Keysight E4980A/AL Precision LCR Meter
User's Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight E4980AL

  • Page 1 Keysight E4980A/AL Precision LCR Meter User’s Guide...
  • Page 2 EULA. governed by United States and Declaration of Conformity Keysight shall be under no obligation international copyright laws. to update, revise or otherwise modify Trademark Acknowledgments Declarations of Conformity for this the Software.
  • Page 3: Table Of Contents

    1. GPIB Interface Connector ..........37 Keysight E4980A/AL User’s Guide...
  • Page 4 MEAS SETUP page ............88 Keysight E4980A/AL User’s Guide...
  • Page 5 SYSTEM INFO Page ............155 Keysight E4980A/AL User’s Guide...
  • Page 6 Parallel/Series Circuit Mode ..........194 Keysight E4980A/AL User’s Guide...
  • Page 7 Trigger System ............249 Keysight E4980A/AL User’s Guide...
  • Page 8 *ESR? ..............322 Keysight E4980A/AL User’s Guide...
  • Page 9 :CORRection:OPEN[:EXECute] ..........338 Keysight E4980A/AL User’s Guide...
  • Page 10 :FUNCtion:SMONitor:VDC[:STATe] ......... .355 Keysight E4980A/AL User’s Guide...
  • Page 11 :SYSTem:COMMunicate:LAN[:SELF]:CONTrol ....... . . 372 :SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess? ..... . . 372 Keysight E4980A/AL User’s Guide...
  • Page 12 Recommended Calibration Period ......... .424 Keysight E4980A/AL User’s Guide...
  • Page 13 List of Default Values ............451 Keysight E4980A/AL User’s Guide...
  • Page 14 Channel Selection Input Signals ..........504 Keysight E4980A/AL User’s Guide...
  • Page 15 Correction Data Confirmation ..........517 G.Functional Comparison between 4284A, 4279A and E4980A/AL Functional Comparison between 4284A, 4279A, and E4980A/AL ....521 Keysight E4980A/AL User’s Guide...
  • Page 16 Contents Keysight E4980A/AL User’s Guide...
  • Page 17: Unpacking And Preparation

    Keysight E4980A/AL Precision LCR Meter User’s Guide Unpacking and Preparation This chapter describes how to set up and start the Keysight E4980A/AL Precision LCR Meter. Contents of this Chapter Checking the Shipment on page? After you receive the E4980A/AL, check all the items in the packing container.
  • Page 18: Checking The Shipment

    1-1, check that all packaged items supplied with the LCR meter have been provided as per the specified options. Step 4. After checking, if one of the following applies, contact your nearest Keysight Technologies sales and service office. 1. The packing box or shock-absorbing material used to package the LCR meter has been damaged or the shock-absorbing material displays traces of where extreme pressure has been applied.
  • Page 19 . A 3p-2p conversion adapter is not supplied. Figure 1-1 2. This accessory is not shown in . For how to install the rack mount kit, refer to the attached manual. Figure 1-1 E4980A/AL Accessories Keysight E4980A/AL Precision LCR Meter...
  • Page 20: Preparations Before Use

    150 VA Setting up the Fuse Please use the following fuse type. UL/CSA type, Slo-Blo, 5x20-mm miniature fuse, 3 A, 250 V (Keysight part number 2110-1017) When you need a fuse, contact your nearest Keysight Technologies sales and service office. To verify and replace the fuse, remove the power cable and pull out the fuse holder.
  • Page 21: Verifying And Connecting The Power Cable

    Use the supplied three-wire power cable with a grounding wire to securely ground the E4980A/AL. A 3p-2p conversion adapter is not supplied with the LCR meter. When you need a 3p-2p conversion adapter, contact your nearest Keysight Technologies sales and service office listed in the back of this manual. Figure 1-3 shows the power cable options.
  • Page 22 Unpacking and Preparation Preparations before Use Figure 1-2 Power cable options Keysight E4980A/AL Precision LCR Meter...
  • Page 23: How To Remove The Handle

    Step 1. Lift the handle perpendicular to the unit while pulling it in the direction of 1. Step 2. While pulling the handle in the direction of 1, lift towards 2. Install the handle with 3 facing toward the front. Installing the handle the other way around may damage it. Keysight E4980A/AL Precision LCR Meter...
  • Page 24: Caution When Using The Handle

    When the handle is set up as shown in Figure 1-5, do not put any additional weight onto the E4980A/AL or lift it suddenly. • Do not touch the handle while the DUT is attached. Figure 1-4 The handle in the set-up position Keysight E4980A/AL Precision LCR Meter...
  • Page 25: Environmental Requirements

    0 to 2,000 m (0 to 6,561 feet) Vibration Max. 0.5 G, 5 Hz to 500 Hz The environmental requirements listed above are NOT for the specifications and measurement accuracy of the E4980A/AL, but for its operating environment. Keysight E4980A/AL Precision LCR Meter...
  • Page 26: Ventilation Requirements

    The LCR meter still conforms to the requirements of the safety standard it is installed with the following cooling clearance: Requirements Back ≥ 180 mm Sides ≥ 60 mm (both right and left) Figure 1-5 Ventilation space at installation Keysight E4980A/AL Precision LCR Meter...
  • Page 27: Protection Against Electrostatic Discharge (Esd)

    E4980A/AL. When installing the E4980A/AL, ensure that there is sufficient free space around the unit to permit quick disconnection of the plug (from the AC outlet or the E4980A/AL unit) in case of emergency. Keysight E4980A/AL Precision LCR Meter...
  • Page 28: Starting The E4980A/Al

    Do not turn the power OFF while saving into or recalling from the internal memory of the E4980A/AL or the USB memory. Doing so may clear the contents from the memory. Keysight E4980A/AL Precision LCR Meter...
  • Page 29: Disconnecting From The Supply Source

    “Ensuring Adequate Free Space around the LCR meter for Immediate Disconnection of the Power Cable in Case of Emergency” on page When turning the power OFF under normal circumstances, always follow the methods described in “Turning the Power ON and OFF” on page Keysight E4980A/AL Precision LCR Meter...
  • Page 30 Unpacking and Preparation Starting the E4980A/AL Keysight E4980A/AL Precision LCR Meter...
  • Page 31: Overview

    Product Introduction The Keysight E4980A/AL is a general-purpose LCR meter for incoming inspection of components, quality control, and laboratory use. The E4980A/AL is used for evaluating LCR components, materials, and semiconductor devices over a wide range of frequencies (20 Hz to 300 kHz/500 kHz/1 MHz/2 MHz) and test signal levels (0.1 mVrms to 2 Vrms, 50 A to 20 mArms).
  • Page 32: Front Panel: Names And Functions Of Parts

    NOT touch UNKNOWN terminals, DC source terminals, or electrodes of test fixtures. Turn off the output of DC Bias and DC source and confirm that LED indicators are off when setting up the E4980A/AL and its accessories. Keysight E4980A/AL Precision LCR Meter...
  • Page 33: Power Switch

    When the cursor is moved to a certain field, that field changes to an inverse video image of the original field. The cursor can only be moved from field to field. For the skip key, refer to the “How to Use Skip Keys” on page Keysight E4980A/AL Precision LCR Meter...
  • Page 34: Entry Keys

    A key used to toggle on and off of the DC source output. When the DC source output is set to on, the DC Source indicator lights up and DCSRC is displayed in the status display area on the screen. Option 001 is required for this function. Keysight E4980A/AL Precision LCR Meter...
  • Page 35: Unknown Terminals

    USB port of printers, devices with a built-in HDD, or USB hubs. To print a screen, refer to “To save a screenshot into USB memory” on page 188. Do not unplug the USB memory while the USB indicator is on. Keysight E4980A/AL Precision LCR Meter...
  • Page 36: Ground Terminal

