CareFusion Alaris 8015 User Manual page 184

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Syringe Module (Continued)
Pressure Sensing Disc NOT Installed
WARNING
When the pressure sensing disc is not being used and an
occlusion occurs, there is a risk of infusing pressurized
buildup of infusates upon correction of the occlusion. To
avoid an inadvertent bolus, relieve the pressure before
restarting the infusion.
1.
Press
CHANNEL SELECT
2.
Press
OPTIONS
3.
Press Pressure Limit soft key.
4.
To select a pressure limit, press appropriate soft key.
5.
Press
CONFIRM
General Programming
2-118
Select Pressure Limit (Continued)
key.
key.
soft key.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - High
Prime Set with Syringe
>Select an Option or EXI T
PAGE
EXIT
DOWN
A
Pressure Limit Selection
Low
Med
High
CONFIRM