Pulse Measurements Mode – Option 330
Pulse Measurement Settings
1 - Average Power
2 - Peak Power
3 - Peak minus
Average
4 - Pulse Top
5 - Rise Time
6 - Fall Time
7 - Pulse Width
8 - Pulse Period
9 - Duty Cycle
10 - Pulse Rep Freq
How to enable Auto Analysis
— Press Mrk ->/Tools.
— Then
— Then choose from the following:
Keysight N9923-90001 User's Guide
Auto Analysis
Auto Analysis reports the following TEN pulse measurements:
Average power level of all
measured data points
Highest reading of all measured
data points
Difference between the above two
results
Power value based on the specified
Pulse Top setting.
The time it takes for a pulse to rise
from 10% to 90% of its peak value.
The time it takes for a pulse to fall
from 90% to 10% of its peak value.
The time between the rising edge
and falling edge of a pulse.
The time of one complete pulse cycle
The ratio of the pulse width / pulse
period.
The rate, in pulses / sec, at which
pulses recur.
Auto Analysis
OFF
—
Analysis OFF
ON
—
The Analysis values appear below trace 1.
157