    Outputs the DC source in the range from -10 V to 10 V. Option 001 is required for this function. The DC source terminal is installed even in the E4980A/AL without Option 001; however, the E4980A/AL without Option 001 cannot use this terminal. Keysight E4980A/AL Precision LCR Meter...
  • Page 37: Rear Panel: Names And Functions Of Parts

    Interfaces with which the E4980A/AL can be equipped as options are shown below. Handler interface (Option E4980A/AL-201) This interface is for data exchange with an automatic machine (handler) used on a production line. Connector type: 36-pin Centronics Keysight E4980A/AL Precision LCR Meter...
  • Page 38: Usb (Usbtmc) Interface Port

    The BNC connector to input the positive TTL pulse to trigger the E4980A/AL with external trigger signals.(The trigger mode must be set to EXTernal.) 6. Serial Number Plate The seal showing the product’s serial number. Keysight E4980A/AL Precision LCR Meter...
  • Page 39: Power Cable Receptacle (To Line)

    “Verifying and Connecting the Power For more on the power supply, see Cable” on page 8. Fan The cooling fan for controlling the temperature inside the E4980A/AL. This fan extracts heated air from inside the LCR meter. Keysight E4980A/AL Precision LCR Meter...
  • Page 40: Screen Area: Names And Functions Of Parts

    MEAS DISPLAY page • BIN No. DISPLAY page • BIN COUNT DISPLAY page • LIST SWEEP DISPLAY page • MEAS SETUP page For the input method in the comment line, refer to “MEAS SETUP page” on page Keysight E4980A/AL Precision LCR Meter...
  • Page 41: Softkey Area

    This means that no measurement has been made or the data format will not hold the measurement results. INFINITY This message is displayed when a deviation mode of the deviation measurement function is % and measurement results cannot be calculated. Keysight E4980A/AL Precision LCR Meter...
  • Page 42: Input Line Area

    Displays the status when DC bias or DC source is on, or any front panel key is locked. When sending SCPI commands from the external controller, “RMT” is displayed and the front panel keys are locked. Keysight E4980A/AL Precision LCR Meter...
  • Page 43: Basic Operation

    The unit changes according to the field selected. How to Use Cursor Keys Move the cursor to the desired field using the cursor keys as shown in Figure 2-5. Figure 2-5 Cursor keys and a field operation example Keysight E4980A/AL Precision LCR Meter...
  • Page 44: How To Use Skip Keys

    On a display page, three lines are collected together as one. By using a skip key, you can select the desired field quickly as the selected field moves from area to area. Figure 2-6 Skip key and a field operation example Keysight E4980A/AL Precision LCR Meter...
  • Page 45: Display Format

    Keysight E4980A/AL Precision LCR Meter User’s Guide Display Format This chapter describes each page of the DISPLAY FORMAT MENU of the E4980A/AL. MEAS DISPLAY Page When you press the [DISPLAY FORMAT] key, the MEAS DISPLAY page appears. The following measurement controls can be set on this page (The field in parenthesis is used to set measurement controls).
  • Page 46 Pressing the [DISPLAY FORMAT] key when is displayed enlarges the measurement result as shown in Figure 3-2. Pressing the [DISPLAY FORMAT] key again returns to the display in Figure 3-1. Figure 3-2 The enlarged MEAS DISPLAY for measurement results Keysight E4980A/AL Precision LCR Meter...
  • Page 47: Measurement Function

    2. This parameter can be set only when the E4980A is equipped with option 001. When the measurement parameter is DC voltage measurement/DC current measurement (Vdc-Idc), set the setting value of the test signal level (LEVEL field) to 0 V for accurate measurement. Keysight E4980A/AL Precision LCR Meter...
  • Page 48 Dissipation factor Quality factor (inverse of dissipation factor) Conductance Equivalent series resistance measured using the series equivalent circuit model Equivalent parallel resistance measured using the parallel equivalent circuit model Reactance Sustenance  Phase angle DC current Keysight E4980A/AL Precision LCR Meter...
  • Page 49 Y-d Cs-D Cp-D Cs-Q Cp-Q Cs-Rs Cp-G Cp-Rp Ls-D Lp-D Ls-Q Lp-Q Ls-Rs Lp-G Ls-Rdc Lp-Rp Lp-Rdc ----- ----- Primary parameter Secondary parameter Measurement function Vdc-Idc does not have the series mode and parallel mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 50 Step 4. Use the softkeys to select the primary parameter. Step 5. If the secondary parameter exists, select the secondary parameter from measurement parameters displayed using the softkeys. For the combinations of measurement parameters, refer to Table 3-1 on page Keysight E4980A/AL Precision LCR Meter...
  • Page 51: Impedance Range

    Option 001 100 m , 1, 10, 100, 300, 1 k, 3k, 10 k, 30 k, 100k 1. This cannot be used when the test signal level is less than or equal to 2 V. Keysight E4980A/AL Precision LCR Meter...
  • Page 52 < Impedance  1.08 108 m 1.08 < Impedance  10.08 10.08 < Impedance < 276 276  Impedance < 0.92 k 0.92 k  Impedance < 2.76 k 2.76 k  Impedance < 9.2 k Keysight E4980A/AL Precision LCR Meter...
  • Page 53 If you set the impedance range first and then frequency, the resulting impedance range may not be the one you wanted to set. Keysight E4980A/AL Precision LCR Meter...
  • Page 54 Display Format MEAS DISPLAY Page Effective measurement range (0.2V or 2 mA < Test signal level  2 V or 20 mA) Figure 3-4 Keysight E4980A/AL Precision LCR Meter...
  • Page 55 Display Format MEAS DISPLAY Page Figure 3-5 Effective measurement range (Test signal level > 2 V or 20 mA) Keysight E4980A/AL Precision LCR Meter...
  • Page 56 Display Format MEAS DISPLAY Page Effective measurement range (Test signal level  0.2 V or 2 mA) Figure 3-6 Keysight E4980A/AL Precision LCR Meter...
  • Page 57 Manually setting the time for the impedance range mode Changing impedance range requires a few milliseconds to several tens of data sheet Chapter 11 milliseconds. For details, refer to Settling Time in “Specifications and Supplemental Information” Keysight E4980A/AL Precision LCR Meter...
  • Page 58: Test Frequency

    Increments the current test frequency to the next sequentially higher frequency point. There are ten frequency points between successive decade values. Refer to Table 3-9 for the sequential frequency points that can be set using this softkey. Keysight E4980A/AL Precision LCR Meter...
  • Page 59 300 kHz 80 Hz 400 Hz 4 kHz 40 kHz 400 kHz 500 Hz 5 kHz 50 kHz 500 kHz 600 Hz 6 kHz 60 kHz 600 kHz 800 Hz 8 kHz 80 kHz 800 kHz Keysight E4980A/AL Precision LCR Meter...
  • Page 60 Time required for setting the test frequency Changing the test frequency requires a few milliseconds to several tens of data sheet Chapter 11 milliseconds. For details, refer to Settling Time in “Specifications and Supplemental Information” Keysight E4980A/AL Precision LCR Meter...
  • Page 61: Test Signal Level

    Oscillator current level and resolution (Std.) Oscillator current level Resolution 1 Arms 0 Arms to 2 mArms 2 Arms 2 mArms to 5 mArms 5 Arms 5 mArms to 10 mArms 10 Arms 10 mArms to 20 mArms Keysight E4980A/AL Precision LCR Meter...
  • Page 62 50 mArms to 100 mArms Test signal level setting procedure Step 1. Press the [Display Format] key Step 2. Press the MEAS DISPLAY softkey. Step 3. Use the cursor keys to select the LEVEL field. Keysight E4980A/AL Precision LCR Meter...
  • Page 63 1 m, 2 m, 3 m, •••, 9 m 10 m, 20 m, 30 m, •••, 90 m 100 m 1. The current level can be set to more than 20 m only when option 001 is installed. Keysight E4980A/AL Precision LCR Meter...
  • Page 64: Dc Bias

