Features and Definitions (Continued)
Feature
Guardrails Suite MX
Patient ID
Profile
System Configuration
Tamper Resist
General Information
1-38
Features and Displays (Continued)
Designed to help prevent programming errors by:
• Customizing device configurable settings to meet need of selected
hospital/facility area/unit (Profile).
• Comparing user-programming with hospital-defined best-practice
guidelines.
• Providing a visual and audio prompt if an out-of-limits entry is made.
An optional alphanumeric
entered and displayed
• When enabled, ID entry defaults to Startup screen.
• When disabled, ID entry is only accessible from System Options
screen.
A unique set of system configuration settings and best-practice guidelines
for a specific patient population or patient type, and can consist of
following components:
• Instrument configuration settings.
• A Drug Library, which includes drug names, standard concentrations,
dosing units, duration limits, and optional associated Clinical Advisories
for both continuous and Bolus Dose infusion.
• An IV Fluid library, an optional library consisting of IV Fluids (for
example, TPN) and limits around rate of delivery.
• A Channel Label Library with text (alphanumeric) labels, which allows
identification (on modules) that can be used to indicate route of delivery
(for example, epidural).
Profile settings are established by the facility's own multi-disciplinary team
prior to system implementation. Profile parameters are used to create a
Data Set, which is then transferred to the PC unit.
Allows system settings to be customized. If Profiles feature is enabled, the
system settings defined for selected Profile are automatically activated.
Provides a quick one-touch lockout of the front panel keypad when the
infusion is running, during a delay, or a SpO
monitoring. You cannot lockout the front panel keypad during KVO. An
alarm can be silenced even though the panel is locked.
Definition
-character patient identifier that can be
16
or EtCO
2
Alaris System User Manual – with v9.33 Model 8015
module is actively
2
PC Unit Section