To ensure that the Alaris System remains in good operating
condition, both regular and preventive maintenance inspections
are required. Refer to the System Maintenance software for
detailed instructions.
REGULAR INSPECTIONS
PROCEDURE
INSPECT FOR DAMAGE
Exterior Surfaces
IUI Connector*
Keypad
Pole Clamp
Power Cord
CLEANING
START-UP
NOTE:
* While visually inspecting the IUI connectors, look for fractures
on the connectors black-colored plastic. If you see any damage,
do not use an instrument with fractured IUI connectors. The IUI
connector must be replaced before the instrument can be used
again.
Inspect IUI Connectors
If any surface contaminants or blue or green deposits are
visible, the connector must be replaced.
Figure 1: Male IUI Connector
(right side)
1.
Visually inspect the right side (male) IUI connector for
cracks on the entire surface of the black colored plastic
housing. See Figures 1 and 2 for the male IUI inspection
area and typical cracks.
2.
Visually inspect the left side (female) IUI connector for
cracks on the edges of the black colored plastic housing.
3.
If cracks are found, replace the IUI connector before use.
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Inspection Requirements
FREQUENCY
Each usage
Each usage
Each usage
Each usage
Each usage
As required
Each usage
Figure 2: Male IUI Connector
WARNING
Failure to perform these
inspections can result in improper
instrument operation.
CAUTION
Preventive maintenance
inspections should only be
performed by qualified service
personnel.
WARNING
Failure to follow these instructions
may result in potential hazards
associated with damaged IUI
connectors.
WARNING
Inserting a finger or other object into
the IUI connector, when the module
is attached to the PC unit, could
result in electrical shock.
A-27