     (0 V to 5 V) 100 V  (5 V to 10 V) 1 mV  (10 V to 20 V) 2 mV  (20 V to 40 V) 5 mV 1. Effective resolution is 330 V. Keysight E4980A/AL Precision LCR Meter...
  • Page 65 Test signal level Vdc (V) Vosc (Vrms) Vosc x √2 x 1.15 + Vdc x 1.002 < 42 V Vdc (V) Iosc (Arms) Iosc x √2 x 115 + Vdc x 1.002 < 42 V Keysight E4980A/AL Precision LCR Meter...
  • Page 66 1 m, 2 m, 3 m, •••, 9 m 10 m, 20 m, 30 m, •••, 90 m 100 m, 200 m, 300 m, •••, 900 m 1, 2, 3, •••, 9 10, 20, 30, 40 Keysight E4980A/AL Precision LCR Meter...
  • Page 67 Settling time for the DC bias signal voltage Changing the DC bias requires a few milliseconds to several tens of milliseconds. For details, refer to DC bias settling time in data sheet Chapter 11 “Specifications and Supplemental Information” Keysight E4980A/AL Precision LCR Meter...
  • Page 68: Measurement Time Mode

    Step 3. Use the cursor keys to select the MEAS TIME field. Step 4. Use the softkeys to set the measurement time. Softkey Function SHORT Shortens the measurement time Sets the measurement time between SHORT and LONG LONG Lengthens the measurement time Keysight E4980A/AL Precision LCR Meter...
  • Page 69: Display Setting For Measurement Results

    Step 2. Press the MEAS DISPLAY softkey. Step 3. Use the cursor keys to select the field in which a measurement result of primary parameter or secondary parameter is displayed. Step 4. Use the following softkeys to set the display. Keysight E4980A/AL Precision LCR Meter...
  • Page 70 Moves the decimal point to the left. Supplementary units are also changed. D.P POS DECL - Moves the decimal point to the right. Supplementary units are also changed. Figure 3-8 Fixed-display sample of a measurement result Keysight E4980A/AL Precision LCR Meter...
  • Page 71: Displaying Errors Instead Of Measurement Results

    ( ) (RANGE field) 0 to 0.11 more than 0.11 0 to 1.1 more than 1.1 0 to 11 more than 11 0 or more n.a. 270 or more less than 270 900 or more less than 900 Keysight E4980A/AL Precision LCR Meter...
  • Page 72 (DCR RNG field) occurs 0 to 11 more than 11 0 or more n.a. 900 or more less than 900 10 k 9,000 or more less than 9,000 100 k 90,000 or more less than 90,000 Keysight E4980A/AL Precision LCR Meter...
  • Page 73 This section describes measurement results and each monitor value when overload occurs. The bold font in the list indicates the measurement parameter for which overload occurs. “*” indicates that measurement results, judgment, and BIN count are normal. Keysight E4980A/AL Precision LCR Meter...
  • Page 74 OVLD measurement (except for Ls/Lp-Rdc) OVLD Impedance measurement (Ls/Lp-Rdc) OVLD DCR measurement (Ls/Lp-Rdc) Vdc-Idc measurement OVLD OVLD OVLD Other than Vdc-Idc measurement Other than Vdc-Idc OVLD measurement OVLD is displayed as OVERLOAD on the display. Keysight E4980A/AL Precision LCR Meter...
  • Page 75: Monitor Information

    The VDC/IDC monitor values are the ones measured under the condition that the setting value of the test signal level (LEVEL field) is output. Therefore, these values are different from the measurement results displayed when the measurement parameter is DC voltage measurement/DC current measurement (Vdc-Idc). Keysight E4980A/AL Precision LCR Meter...
  • Page 76: Bin No. Display Page

    Test Signal Level (LEVEL) • DC Bias (BIAS) • Measurement Time Mode (MEAS TIME) • Correction Information (CORR) Figure 3-9 shows the available fields and the softkeys that correspond to each field on this page. Keysight E4980A/AL Precision LCR Meter...
  • Page 77: Comparator Function On/Off

    ON or OFF. Procedure for setting the comparator function Step 1. Press the [Display Format] key. Step 2. Press the BIN No. softkey. Step 3. Use the cursor keys to select the COMP field. Keysight E4980A/AL Precision LCR Meter...
  • Page 78 Display Format BIN NO. DISPLAY Page Step 4. Use the softkeys to set the comparator function to ON or OFF. Softkey Function Sets the comparator to ON Sets the comparator to OFF Keysight E4980A/AL Precision LCR Meter...
  • Page 79: Bin Count Display Page

    You can set it so that a beep sounds when the sorting result is output. For “Beep Feature” on page 143. details, refer to Keysight E4980A/AL Precision LCR Meter...
  • Page 80 Display Format BIN COUNT DISPLAY Page Figure 3-10 BIN COUNT DISPLAY Page Keysight E4980A/AL Precision LCR Meter...
  • Page 81: Counter Function

    Step 2. Press the BIN COUNT softkey. Step 3. Use the cursor keys to select the COUNT field. Step 4. Use the softkeys to press RESET COUNT. The counter is reset when the instrument setting state is recalled. Keysight E4980A/AL Precision LCR Meter...
  • Page 82: List Sweep Display Page

    You can set it so that a beep sounds when the sorting result is output. For “Beep Feature” on page 143. details, refer to Keysight E4980A/AL Precision LCR Meter...
  • Page 83 Display Format LIST SWEEP DISPLAY Page Figure 3-11 LIST SWEEP DISPLAY Page Keysight E4980A/AL Precision LCR Meter...
  • Page 84: Sweep Mode

    Sequential mode and step mode When two or more sweep points are the same and are adjacent, the E4980A/AL measures all of the listed points, and then compares the measurement result to limits set for each sweep point. Keysight E4980A/AL Precision LCR Meter...
  • Page 85 Step 3. Use the cursor keys to select the No. of fields. Step 4. Use the softkeys to confirm the sweep points on each page. Softkey Function PREV PAGE Displays the previous page NEXT PAGE Displays the next page Keysight E4980A/AL Precision LCR Meter...
  • Page 86: Display Blank Page

    The LCD backlight cannot be turned off. Keys other than the DISPLAY NORMAL softkey are not available. Even when the screen is in the non-display state and the front panel keys are unavailable (UNLOCK state), the DISPLAY NORMAL key is available. Keysight E4980A/AL Precision LCR Meter...
  • Page 87: Configuring Measurement Conditions (Display And Function Related Settings)

    Keysight E4980A/AL Precision LCR Meter User’s Guide Configuring Measurement Conditions (Display and Function Related Settings) This chapter provides information on the various settings accessible through the MEAS SETUP page of the E4980A/AL. Initializing the Instrument The E4980A/AL can be initialized into one of the following four default states:...
  • Page 88: Meas Setup Page

    Reference value for deviation measurement mode B (REF B field) The following fields are accessible through both the MEAS SETUP and MEAS DISPLAY pages: • Measurement function (FUNC field) • Measurement range (RANGE field) • Measurement frequency (FREQ field) Keysight E4980A/AL Precision LCR Meter...
  • Page 89 The following sections gives descriptions on the fields that are only accessible through the MEAS SETUP. Figure 4-1 shows the fields available on this page along with the softkeys corresponding to them. Figure 4-1 MEAS SETUP page Keysight E4980A/AL Precision LCR Meter...
  • Page 90: Comment Line

