:MEASure:VSTArt

Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
The short form of this command, VSTA, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:Y1Position command (see
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:Y1Position"
":MARKer:Y2Position"
":MARKer:YDELta"
on page 375
":MARKer:X1Y1source"
":MEASure:SOURce"
":MEASure:TDELta"
on page 1019
":MEASure:TSTArt"
on page 1023
Obsolete and Discontinued Commands
page
1090). The normal short form, VST, would be the same for
on page 364
on page 388
on page 373
on page 374
on page 367
on page 421
34
page
373) instead.
1029

Advertisement

Table of Contents
loading

Table of Contents