Guardrails PCA Pause Protocol Feature (Continued)
Program an Infusion (Continued)
NOTES:
To review PCA pause limits, see "Review or Change PCA
Pause Alarm Limits."
Once the
START
alternates between volume remaining (VTBI - Volume to be
Infused) and PCA drug name with the infusion code.
The Main Display displays PCA Pause Protocol ON.
If Patient ID is entered, Patient ID alternates with PCA
Pause Protocol ON.

Review or Change PCA Pause Alarm Limits

1.
From Main Display press
2.
Press
OPTIONS
3.
Press PCA Pause Limits soft key.
4.
Verify that PCA pause limits as per facility protocol or
physician order.
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
soft key is pressed, the Main Display screen
CHANNEL SELECT
key.
.
Midtown Hospital
Med Surg
A
PCA + Cont Morphine
B
VTBI = ALL (30 mL)
C
ETCO2
D
SPO2
PCA Pause Protocol ON
AUDIO
VOLUME
INFUSED
ADJUST
A
Channel Options
1 of 2
PCA Pause Limits
Infusion
Modes
Patient History
Dose
Request Setup
Drug Event
History
>Select an Option or
EXIT
PAGE
EXIT
DOWN
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
97
%SPO2
88
LOW
5
RR
4
LOW
>Press CONFIRM to Apply
Changes
DISABLE
DISABLE
CONFIRM
SPO2
ETCO2
Programming