Trigger[:Edge]:Coupling - Agilent Technologies InfiniiVision 3000 Programmer's Manual

X-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

:TRIGger[:EDGE]:COUPling

Command Syntax
:TRIGger[:EDGE]:COUPling <coupling>
<coupling> ::= {AC | DC | LFReject}
The :TRIGger[:EDGE]:COUPling command sets the input coupling for the
selected trigger sources. The coupling can be set to AC, DC, or LFReject.
• AC coupling places a high- pass filter (10 Hz for analog channels, and
• LFReject coupling places a 50 KHz high- pass filter in the trigger path.
• DC coupling allows dc and ac signals into the trigger path.
The :TRIGger[:EDGE]:COUPling and the :TRIGger[:EDGE]:REJect selections are coupled.
N O T E
Changing the setting of the :TRIGger[:EDGE]:REJect can change the COUPling setting.
Query Syntax
:TRIGger[:EDGE]:COUPling?
The :TRIGger[:EDGE]:COUPling? query returns the current coupling
selection.
Return Format
<coupling><NL>
<coupling> ::= {AC | DC | LFR}
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1088)
3.5 Hz for all External trigger inputs) in the trigger path, removing dc
offset voltage from the trigger waveform. Use AC coupling to get a
stable edge trigger when your waveform has a large dc offset.
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 851
":TRIGger[:EDGE]:REJect"
on page 843
on page 869
30
:TRIGger Commands
867

Advertisement

Table of Contents
loading

Table of Contents