Specifications and Symbols (Continued)
Feature
Max Accumulated Dose Range
Max Rate (for Continuous Dose)
NEOI
Alert Time
NEOI Snooze
Occlusion Pressure Set Point
PCA Pause Protocol:
PCA Pause Protocol
Monitoring Module Attach
Enforcement
PCA Pause Protocol Text
SpO
Settings:
2
% SpO
Low Limit
2
Initial Value
EtCO
Settings:
2
Respiratory Rate Lower Limit
(bpm)
Initial Value
Priming
Security Access Level
NOTES:
This configuration setting is a shared setting between the PCA
module and the Syringe module.
These values are configured in the SpO
within the Editor Software and can be changed by the clinician
by accessing Channel Options on the PCA module.
These values are configured in the EtCO
within the Editor Software and can be changed by the clinician
by accessing Channel Options on the PCA module.
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Default Setting
4-hour limit
999 mL/h
Disabled
Disabled
High (800 mmHg)
Disabled
None
PCA infusion has paused due
to a decline in respiratory
status. Check patient.
None
None
None
None
Enabled
Level 1
module settings
2
module settings
2
Options
Disabled; 1, 2, or 4-hour limit
0.1 - 99.9 mL/h in 0.1 mL/h increments;
100 - 999 mL/h in 1 mL/h increments
Enabled - Disabled
5 - 25% of syringe size
5, 10 or 15 minutes
Low (200 mmHg)
Medium (500 mmHg)
High (800 mmHg)
Enabled - Disabled
Enabled - Disabled
Editable per hospital protocol
20 - 99
20 - 99
0 - 149
0 - 149
Enabled- Disabled
Level 1, 2, 3
General Information
3-57