Set Up Pca Dose Only - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Set Up PCA Dose Only

1.
Perform steps in "Prepare Infusion."
2.
Press PCA Dose Only soft key from Infusion Mode
screen.
3.
To enter PCA dose, use numeric data entry keys.
4.
To enter lockout interval, press
key and use numeric data entry keys.
5.
To enter maximum limit, press
then Yes soft key.
If No is selected - then no MAX LIMITS will be available for
this infusion.
6.
Enter maximum limit using numeric data entry keys.
Time (in hours) associated with Max Limit is automatically
entered based on setup in system configuration.
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Infusion Modes (Continued)
LOCKOUT INTERVAL
soft key and
MAX LIMIT
A
INFUSION MODES
PCA Dose only
Continuous Infusion
PCA
Loading Dose Only
>Select an Option
A
LOCKOUT
INTERVAL
LOADING
[Conc]: 1 mg/mL
>Select LOCKOUT
INTERVAL
CHANGE
soft
A
LOCKOUT
INTERVAL
LOADING
[Conc]: 1 mg/mL
>Select Max Limit Option
CHANGE
Guardrails Drug Setup
Morphine
Dose + Continuous
DRUG
SETUP
LIBRARY
PCA Only
Morphine
PCA
_ 20
mg
DOSE
_ _ _
minutes
MAX
None
LIMIT
None
DOSE
MODE
Guardrails Drug Setup
IMS Pre-fill 30mL
20
PCA
mg
DOSE
10
minutes
Yes
MAX
> > > >
No
LIMIT
None
DOSE
MODE
Programming
3-13