Manual Programming—Secondary Infusion (Pump Module) (Continued)
6.
VTBI entry:
When
is less than 10 mL/h, entry can be to
VTBI
two decimal places (one-hundredth of a mL).
VTBI is pre-populated with diluent volume of infusion.
To change VTBI, press
data entry keys.
7.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
numeric data entry keys (rate value is calculated and
displayed).
To enter rate, press
numeric data entry keys.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Infusion (Continued)
soft key and use numeric
VTBI
DURATION
RATE VOLUME
soft key and use
soft key and use
Manual Programming with Guardrails Suite MX
WARNING
The secondary
VTBI
require consideration of such
variables as factory
overfill, medication additions.
Underestimating the volume
causes the remaining secondary
solution to be infused at the primary
rate; overestimating results in the
primary solution being infused at the
secondary rate. Multiple doses from
a single container are not possible.
A
Guardrails Drug Setup
Methotrexate
SECONDARY
_ _ _
RATE =
50
VTBI
mL
_ _ _ _
DURATION
1200
DOSE =
mg/m
[Conc]: 136.8 mg/mL
>Enter Duration
RATE
SETUP
VOLUME
A
Guardrails Drug Setup
Methotrexate
SECONDARY
4
RATE =
mL/h
50
VTBI
mL
6:00
DURATION
1200
DOSE =
mg/m
[Conc]: 136.8 mg/mL
>Verify Secondary Clamp
Open, Then Press START
RATE
PAUSE
SETUP
VOLUME
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
settings
mL/h
h
min
2
hh:mm
2
START
2-55