View Patient History (Continued)
7.
To return to Main Display, press
NOTES:
Total drug delivered includes applicable loading dose, PCA
dose, continuous dose, and Bolus Dose. Total drug delivered
does not include priming volume.
Patient History stores a rolling 24-hour log and is
automatically cleared when selecting:
Yes to New Patient? during startup.
A different drug from the Drug Library.
The same drug with different dosing units from the Drug
Library.
A new Therapy for an mL-based drug.
Same patient with a new Profile.
The Patient History defaults to the 8 hour view.
1.
Press
CHANNEL SELECT
2.
From Main Display, press
3.
Press Patient History soft key.
4.
Press
CLEAR HISTORY
A confirmation screen appears.
5.
To continue and clear patient history, press Yes soft key.
To cancel and return to patient history, press No soft key.
Programming
3-28
Infusion Modes (Continued)
EXIT

Clear Patient History

key.
OPTIONS
soft key.
soft key.
key.
Alaris System User Manual – with v9.33 Model 8015
A
Patient History 09:01
History from 07:01 - 09:01
LAST
07:00
2003-6-10
CLEARED
SHIFT TOTALS
Total Drug:
22 mg
Total Demands:
5
Delivered:
2
2
ZOOM: 24
12
8
4
1 hours
>Press ZOOM to Change
Time Scale
CLEAR
ZOOM
EXIT
DETAIL
HISTORY
PCA Module Section