    Step 5. Press the ADD CHAR softkey.Your selected single character appears in the input line area. Step 6. Repeat Step 4 Step 5 to input subsequent characters. Step 7. Press the ENTER softkey to enter your text in the USER COMMENT field. Keysight E4980A/AL Precision LCR Meter...
  • Page 91: Trigger Mode

    The E4980A/AL performs one cycle of measurement each time it receives a trigger command sent via GPIB/USB/LAN. E4980A/AL ignores any trigger that is input during a measurement cycle. Be sure to trigger the instrument when it is not in a measurement cycle. Keysight E4980A/AL Precision LCR Meter...
  • Page 92 Step 3. Select your desired trigger mode by pressing the appropriate softkey: Softkey Description Puts the instrument into internal trigger (INT) mode. Puts the instrument into manual trigger (MAN) mode. Puts the instrument into external trigger (EXT) mode. Puts the instrument into bus mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 93: Automatic Level Control

    1. If the DUT has a very high linearity, the automatic level control feature may turn inactive before it enters the 9th iteration of the feedback loop of the level measure- ment cycle and the level may change. Keysight E4980A/AL Precision LCR Meter...
  • Page 94 When test signal level is set close to 2 Vrms/20 mArms or 20 Vrms/100 mArms or less then 5 mVrms, the warning message “ALC unable to regulate” may appear. Keysight E4980A/AL Precision LCR Meter...
  • Page 95 Configuring Measurement Conditions (Display and Function Related Settings) MEAS SETUP page Figure 4-4 Working ranges of the automatic level control feature Keysight E4980A/AL Precision LCR Meter...
  • Page 96 Step 1. Press [Meas Setup]. Step 2. Using the cursor keys, select the ALC field. Step 3. Use the following softkeys: Softkey Description Turns ON the automatic level control feature. Turns OFF the automatic level control feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 97: Dc Bias Current Isolation

    Turn ON the DC bias current isolation feature. Turn OFF the DC bias current isolation feature. Step 4. Set the DCI measurement range. For more information, see “To set the DCI range:” on page 106. Keysight E4980A/AL Precision LCR Meter...
  • Page 98: Averaging Factor

    Increments the averaging factor in steps of 1. DECR - Decrements the averaging factor in steps of 1. DECR -- Decrements the averaging factor in steps of 1, 2, 4, 8, 16, 32, 64, 128, and 256. Keysight E4980A/AL Precision LCR Meter...
  • Page 99: Trigger Delay Time

    INCR ++ / DECR -- (s) 1 m, 2 m, 3 m, •••, 9 m 10 m, 20 m, 30 m, •••, 90 m 100 m, 200 m, 300 m, •••, 900 m 1, 2, 3, •••, 9 Keysight E4980A/AL Precision LCR Meter...
  • Page 100 1 s through 10 s 10 ms 10 s through 100 s 100 ms 100 s through 999 s For the relationship between the trigger delay time and step delay time, refer to Figure 4-5 on page 102. Keysight E4980A/AL Precision LCR Meter...
  • Page 101: Step Delay Time

    1 m, 2 m, 3 m, •••, 9 m 10 m, 20 m, 30 m, •••, 90 m 100 m, 200 m, 300 m, •••, 900 m 1, 2, 3, •••, 9 10, 20, 30, •••, 90 100, 200, 300, •••, 900, 999 Keysight E4980A/AL Precision LCR Meter...
  • Page 102 Rdc parameter such as Ls-Rdc, since the stimulus setting and measurement procedure is executed three times in one measurement, the step delay time takes three times as long as the setting value. Keysight E4980A/AL Precision LCR Meter...
  • Page 103: Dc Bias Voltage Monitor

    Step 2. Using the cursor keys, select the VDC MON field. Step 3. Use the following softkeys: Softkey Description Turn ON the DC bias voltage monitor feature. Turn OFF the DC bias voltage monitor feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 104: Dc Bias Current Monitor

    Step 2. Using the cursor keys, select the IDC MON field. Step 3. Use the following softkeys: Softkey Description Turn ON the DC bias current monitor feature. Turn OFF the DC bias current monitor feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 105: Dcr Range

    Sets the DCR range to HOLD (manual). INCR + Use this softkey to increment the measurement range. The DCR range changes to HOLD. DECR - Use this softkey to decrement the measurement range. The DCR range changes to HOLD. Keysight E4980A/AL Precision LCR Meter...
  • Page 106: Dci Range

    Sets the DCI range to HOLD (manual). INCR + Use this softkey to increment the measurement range. The DCI range changes to HOLD. DECR - Use this softkey to decrement the measurement range. The DCI range changes to HOLD. Keysight E4980A/AL Precision LCR Meter...
  • Page 107: Dc Source

    INCR ++ / INCR -- (V) 1 m, 2 m, 3 m, •••, 9 m 10 m, 20 m, 30 m, •••, 90 m 100 m, 200 m, 300 m, •••, 900 m 1, 2, 3, •••, 9 10 Keysight E4980A/AL Precision LCR Meter...
  • Page 108: Automatic Bias Polarity Control

    UNKNOWN terminal. When the DC bias feature is OFF and you set the automatic bias polarity control feature to AUTO, the automatic bias polarity control feature does not function. Figure 4-6 Varactor diode (normal polarity) Keysight E4980A/AL Precision LCR Meter...
  • Page 109 Step 2. Using the cursor keys, select the BIAS POL field. Step 3. Use the following softkeys: Softkey Description AUTO Sets the automatic bias polarity control feature to AUTO mode. Sets the automatic bias polarity control feature to FIX mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 110: Deviation Measurement

    % = (X-Y)/Y100 (%) actual measurement of the DUT stored reference value To set up the deviation measurement feature: Step 1. Press [Meas Setup]. Step 2. Using the cursor keys, select the REF A field. Keysight E4980A/AL Precision LCR Meter...
  • Page 111 Step 7. Select the deviation mode for the primary parameter using the softkeys described in Step 1. If the primary parameter of the measurement function is C, the softkey labels show these units: p, n, u, m, x1. Keysight E4980A/AL Precision LCR Meter...
  • Page 112: Correction Page

    (REF A, REF B, OPEN A, OPEN B, SHORT A, SHORT B, LOAD A, and LOAD B fields) Figure 4-8 shows the fields available on this page along with the softkeys corresponding to them. Keysight E4980A/AL Precision LCR Meter...
  • Page 113: To Set The Correction Function To On Or Off

    Open/short /load correction at user-specified spot frequency points takes priority over open/short correction at all frequency points when test frequency is identical with one of the user-specified spot frequency points. Keysight E4980A/AL Precision LCR Meter...
  • Page 114: The Correction Functions Of The E4980A/Al Are Operated As Follows

    Open/short/load correction at at user-specified spot Performed user-specified spot frequency points is frequency points performed in place of open/short correction at all frequency points when test frequency is identical with one of the user-specified spot frequency points. Keysight E4980A/AL Precision LCR Meter...
  • Page 115: Open Correction

