Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Auto Analysis
Auto Analysis reports the following TEN pulse measurements:
1 - Average Power
2 - Peak Power
3 - Peak minus
Average
4 - Pulse Top
5 - Rise Time
6 - Fall Time
7 - Pulse Width
8 - Pulse Period
9 - Duty Cycle
10 - Pulse Rep Freq
How to enable Auto Analysis
— Press Mrk ->/Tools.
— Then
— Then choose from the following:
Keysight N9912-90001 User's Guide
Average power level of all
measured data points
Highest reading of all measured
data points
Difference between the above two
results
Power value based on the specified
Pulse Top setting.
The time it takes for a pulse to rise
from 10% to 90% of its peak value.
The time it takes for a pulse to fall
from 90% to 10% of its peak value.
The time between the rising edge
and falling edge of a pulse.
The time of complete pulse cycle.
The ratio of the pulse width / pulse
period.
The rate, in pulses / sec, at which
pulses recur.
Auto
Analysis.
—
OFF
Analysis OFF
—
ON
The Analysis values appear below trace 1.
13- 13