Features and Definitions (Continued)
Feature
Drug Library
Duration Limits
Dynamic Pressure
Display
Event Logging
Initial Value
IV Fluid Library
Limit
Multidose Mode
Rapid Bolus
Restore
General Information
2-138
Features and Displays (Continued)
Definition—Pump and Syringe Modules
When Profiles feature is enabled, it provides a hospital-defined list of drugs and
concentrations appropriate for use in as many as 30 Profiles. Drug Library use
automates programming steps, including drug name, drug amount and diluent
volume, and activates hospital-established best-practice limits. Drug Library
entries can be delivered as a primary or secondary, or both, as determined by
hospital-health system.
Hospital-established limits around duration of an Intermittent Infusion.
Appears on Main Display. If enabled, it graphically displays current patient-side
occlusion pressure set point and current patient-side operating pressure for that
module. (See "Displays" for additional "Dynamic Pressure Display" information.)
Event Logging records instrument operations.
An optional and editable starting value for Continuous Infusion dose, duration,
Bolus Dose, bolus rate of administration or Bolus Dose duration.
An optional library consisting of IV Fluids (for example, TPN) and limits around
rate of delivery.
A programming Limit or best-practice guideline determined by hospital/health
system and entered into system's Data Set. Supports concentration Limits for
all infusions that utilize concentration. Profile-specific Limits can be defined
for flow rate, patient weight, body surface area (
continuous dose, or total dose and duration for each drug in a Drug Library.
Dose and duration Limits can be defined by hospital/health system as a Hard
and/or Soft Limits.
• A Hard Limit is a programmed Limit that cannot be overridden, except in
anesthesia mode.
• A Soft Limit is a programmed Limit that can be overridden.
Allows 2 - 24 doses to be programmed at equally spaced intervals on the
same module over a 24-hour period. This mode is designed to allow delivery
of multiple, equal doses from the same IV container at regularly scheduled
intervals.
Fastest rate at which Bolus Dose should be delivered, as defined by facility's
clinical best-practice guidelines.
To simplify programming, can be used to recall previous rate and volume
settings for same patient. This option is only available if patient is not new and
system is powered up within 8 hours of last usage.
), maximum and minimum
BSA
Pump and Syringe Modules Section