    300 kHz 80 Hz 400 Hz 4 kHz 40 kHz 400 kHz 500 Hz 5 kHz 50 kHz 500 kHz 600 Hz 6 kHz 60 kHz 600 kHz 800 Hz 8 kHz 80 kHz 800 kHz Keysight E4980A/AL Precision LCR Meter...
  • Page 116 • When the measurement has finished, the “OPEN measurement in progress” message disappears. During the measurement, the ABORT softkey is shown. Use this key when • you want to abort open correction. Keysight E4980A/AL Precision LCR Meter...
  • Page 117 If one of them happens, try its suggestion. "ALC unable to regulate" warning message is displayed (suggestion: turn off ALC) "Signal source overload" warning message is displayed (suggestion: reduce test signal) "Overload" is displayed (Measurement result becomes overload) (suggestion: check cabling/fixturing) Keysight E4980A/AL Precision LCR Meter...
  • Page 118 Step 6. Press the ON key to enable open correction in successive measurements. Description of Softkeys To enable/disable or otherwise control the behavior of open correction, use the following softkeys: Softkey Description Enables open correction. Disables open correction. MEAS OPEN Starts open correction. Keysight E4980A/AL Precision LCR Meter...
  • Page 119: Short Correction

    (inductances and resistances) at the 51 test frequency points. • During the measurement, a “SHORT measurement in progress” message is shown on the display. • When the measurement has finished, the “SHORT measurement in progress” message disappears. Keysight E4980A/AL Precision LCR Meter...
  • Page 120 Step 6. Press the ON key to enable short correction in successive measurements. Description of the Softkeys To enable/disable or otherwise control the behavior of short correction, use the following softkeys: Softkey Description Enables short correction. Disables short correction. MEAS Starts short correction. SHORT Keysight E4980A/AL Precision LCR Meter...
  • Page 121: Correction Based On User-Specified Frequency Points

    Step 3. Using the cursor keys, select the SPOT No. field. Step 4. Enter the measurement point number using the softkeys or entry keys. If you have used the entry keys to enter the value, the softkey labels change to unit labels (x1). Keysight E4980A/AL Precision LCR Meter...
  • Page 122 Step 9. Press the MEAS OPEN softkey. • During the measurement, an “OPEN measurement in progress” message is shown on the display. • When the measurement has finished, the “OPEN measurement in progress” message disappears. Keysight E4980A/AL Precision LCR Meter...
  • Page 123 Step 3 Step 9 until you have specified all of your desired measurement points. Step 11. Move the cursor into the OPEN field. Step 12. Press the ON key to enable the open correction feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 124 Step 9. Press the MEAS SHORT softkey. • During the measurement, a “SHORT measurement in progress” message is shown on the display. • When the measurement has finished, the “SHORT measurement in progress” message disappears. Keysight E4980A/AL Precision LCR Meter...
  • Page 125 Step 3 Step 9 until you have specified all of your desired measurement points. Step 11. Move the cursor into the SHORT field. Step 12. Press the ON key to enable the short correction feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 126 Step 14. Press the MEAS LOAD softkey. • During the measurement, a “LOAD measurement in progress” message is shown on the display. • When the measurement has finished, the “LOAD measurement in progress” message disappears. Keysight E4980A/AL Precision LCR Meter...
  • Page 127 Step 3 Step 14 until you have specified all of your desired measurement points. Step 16. Move the cursor into the LOAD field. Step 17. Press the ON key to enable the load correction feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 128: Relationships Between Correction Based On All Frequency Points And Correction Based On Specified Frequency Points

    Which data is used depends on the test frequencies. • Obtain the open/short correction data. 2. This area is used to: • Obtain the open/short/load correction data at each frequency point specified in the FREQ field. Keysight E4980A/AL Precision LCR Meter...
  • Page 129 When you specify a frequency point whose frequency value is identical to that of an existing frequency point and turn on the correction feature for both points, the frequency point with the smaller measurement point number (SPOT No. field) is selected as the correction data. Keysight E4980A/AL Precision LCR Meter...
  • Page 130: Reading/Writing Correction Data

    B, SHORT B, and LOAD B. Step 6. Enter the correction data using the softkeys or entry keys. If you use the entry keys to enter the value, the softkey labels change to unit labels (p, n, u, m, x1). Keysight E4980A/AL Precision LCR Meter...
  • Page 131: Measurement Functions For The Standard

    Step 5. If there is a secondary parameter and you are presented with another list of measurement parameters, select one of them by using the softkeys. For more information on measurement parameters, see “Measurement Function” on page Keysight E4980A/AL Precision LCR Meter...
  • Page 132: Selecting Single/Multiple Correction Mode

    For more information on measurement parameters, see “Measurement Function” on page When Option 301 (Scanner Interface) is not installed and you attempt to switch to multiple correction mode, a Scanner I/F not installed error message appears in the system message area. Keysight E4980A/AL Precision LCR Meter...
  • Page 133: Selecting The Cable Length

    Step 4. Use the following softkeys: Softkey Description Sets the cable length to 0 m (none). Sets the cable length to 1 m. Sets the cable length to 2 m. Sets the cable length to 4 m. Keysight E4980A/AL Precision LCR Meter...
  • Page 134: Limit Table Setup Page

    Beep feature (BEEP field) • Low limit for each bin (LOW field) • High limit for each bin (HIGH field) Figure 4-14 shows the fields available on this page along with the softkeys corresponding to them. Keysight E4980A/AL Precision LCR Meter...
  • Page 135: Parameter Swap Feature

    Cp-D and you use the parameter swap feature, the measurement function is changed to D-Cp (see Figure 4-15); after the swap, the comparator uses up to nine sets of limits for parameter D and one set of limits for parameter Cp. Keysight E4980A/AL Precision LCR Meter...
  • Page 136 You can revert to the previous setting of primary and secondary parameters by pressing the SWAP PARAM softkey again. Using the parameter swap feature does not change the primary and secondary parameters for the measurement function (on the MEAS DISPLAY and MEAS SETUP pages). Keysight E4980A/AL Precision LCR Meter...
  • Page 137: Comparator Limit Mode

    In tolerance mode, the lower limit need not be smaller than the nominal value, and the upper limit value need not be larger than the nominal value. As you can see in Figure 4-17, as sorting occurs, there may be blanks or overlaps. Keysight E4980A/AL Precision LCR Meter...
  • Page 138 Step 4. Select your desired limit mode by pressing the appropriate softkey: Softkey Description Switches the comparator into tolerance mode based on the absolute parameter values. Switches the comparator into tolerance mode based on the deviation percentages. Switches the comparator into sequential mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 139: Tolerance Mode Nominal Value

    Increments the nominal value in steps of 1. DECR - Decrements the nominal value in steps of 1. DECR -- Decrements the nominal value in steps of your selected number (1, 2, 5, 10, 20, 50, 100, 200, 500). Keysight E4980A/AL Precision LCR Meter...
  • Page 140: Turning On/Off The Comparator

    Step 2. Press the LIMIT TABLE softkey. Step 3. Using the cursor keys, select the COMP field. Step 4. Turn on or off the comparator by pressing the appropriate softkey: Softkey Description Turns ON the comparator. Turns OFF the comparator. Keysight E4980A/AL Precision LCR Meter...
  • Page 141: Turning On/Off The Auxiliary Bin

    DUTs that do not fall within the primary parameter limit values are sorted as OUT OF BINS. In addition, DUTs that fall within the primary parameter limits but are out of the secondary parameter limits are sorted into the auxiliary (AUX) bin. Keysight E4980A/AL Precision LCR Meter...
  • Page 142 Step 3. Using the cursor keys, select the AUX field. Step 4. Turn on or off the auxiliary (AUX) bin by pressing the appropriate softkey: Softkey Description Turns ON the auxiliary (AUX) bin. Turns OFF the auxiliary (AUX) bin. Keysight E4980A/AL Precision LCR Meter...
  • Page 143: Beep Feature

    Step 4. Set the beep output conditions by pressing the appropriate softkey: Softkey Description FAIL Beep is output when the comparator sorting result is OUT OF BIN. PASS Beep is output when the comparator sorting result is one of BIN1 through BIN9 or AUX BIN. Keysight E4980A/AL Precision LCR Meter...
  • Page 144: Lower And Upper Limits

