Manual Programming—Primary Infusion (Continued)
Bolus Dose (Continued)
6.
To enter bolus duration, use numeric data entry keys.
OR
To deliver Bolus Dose at maximum safe rate possible for
selected drug and setup, and automatically calculate
bolus duration, press Rapid Bolus soft key.
TOTAL DOSE
7.
Verify correct parameters and press
If the programmed Bolus Dose and/or Bolus Dose
duration is outside the Soft Limit for that care area, an
audio alert sounds and a visual prompt appears before
programming can continue. If Yes soft key is pressed,
programming continues; if No soft key is pressed,
infusion needs to be reprogrammed.
If the programmed Bolus Dose and/or Bolus Dose
duration is outside the Hard Limit for that care area, an
audio alert sounds and a visual prompt appears before
programming can continue. Infusion needs to be
reprogrammed.
Manual Programming with Guardrails Suite MX
2-36
alternates with
INFUSE AT
START
rate.
soft key.
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A A A
Alteplase
Stroke
BOLUS DOSE
0.090
DOSE
mg/kg
PATIENT
70
kg
WEIGHT
(420 mL/h)
Rapid
_ _
DURATION
min
Bolus
TOTAL DOSE = 6.3 mg
BOLUS VTBI = 6.3 mL
>Enter Duration
CONT-
SETUP
INUOUS
A A A
A A A
Guardrails Drug Library
Alteplase
Dopamine
Stroke
BOLUS DOSE
BOLUS DOSE
5
0.090
mcg/kg
DOSE
DOSE
mg/kg
PATIENT
PATIENT
71
70
kg
kg
WEIGHT
WEIGHT
(420 mL/h)
Rapid
<1
1
DURATION
DURATION
min
min
Bolus
INFUSE AT: 0.09 mg/kg/min
BOLUS VTBI = 50.0 mL
[Conc]: 1 mg/mL
[Conc]: 800 mcg/mL
>Press START to Begin
>Press Start to Begin
Infusing Bolus Dose
Infusing Bolus Dose
CONT-
CONT-
PAUSE
PAUSE
SETUP
SETUP
START
START
INUOUS
INUOUS
Pump and Syringe Modules Section
OR

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents