Audio Characteristics
Profile
Priority
High
Profile 1
Medium
Low
High
Profile 2
Medium
Low
High
Profile 3
Medium
Low
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Alarms and Alerts (continued)
Audio Characteristics
Repeating sequence of 10 beeps followed by an approximate 7 second pause for a monitoring
alarm
Repeating sequence of 2 beeps followed by an approximate 1.5 second pause for an infusion
alarm
Repeating sequence of 6 beeps followed by an approximate 4 second pause for a monitoring
alarm
Repeating sequence of 1 beep followed by an approximate 2 second pause for an infusion
alarm
Repeating sequence of 1 beep followed by an approximate 3.5 second pause.
Repeating sequence of 10 beeps followed by an approximate 7 second pause for a monitoring
alarm
Repeating sequence of 2 beeps followed by an approximate 1 second pause for an infusion
alarm
Repeating sequence of 6 beeps followed by an approximate 4 second pause for a monitoring
alarm
Repeating sequence of 1 beep followed by an approximate 2 second pause for an infusion
alarm
Repeating sequence of 1 beep followed by an approximate 3.5 second pause.
Repeating sequence of 10 beeps followed by an approximate 7 second pause for a monitoring
alarm
Repeating sequence of 2 beeps followed by an approximate 0.5 second pause for an infusion
alarm
Repeating sequence of 6 beeps followed by an approximate 4 second pause for a monitoring
alarm
Repeating sequence of 1 beep followed by an approximate 2 second pause for an infusion
alarm
Repeating sequence of 1 beep followed by an approximate 3.5 second pause.
Troubleshooting and Maintenance
Visual
Indicator
Flashing Red
Flashing
Yellow
Steady
Yellow
Flashing Red
Flashing
Yellow
Steady
Yellow
Flashing Red
Flashing
Yellow
Steady
Yellow
A-9