Guardrails PCA Pause Protocol Feature (Continued)
Review or Change PCA Pause Alarm Limits (Continued)
5.
To change PCA pause limits, press soft key that
corresponds to alarm limit and enter a value within
acceptable range.
6.
Press
CONFIRM
7.
Press
soft key.
START
NOTE:
The acceptable range for PCA Pause Protocol is configurable
and defined by the hospital within the Data Set using the
Guardrails Suite MX.
The
PCA PAUSE LIMITS
. A prompt is provided if the
ALARM LIMITS
must be modified.

Disable PCA Pause Alarm

1.
From Main Display press
2.
Press
OPTIONS
3.
Press PCA Pause Limits soft key.
4.
Press
DISABLE SPO2
appropriate.
Disabling SpO
PCA Pause feature only, without interrupting
monitoring functionality.
Once disabled, alarm limits are grayed out and are not
editable.
Programming
3-40

soft key.
must be lower than the
CHANNEL SELECT
key.
or
DISABLE ETCO2
or EtCO
from this screen discontinues
2
2
SPO2/ETCO2
PCA PAUSE LIMITS
.
soft key, as
Alaris System User Manual – with v9.33 Model 8015
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
%SPO2
97
88
LOW
RR
_4
5
LOW
>Enter a Pause Limit
Between 2 - 4
DISABLE
DISABLE
CONFIRM
SPO2
ETCO2
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
%SPO2
97
88
LOW
RR
5
_4
LOW
>Enter a Pause Limit
Between 2 - 4
DISABLE
DISABLE
CONFIRM
SPO2
ETCO2