Wgen:modulation:am:depth - Agilent Technologies InfiniiVision 2000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 2000 X-Series:
Table of Contents

Advertisement

31
:WGEN Commands

:WGEN:MODulation:AM:DEPTh

Command Syntax
:WGEN:MODulation:AM:DEPTh <percent>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used
by the modulation. For example, a depth setting of 80% causes the output
amplitude to vary from 10% to 90% (90% – 10% = 80%) of the original
amplitude as the modulating signal goes from its minimum to maximum
amplitude.
Query Syntax
:WGEN:MODulation:AM:DEPTh?
The :WGEN:MODulation:AM:DEPTh? query returns the AM depth
percentage setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also
672
(see
page
790)
":WGEN:MODulation:AM:FREQuency"
":WGEN:MODulation:FM:DEViation"
":WGEN:MODulation:FM:FREQuency"
":WGEN:MODulation:FSKey:FREQuency"
":WGEN:MODulation:FSKey:RATE"
":WGEN:MODulation:FUNCtion"
":WGEN:MODulation:FUNCtion:RAMP:SYMMetry"
":WGEN:MODulation:STATe"
":WGEN:MODulation:TYPE"