Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Then choose from the following:
— Target, Peak, and Minimum. Learn more about these standard features
Rise and Fall time in Marker Search uses different criteria than in Auto
Analysis and will therefore result in different reported values.
1 - Average Power
2 - Peak Power
3 - Peak minus Average
4 - Pulse Top
5 - Rise Time
6 - Fall Time
7 - Pulse Width
8 - Pulse Period
9 - Duty Cycle
10 - Pulse Rep Freq
Keysight N9927-90001 User's Guide
beginning in
"Searching with Markers" on page
Rise Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value BEFORE the Peak.
Fall Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value AFTER the Peak.

Auto Analysis

Auto Analysis reports the following TEN pulse measurements:
Average power level of all measured data
points
Highest reading of all measured data
points
Difference between the above two results
Power value based on the specified Pulse
Top setting.
The time it takes for a pulse to rise from
10% to 90% of its peak value.
The time it takes for a pulse to fall from
90% to 10% of its peak value.
The time between the rising edge and
falling edge of a pulse.
The time of one complete pulse cycle
The ratio of the pulse width / pulse period.
The rate, in pulses / sec, at which pulses
recur.