    In sequential mode, you can omit the lower or upper limit or both for bin 1, as shown in the following examples: • Bin 1: lower limit (L1) only Bin 2: upper limit (H2) Bin 3: upper limit (H3) • Bin 1: upper limit (H1) only Bin 2: upper limit (H2) Keysight E4980A/AL Precision LCR Meter...
  • Page 145 9. 1. If the primary parameter of the measurement function is C, the softkey labels show these units: p, n, u, m, x1. Keysight E4980A/AL Precision LCR Meter...
  • Page 146 To clear the entire table: Step 1. Press [Meas Setup]. Step 2. Press the LIMIT TABLE softkey. Step 3. Using the cursor keys, select the BIN field. Step 4. Press the CLEAR TABLE softkey. Keysight E4980A/AL Precision LCR Meter...
  • Page 147: List Sweep Setup Page

    Limit parameter selection (LMT field) • Lower and upper limits (LOW and HIGH fields) Figure 4-20 shows the fields available on this page along with the softkeys corresponding to them. Figure 4-20 LIST SWEEP SETUP Page Keysight E4980A/AL Precision LCR Meter...
  • Page 148: Sweep Mode

    Step 4. Select your desired list sweep measurement mode by pressing the appropriate softkey: Softkey Description Set the list sweep measurement mode to sequential mode. STEP Set the list sweep measurement mode to step mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 149: List Sweep Parameters

    When the sweep parameter is set using the SCPI commands, the secondary sweep parameter can be set. In addition, the measurement range can be set for the secondary parameter. For details, refer to “:LIST:STIMulus:TYPE” on page 362. Keysight E4980A/AL Precision LCR Meter...
  • Page 150: Sweep Points And Limit Modes

    HIGH. If you specify the limit parameter for a sweep point but omit both lower and upper limit values, DUTs always pass the sweep point evaluated as IN. Keysight E4980A/AL Precision LCR Meter...
  • Page 151 (u, m, x1, k, M). Step 6. Using the cursor keys, select the LMT field. Step 7. Configure the limit parameter by pressing the appropriate softkey: Softkey Description Uses the measurement function’s primary parameter as the limit parameter. Keysight E4980A/AL Precision LCR Meter...
  • Page 152 -10 - 10 0.1 - 100 k 0.1 - 100 k 1 - 100 k 1 - 100 k IMP RNG 1. An error message “DC bias opt not installed” appears. 2. GPIB command is available. Keysight E4980A/AL Precision LCR Meter...
  • Page 153: Sweep Parameter Auto-Completion

    Step 4. Specify the start point by selecting the corresponding sweep point field (n) with the cursor keys. Step 5. Specify the end point by selecting the corresponding sweep point field (m) with the cursor keys. Keysight E4980A/AL Precision LCR Meter...
  • Page 154 1. Available only when using the frequency as the list sweep parameter. If n is not found, calculation is made on the assumption that n=1 represents the minimum value. If m is not found, calculation is made on the assumption that m=201 represents the maximum value. Keysight E4980A/AL Precision LCR Meter...
  • Page 155: System Configurations

    Keysight E4980A/AL Precision LCR Meter User’s Guide System Configurations This chapter provides information on the various settings accessible through the SYSTEM MENU page of the E4980A/AL. SYSTEM INFO Page Press the [System] key to open the SYSTEM INFO page. This page displays the...
  • Page 156: Handler Interface

    When the E4980A/AL is not equipped with the Handler Interface, the HANDLER I/F field shows a “NOT INSTALLED” message and the ON/OFF settings are not available. Softkey Description Turns ON the handler interface feature. Turns OFF the handler interface feature. Keysight E4980A/AL Precision LCR Meter...
  • Page 157: Scanner Interface

    FW DATE Displays the date of the instrument’s most recent update. OPTION Displays the option number of any option installed in the instrument. MAC ADDR Displays the MAC address. USB ID Displays the USB address. Keysight E4980A/AL Precision LCR Meter...
  • Page 158: System Config Page

    In addition, this page displays the IP address, subnet mask, and gateway settings obtained by setting the IP address to AUTO. Figure 5-2 shows the fields available on this page along with their corresponding softkeys. Figure 5-2 SYSTEM CONFIG Page Keysight E4980A/AL Precision LCR Meter...
  • Page 159: Turning On/Off The Beep Feature

    Turns ON the beep feature. Turns OFF the beep feature. 1. You can use the BEEP field on the LIST SWEEP SETUP page or the COMP:BEEP command to turn on/off the beep feature for the comparator. Keysight E4980A/AL Precision LCR Meter...
  • Page 160: Changing The Beep Tone

    Step 4. Change the beep tone by pressing the appropriate softkey: Softkey Description TONE 1 Selects tone 1. TONE 2 Selects tone 2. TONE 3 Selects tone 3. TONE 4 Selects tone 4. TONE 5 Selects tone 5. Keysight E4980A/AL Precision LCR Meter...
  • Page 161: Changing The Beep Tone

    Step 4. Use the following softkeys: Softkey Description DATE Selects the date field so that you can change the year, month, and day. TIME Selects the time field so that you can change the hour, minute, and second digits. Keysight E4980A/AL Precision LCR Meter...
  • Page 162 (x1). Softkey Description HOUR INCR + Increments the hours in steps of 1. HOUR DECR - Decrements the hours in steps of 1. Keysight E4980A/AL Precision LCR Meter...
  • Page 163: Configuring The Gpib Address

    Step 2. Press the SYSTEM CONFIG softkey. Step 3. Use the cursor keys to select the GPIB ADDR field. Step 4. Enter a value with the entry keys in the range of 0 through 30. Keysight E4980A/AL Precision LCR Meter...
  • Page 164: Configuring The Lan Ip Address

    Once you have automatically obtained an IP address, you can check the following monitor areas in the SYSTEM CONFIG page to see the address, subnet mask, and gateway: • CURRENT IP ADDR • CURRENT SUBNET MASK • CURRENT GATEWAY Keysight E4980A/AL Precision LCR Meter...
  • Page 165 E4980A/AL. The status is expressed as one of the following: Table 5-2 LAN connection status Status Description NORMAL LAN connection is OK. FAULT Disconnected from the LAN or LAN connection has failed. LAN connection is being initialized. Keysight E4980A/AL Precision LCR Meter...
  • Page 166: Self Test Page

    (denoted in parentheses). • Choosing a Test Item (TEST No. field) Figure 5-3 shows the fields available on this page along with their corresponding softkeys. Figure 5-3 SELF TEST Page Keysight E4980A/AL Precision LCR Meter...
  • Page 167: Choosing A Test Item

    SCANNER Checks the scanner interface. (Visual confirmation only; no INTERFACE pass/fail result shown on screen.) To run a self test: For step-by-step instructions, see “Daily Checks (Executing the self-test)” on page 421. Keysight E4980A/AL Precision LCR Meter...
  • Page 168: Service Page

    USB memory, but it does NOT allow you to modify or delete the information. The system information of the E4980A/AL may be used for support and repair by Keysight Technologies but will never be used for any other purpose. Figure 5-4...
  • Page 169: Monitor Information

    Saving the System Information into External Memory The system information of the E4980A/AL can be saved into external memory. The system information may be useful for support and repair by Keysight Technologies. For details on the system information, refer to the Service Guide.
  • Page 170 System Configurations SERVICE Page Keysight E4980A/AL Precision LCR Meter...
  • Page 171: Save/Recall

