Electromagnetic Environment (Continued)
Alaris System (Continued)
Immunity Test
IEC 61000-4-2
Electrostatic Discharge
(ESD)
IEC 61000-4-4
Electrical Fast Transient,
Burst (EFT)
IEC 61000-4-5
Power Line Surge
IEC 61000-4-8
Power Frequency
Magnetic Field
(50/60 Hz)
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Compliance (Continued)

Electromagnetic Immunity

IEC
60601-1-2:2007
Test Level
For Model 8015,
±8 kV contact
Infusion modules, and
±15 kV air
Auto-ID module:
For Model 8015,
±8 kV contact
Infusion modules and
±15 kV air
Auto-ID module:
For Model 8015 and
±8 kV contact
monitoring modules:
±15 kV air
±6 kV contact
For Model 8015 and
±8 kV air
monitoring modules:
±6 kV contact
±8 kV air
±2 kV for power supply
±2 kV for power
lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV differential mode
±1 kV differential
±2 kV common mode
mode
±2 kV common mode
3 A/m
400 A/m 50 Hz
400 A/m 60 Hz
Table 2
Compliance
Level
Floors should be wood, concrete, or ceramic
tile.
If floors are covered with synthetic material, relative
humidity should be at least 30%.
If connector testing exemption is used, following ESD
sensitivity symbol appears adjacent to each connector.
Alaris System User Manual – with v9.33 Model 8015" -
Do Not Touch"
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
Electromagnetic Environment -