Table of Contents

Advertisement

B - Verification and Calibration
Equipment
Digital Voltmeter
1
Current Monitor
2
Ratio Transformer
Load Resistor
Impedance Resistor
GPIB Controller
1
The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the
load leads and connections. It has special current-monitoring terminals inside the load connection terminals. Connect
the voltmeter directly to these current-monitoring terminals.
2
A ratio transformer is required only when verifying output voltage readback to MIL-STD-45662A 4:1 test
equipment ratio requirements.

Test Setup

Figure B-1 shows the setup for the tests. Be certain to use load leads of sufficient wire gauge to carry the
full output current (see Chapter 2).
458A
DMM
R = Load resistor for CC test (20 ohms)
L
Rs = Current Monitor resistor ( 0.01 ohms )
Rz = Impedance resistor ( 1 ohm )
S1 = Switch is for convenience (not required).
66
Table B-1. Equipment Required
Characteristics
Resolution: 10 nV @ 1 V
Readout: 8.5 digits
Accuracy: >20 ppm
0.01 , 200 ppm, 10 Watts
30:1 ratio, 50 ppm, 45 Hz to 1 kHz
20 , 10 A, 1800 Watts min.
1 , 100 Watts min.
Full GPIB capabilities
R
Rs
L
3458A
Figure B-1. Verification & Calibration Test Setup
Keysight 3458A
Guildline 7320/0.01
HP Series 200/300 or equivalent
01
01
S1
1
30
Recommended Model
811B/6812B/6813B
SENSE
COM
COM
R
Z

Advertisement

Table of Contents
loading

This manual is also suitable for:

6812b6813b ac

Table of Contents