    Keysight E4980A/AL Precision LCR Meter User’s Guide Save/Recall This chapter provides information on the save/recall functionality of the E4980A/AL. Overview of Save/Recall Functionality Configurations and measurement results can be saved into, and recalled from, the E4980A/AL’s internal memory or external USB memory through the save/recall functionality.
  • Page 172: Folder/File Structure On Usb Memory

    Contains instrument configuration states. This single file, whose name is always system “system,” contains the system information. 1. This folder can be operated from the SYSTEM page. These folders are automatically created in the memory. Keysight E4980A/AL Precision LCR Meter...
  • Page 173: Usb Memory Notes

    If you cannot save into or recall from the USB memory, use another USB memory device. • Keysight Technologies shall not be responsible for nor assume any liability for data loss in your USB memory device after using it with the E4980A/AL. Keysight E4980A/AL Precision LCR Meter...
  • Page 174: Saving/Recalling Instrument Configuration States

    11 through 19 In this page, you can configure each of the following controls with the cursor placed in the corresponding field (denoted in parentheses). • Medium mode (MEDIA field) • Register number (No. field) Keysight E4980A/AL Precision LCR Meter...
  • Page 175: Medium Mode

    (in the No. field) change to 0 through 9. Uses USB memory as the destination or source. Once you have selected this mode, the register numbers (in the No. field) change to 10 through 19. Keysight E4980A/AL Precision LCR Meter...
  • Page 176: Choosing A Register Number

    E4980A/AL with a different option(s) equipped. 1. This value is not applicable when the medium mode is INT (internal memory). 2. An “Incompatible state file” warning message is displayed. Status information may fail to be correctly recalled. Keysight E4980A/AL Precision LCR Meter...
  • Page 177: Comment Information

    Step 2. Use the cursor keys to select the MEDIA field. Step 3. Press the INT softkey. Step 4. Use the cursor keys to select the No. field (0 through 9) for your desired register. Keysight E4980A/AL Precision LCR Meter...
  • Page 178 Hard key Description Recall A Recalls configuration states from register 0 (#0 in the No. field). Recall B Recalls configuration states from register 1 (#1 in the No. field). Keysight E4980A/AL Precision LCR Meter...
  • Page 179: Saving/Recalling Instrument Configuration States Into/From Usb Memory

    Step 2. Use the cursor keys to select the MEDIA field. Step 3. Press the EXT softkey. Step 4. Use the cursor keys to select the No. field (10 through 19) for your desired register. Keysight E4980A/AL Precision LCR Meter...
  • Page 180 DC bias and DC source settings are automatically turned OFF. If you attempt to recall configuration states under the following condition, a “No data to load” error message is displayed. • The checksum has changed. Keysight E4980A/AL Precision LCR Meter...
  • Page 181: Using The Auto Recall Feature

    Step 1. Plug the appropriate USB memory into the E4980A/AL. Step 2. Turn on the power to the E4980A/AL. Step 3. Upon start-up, the instrument configures itself using the instrument configuration information previously saved in register number 10 on the USB memory device. Keysight E4980A/AL Precision LCR Meter...
  • Page 182: Saving Measurement Results Into Usb Memory

    Outputs the measurement data for the primary parameter. The Data A field uses the following two different fixed-length ASCII formats: When the “effective digits” SN.NNNNNESNN setting is off: (S:+/-, N:0 to 9, E: exponent character) (:FORM:ASC:LONG OFF) Keysight E4980A/AL Precision LCR Meter...
  • Page 183 -1), but it is excluded when the data is saved into USB memory. Bin No. (IN/OUT) Represents the bin sorting results as well as IN/OUT evaluation results in list sweep measurement, as shown below: OUT_OF_BINS +1 to +9 BIN 1 through BIN 9 Keysight E4980A/AL Precision LCR Meter...
  • Page 184 The <BIN No.> data output format is either 2- or 3-character fixed-length ASCII format: SN or SNN (S:+/-, N:0 to 9) Example of measurement result output Example 6-1 Example of saved measurement result data +1.059517689E-24,+1.954963777E+00,+0,+0 +9.706803904E-25,+2.095857894E-01,+0,+0 +2.172725184E-24,+2.072965495E-01,+0,+0 +3.660460872E-25,+7.172688291E+00,+0,+0 +1.135428381E-24,+6.490636201E-01,+0,+0 +1.384790632E-24,+2.193020669E+00,+0,+0 +3.829879310E-26,+2.788435221E+01,+0,+0 Keysight E4980A/AL Precision LCR Meter...
  • Page 185: To Save Measurement Results Into Usb Memory

    Step 6. Press the SAVE & STOP softkey to save results into the USB memory. Step 7. When the data has been saved into the USB memory, a “Storing data completed. : E498xXXX.csv” message appears in the system message area. Keysight E4980A/AL Precision LCR Meter...
  • Page 186: How To Save The Measurement Result Of List Sweep Measurement To Usb Memory

    Step 7. Press soft key SAVE DATA. Step 8. Press soft key START LOG, then set the mode to accumulate the measurement result to the data buffer memory. Step 9. Press [Display Format]. Step 10. Press soft key LIST SWEEP. Keysight E4980A/AL Precision LCR Meter...
  • Page 187 For example, when measurement point of List Sweep is 100, 100 measurement results are saved to USB memory. Measurement result files are automatically assigned file names E498x001.csv through E498x999.csv, and you cannot change the file names. Keysight E4980A/AL Precision LCR Meter...
  • Page 188: Saving A Screenshot Into Usb Memory

    Step 5. When the data has been saved into the USB memory device, a “Storing image completed. : E498xXXX.gif” message appears in the system message area. Screenshot files are automatically assigned file names E498x001.gif through E498x999.gif, and you cannot change the file names. Keysight E4980A/AL Precision LCR Meter...
  • Page 189: Measurement Procedure And Examples

    Keysight E4980A/AL Precision LCR Meter User’s Guide Measurement Procedure and Examples This chapter covers basic measurement procedures as well as basic L, C, and R measurement theory. It also offers various measurement hints. After the descriptions of basic measurement procedures, practical measurement examples are shown using the E4980A/AL.
  • Page 190 Measurement Procedure and Examples Basic Measurement Procedure Figure 7-1 Basic Measurement Procedure Keysight E4980A/AL Precision LCR Meter...
  • Page 191: Impedance Parameters

    Phase as follows.      ----- - atan   Where, Impedance [] Resistance [] Reactance [] Absolute Value of Impedance []  Phase of Impedance [deg or rad] Series Resistance [] Keysight E4980A/AL Precision LCR Meter...
  • Page 192 Conductance and Susceptance, or in polar form as magnitude of Admittance and Phase. The following are expressions for Admittance. -- - Keysight E4980A/AL Precision LCR Meter...
  • Page 193 Phase of Admittance [deg or rad] Capacitance [F] Parallel Resistance [] measurement function of the E4980A/AL can obtain the |Y| and  The |Y|-  parameters given in the above equations. Figure 7-3 Vector Representation of Admittance Keysight E4980A/AL Precision LCR Meter...
  • Page 194: Parallel/Series Circuit Mode

    Cp-G Cp-Rp ------ - Cs mode Cs-D Cs-Q 2fCsRs --- - Cs-Rs Lp mode Lp-D Lp-Q -------------- - --- - Lp-G 2fLp Lp-Rp ------ - Ls mode Ls-D Ls-Q 2fLs -------------- - --- - Ls-Rs Keysight E4980A/AL Precision LCR Meter...
  • Page 195: Selecting Circuit Mode Of Capacitance

    Below approx. 10  Use series circuit mode Between above values Follow the manufacturer’s recommendation For example, to measure a 20 F capacitor at 1 kHz (impedance will be approx.8 ), the Cs-D or Cs-Q function is suitable. Keysight E4980A/AL Precision LCR Meter...
  • Page 196: Selecting Circuit Mode Of Inductance

