Respiration Rate Test (continued)
Test Action
Measure Accuracy/RR Tolerance mmHg / bpm for
Adult/Pediatric FilterLine Sets at respiration rates of
10, 40, and 60 bpm.
Measure CO2 (mmHg) / RR (bpm) for
Adult/Pediatric FilterLine Sets at respiration rates of
10, 40, and 60 bpm.
Measure Accuracy/RR Tolerance mmHg / bpm for
Infant/Neonatal FilterLine H Sets at respiration rates
of 60, 80, 120 and 150 bpm.
Measure CO2 (mmHg) / RR (bpm) for
Infant/Neonatal FilterLine H Sets at respiration rates
of 60, 80, 120 and 150 bpm.
NOTE:
At sea level.
Accuracy applies for breath rates of up to 80 bpm. For breath
rates above 80 bpm, accuracy is 4 mmHg or ±12 % of reading
whichever is greater, for EtCO
mmHg.This is tested according to and is compliant with ISO
80601-2-55. To achieve the specified accuracies for breath
rates above 60 breaths/minute, the Microstream FilterLine H
Set for Infant/Neonatal must be used.
General Information
5-26
Measurement Accuracy (continued)
values exceeding 18
2
Expected Results
10 bpm
40 bpm
60 bpm
CO2 Partial Pressure
0-38 mmHg
39-99 mmHg
100-150 mmHg***
60 bpm
80 bpm
120 bpm
150 bpm
CO2 Partial Pressure
0-38 mmHg
39-99 mmHg
100-150 mmHg***
Alaris System User Manual – with v9.33 Model 8015
±2mmHg/±1bpm
±2mmHg/±1bpm
Accuracy
± 2 mmHg
± (5% of reading + 0.08 for
every 1 mmHg above 38
mmHg)
± (5% of reading + 0.08 for
every 1 mmHg above 38
mmHg)
±2mmHg/±1bpm
±2mmHg/±2bpm
±4.1mmHg/±2bpm
±4.1mmHg/±3bpm
Accuracy
± 2 mmHg
± (5% of reading + 0.08 for
every 1 mmHg above 38
mmHg)
± (5% of reading + 0.08 for
every 1 mmHg above 38
mmHg)
EtCO
Module Section
2