Displays; Configurable Settings - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

The displays illustrated throughout this document are for
illustration purposes only. The display content varies,
depending on configuration settings, type of administration
set in use, hospital-defined Data Set uploaded using the
Guardrails Suite MX, programmed drug calculation
parameters, and many other variables. Refer to specific
drug product labeling for information concerning appropriate
administration techniques and dosages.
See the PC unit section of this User Manual for system

configurable settings.

The configuration settings are selected during Data Set
development and then uploaded to the Alaris System as part of
the data.
With the Profiles feature enabled, the settings are configured
independently for each Profile. A hospital-defined, best-
practice Data Set must be uploaded to enable the Profiles
feature. Date and Time is a system setting and is the same in
all Profiles.
Feature
Authorization Code
Bolus Delivery Rate
Bolus Dose
Bolus Dose include in Max. Limit
Dose Request Cord Configuration
Forced Module Location
Loading Dose
Lockout Interval
General Information
3-56
Features and Displays (Continued)

Displays

Configurable Settings
Default Setting
1 - 99 minutes in 1-minute
None
150 mL/h
Disabled
Disabled
Profile 2
Enabled
Disabled
increments
Alaris System User Manual – with v9.33 Model 8015
Options
4 digits (0 - 9)
One code applies to all Profiles
75 - 500 mL/h (limited by syringe size)
Enabled - Disabled
Enabled - Disabled
Profile 1, 2, 3
Enabled - Disabled
Enabled- Disabled
Min/Max 1 - 99 minutes
PCA Module Section