Prepare Administration Set (Pump Module)

For instructions on how to go from checking in a Pump module
to preparing it for an infusion setup, see "General Setup and
Operation."
1.
Ensure that the instrument is as close
to the level of patient's heart as
possible - the patient's heart should
be in line with
CHANNEL SELECT
key.
Align with Patient's heart level
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015

Load

WARNING
• Administration Sets:
Use only Alaris Pump
module/GEMINI
System administration sets.
The use of any other set can
cause improper instrument
FU
operation, resulting in an
inaccurate fluid delivery or other
RATE (mL/h)
potential hazard. For a list of
compatible administration sets,
please refer to the Alaris
CHANNEL
SELECT
Module Set Compatibility Card.
PAUSE
Discard if packaging is not intact or
CHANNEL
OFF
RESTART
protector caps are unattached.
CAUTION
• Ensure that the device is as close
to level of patient's heart as
possible. Patient's heart level
should be in line with the
CHANNEL SELECT key.
• Use compatible components
which have the smallest internal
volume or "deadspace" to
minimize residual volumes
between the syringe and the
patient when administering
medications or fluids at low
infusion rates (e.g.,< 5 mL/h, and
especially flow rates < 0.5 mL/h).
This reduces the amount of time it
takes for fluid to reach the patient,
maintains delivery accuracy, and
reduces occlusion detection
times.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Infusion
TM
TM
Pump
Getting Started
2-3