:MEASure Commands

:MEASure:SOURce

Command Syntax
:MEASure:SOURce <source1>[,<source2>]
<source1>,<source2> ::= {<digital channels> | CHANnel<n> | FUNCtion
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :MEASure:SOURce command sets the default sources for
measurements. The specified sources are used as the sources for the
MEASure subsystem commands if the sources are not explicitly set with
the command.
If a source is specified for any measurement, the current source is
changed to this new value.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, or MATH will also set :MARKer:X1Y1source to
source1 and :MARKer:X2Y2source to source2.
EXTernal is only a valid source for the counter measurement (and
<source1>).
Query Syntax
:MEASure:SOURce?
The :MEASure:SOURce? query returns the current source selections. If
source2 is not specified, the query returns "NONE" for source2. If all
channels are off, the query returns "NONE,NONE". Source2 only applies to
:MEASure:DELay and :MEASure:PHASe measurements.
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
N O T E
MATH.
Return Format
<source1>,<source2><NL>
<source1>,<source2> ::= {<digital channels> | CHAN<n> | FUNC | WMWM<r>
See Also:
458
(see
page
1208)
"Introduction to :MEASure Commands"
":MARKer:MODE"
on page 394
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:DELay"
on page 430
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
| MATH | WMEMory<r> | EXTernal}
| EXT | NONE}
on page 420
on page 396
on page 398

Advertisement

Table of Contents
loading

Table of Contents