31
:WGEN Commands
:WGEN:MODulation:AM:DEPTh
Command Syntax
:WGEN:MODulation:AM:DEPTh <percent>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used
by the modulation. For example, a depth setting of 80% causes the output
amplitude to vary from 10% to 90% (90% – 10% = 80%) of the original
amplitude as the modulating signal goes from its minimum to maximum
amplitude.
Query Syntax
:WGEN:MODulation:AM:DEPTh?
The :WGEN:MODulation:AM:DEPTh? query returns the AM depth
percentage setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also
•
•
•
•
•
•
•
•
•
672
(see
page
790)
":WGEN:MODulation:AM:FREQuency"
":WGEN:MODulation:FM:DEViation"
":WGEN:MODulation:FM:FREQuency"
":WGEN:MODulation:FSKey:FREQuency"
":WGEN:MODulation:FSKey:RATE"
":WGEN:MODulation:FUNCtion"
":WGEN:MODulation:FUNCtion:RAMP:SYMMetry"
":WGEN:MODulation:STATe"
":WGEN:MODulation:TYPE"
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
on page 673
on page 674
on page 675
on page 676
on page 677
on page 678
on page 681
on page 682
on page 679