size and manufacturer)
With Pressure Sensing Disc: 1060 mmHg
Factory default setting is 1 mL/h if set rate is 1 mL/h or above; or set rate, if
rate is 0.9 mL/h or below.
KVO rate can be set in System Configuration, in 0.01 mL/h increments, as
follows:
0.01 - 2.5 mL/h (0.01 - 0.09 mL/h available for 1 mL and 3 mL syringes)
Without Pressure Sensing Disc: Three settings—Low, Medium, High
With Pressure Sensing Disc: User selected, 25 - 1000 mmHg in 1 mmHg
increments.
Positive displacement
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Selectable Increments (mL/h)
0.01
0.1
1
Flow Rate Range (mL/h)
0.1 - 999
0.1 - 650
0.1 - 500
0.1 - 250
0.1 - 150
0.01 - 100
0.01 - 30
General Information
2-157