Programming with Interoperability—Initial Primary Infusion (Continued)
The generic workflow for pre-population of infusion parameters
is as follows. Contact your EMR/HIS vendor for detailed
instructions. Using the EMR/HIS system scanner used for
barcode medication administration:
•
Scan Patient's ID Band
•
Scan fluid/medication barcode label
•
Scan barcode label on the Pump or Syringe module
◦
Pump module: The Pump module must be in an idle
state or the clinician has pressed
Pre-population will fail if the module is scanned after the
user has manually programmed the infusion beyond
pressing
CHANNEL SELECT
◦
Syringe module: Pre-population of infusion parameters
for a Syringe module may occur when the module is
already infusing, in an idle state, and/or the clinician has
pressed
CHANNEL SELECT
the module is scanned after the user has manually
programmed the infusion after pressing
.
SELECT
NOTE:
If the automated programming request is sent while the Syringe
module is infusing, a prompt appears on the PC unit to install the
scanned Syringe.