CareFusion Alaris 8015 User Manual page 95

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Manual Programming—Primary Infusion (Continued)
Continuous Infusion (Continued)
If
was selected and facility has defined a Clinical
YES
Advisory for that drug, a message appears. To indicate
information has been noted and continue
programming, press
If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
If selected drug is not weight-based, Not Used is
displayed in
PATIENT WEIGHT
If hospital/facility practice guidelines identify selected
drug as weight-based, prompt for a patient weight in
kilograms appears (as in illustrated example, which
reflects use of Alteplase).
4.
Verify correct parameters and press
confirm.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key.
CONFIRM
field.
soft key to
NEXT
A
Alteplase
Stroke
Clinical Advisory:
This dosing is for Acute
Ischemic STROKE
>Press CONFIRM
A
Alteplase
Stroke
DRUG
AMOUNT
DILUENT
VOLUME
PATIENT
WEIGHT
TIME
UNITS
DOSING
UNITS
[Conc]: 1 mg/mL
>Press NEXT to Confirm
A
Alteplase
Stroke
DRUG
AMOUNT
DILUENT
VOLUME
PATIENT
WEIGHT
TIME
UNITS
DOSING
UNITS
[Conc]: 1 mg/mL
>Enter Patient Weight
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
CONFIRM
100
mg
100
mL
Used
hour
mg/kg/h
DRUG
NEXT
LIBRARY
100
mg
100
mL
_ _ _ _
kg
hour
mg/kg/h
DRUG
LIBRARY
2-29

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents

Save PDF