Syringe Module (Continued)
Pressure Sensing Disc Installed
(Continued)
4.
Press Pressure Limit soft key.
5.
To enter a new pressure limit value, press Change Value
soft key.
OR
If Auto Pressure feature is enabled, press Auto Pressure
soft key.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Select Pressure Limit (Continued)
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - 550 mmHg
Prime Set with Syringe
>Select an Option or EXIT
EXIT
A
Pressure Limit Selection
Current pressure
30
mmHg
Current limit
_
550
mmHg
General Programming
PAGE
DOWN
Auto
Pressure
Change
value
CONFIRM