    Use parallel circuit mode Between above values Follow the manufacturer’s recommendation For example, to measure a 1 mH inductor at 1 kHz (impedance will be approx. 6.3 ), the Ls-D or Ls-Q function is suitable. Keysight E4980A/AL Precision LCR Meter...
  • Page 197: Test Signal Level

    Source resistance of E4980A/AL (= 100 ) |Vm| Signal voltage level applied on DUT |Im| Signal current level flowed in DUT |Zx| Impedance of DUT Figure 7-6 Simplified model of test signal level and DUT Keysight E4980A/AL Precision LCR Meter...
  • Page 198: Test Signal Level Setting

    To make constant current level measurements, set an appropriate test signal level in the current value and set ALC to ON. The signal current level through the inductor will be constant. Keysight E4980A/AL Precision LCR Meter...
  • Page 199: Four-Terminal Pair Configuration

    1. The outer shield conductors work as the return path for the test signal current (they are not grounded). The magnetic fields produced by the inner and outer currents cancel each other because of the opposite directions Keysight E4980A/AL Precision LCR Meter...
  • Page 200 2. The vector voltmeter measures the differential voltage between the inner and outer conductors. The differential measurement method can minimize the influence of the mutual inductance. Keysight E4980A/AL Precision LCR Meter...
  • Page 201: Measurement Contacts

    In measuring capacitors of 10 pF or less, the stray capacitance (when the conductors are grounded, this is capacitance to ground) between the measurement contacts and the conductors near the capacitor influences the Figure 7-9. measurement as shown in Keysight E4980A/AL Precision LCR Meter...
  • Page 202 DUT, interconnect the shields of the measurement terminals to the conductor to reduce the influence of the stray capacitance to ground, as shown in Figure 7-10 (B). Figure 7-10 Reducing capacitance to ground Keysight E4980A/AL Precision LCR Meter...
  • Page 203: Contact Resistance

    When measuring large capacitance values, the four-terminal measurement contacts have the advantage of less measurement error as compared to the two-terminal method. Select a test fixture which can hold the DUT tightly to stabilize the connection. Figure 7-11 Contact resistance Keysight E4980A/AL Precision LCR Meter...
  • Page 204: Extending Test Leads

    Figure 7-13 to make the measurement contact. Figure 7-12 Extending the four-terminal pair test leads Keysight E4980A/AL Precision LCR Meter...
  • Page 205 Measurement Procedure and Examples Measurement Contacts Figure 7-13 Measurement contacts for test leads extension Keysight E4980A/AL Precision LCR Meter...
  • Page 206: Guarding For Measurement Of Low Capacitance Values

    Figure 7-14 shows an example of measurement contacts using a guard plate in the four-terminal pair measurement configuration. Figure 7-14 Example DUT Guard Plate Connection Keysight E4980A/AL Precision LCR Meter...
  • Page 207: Shielding

    Shielding minimizes the effects of electrical noise picked up by the test leads. Therefore, connect a shield plate to the outer shield conductors of the four-terminal pair test leads as shown in Figure 7-15. Figure 7-15 Guard shield Keysight E4980A/AL Precision LCR Meter...
  • Page 208: Correction Functions

    Compensate any error due to • Measurements to be Correction the test fixture and test leads referenced to a by using the standard. standard • Measurements using a test fixture that has complicated impedance characteristics Keysight E4980A/AL Precision LCR Meter...
  • Page 209 Measurement Procedure and Examples Correction Functions Table 7-3 Examples of correction functions • Simple measurements using a • Measurements using Keysight test leads Keysight-supplied direct-connecting test and a test fixture fixture • Precise measurements to be referenced • Measurements using a test fixture that has...
  • Page 210: Performing Open Correction

    UNKNOWN terminals. Figure 7-16 shows a sample shorting bar for the Keysight 16047E test fixture. The shorting bar should have very low residual impedance, so be sure to use a high-conductivity metal plate that does not easily corrode for the shorting plate.
  • Page 211 E4980A/AL: Do not use a test fixture. If the device does not have four-terminal pair measurement terminals, measure the device using a direct coupling test fixture (such as the Keysight 16047A/C/D/E). Step 3. On the CORRECTION page, enter the measured values obtained in step 2 as the REF A and REF B values with the function used in step 2.
  • Page 212: Parasitics Incident To Dut Connection

    Figure 7-17 shows parasitic impedance model after corrections are performed using the Keysight 16047A/D/E test fixture. In this case, to minimize the influence of parasitics on measurement values, insert the DUT completely into the test fixture while keeping the leads of the DUT as short as possible.
  • Page 213: Characteristics Example

    Table 7-4 Typical characteristics of components Characteristics Example Measurement Functions Large C Cs-Rs, Cs-D, Cs-Q, R-X, |Z|- Small C Cp-D, Cp-G, G-B, |Y|- Large L Lp-Rp, Lp-D Lp-Q, G-B, |Y|- Keysight E4980A/AL Precision LCR Meter...
  • Page 214 Measurement Procedure and Examples Characteristics Example Table 7-4 Typical characteristics of components Characteristics Example Measurement Functions Small L Ls-Rs, Ls-D, Ls-Q, R-X, |Z|- Large R Cp-Rp, G-B, |Y|- Small R Ls-Rs, R-X, |Z|- Keysight E4980A/AL Precision LCR Meter...
  • Page 215: Capacitor Measurements

    16047E Direct Couple Test Fixture (general purpose) is used for this measurement. Figure 7-18 Connecting the 16047E Step 4. To compensate for the Keysight 16047E’s residuals and strays, an OPEN/SHORT correction is required. 1. Leave the Keysight 16047E in an OPEN condition as shown in Figure 7-18.
  • Page 216 Wait until the message “OPEN measurement in progress” disappears. 4. Press the ON softkey to enable an OPEN correction. 5. Connect a shorting bar to the Keysight 16047E to set up the SHORT Figure 7-19. condition as shown in...
  • Page 217 Step 6. Press the [Meas Setup] key. Measurements are performed continuously by the internal trigger, and the measured Cp and D values of the capacitors are displayed as shown in Figure 7-21. Figure 7-21 Measurement results of Capacitor Keysight E4980A/AL Precision LCR Meter...
  • Page 218: Inductance Measurements

    Wait until the message “OPEN measurement in progress” disappears. 4. Press the ON softkey to enable an OPEN correction. 5. Connect a shorting bar to the Keysight 16047E to set up the SHORT condition as shown in Figure 7-19.
  • Page 219 (the value set in the LEVEL field). Step 7. Press the [Meas Setup] key. Measurements are performed continuously by the internal trigger, and the measured Ls and Rdc values of the magnetic core inductor are displayed. Keysight E4980A/AL Precision LCR Meter...
  • Page 220 Measurement Procedure and Examples Inductance Measurements Figure 7-23 Measurement results for magnetic core inductor Keysight E4980A/AL Precision LCR Meter...
  • Page 221: Measurements Using Dc Source

    Wait until the message “OPEN measurement in progress” disappears. 4. Press the ON softkey to enable an OPEN correction. 5. Connect a shorting bar to the Keysight 16047E to set up the SHORT condition as shown in Figure 7-19.
  • Page 222 1. Set the list sweep parameter to DC SRC[V]. 2. Enter the level of the sweep point. Step 8. Press the [DC BIAS] key. DCSRC is displayed in the status display area and the LED indicator of the DC source is ON (orange). Keysight E4980A/AL Precision LCR Meter...
  • Page 223 Measurement Procedure and Examples Measurements Using DC source Figure 7-25 Measurement results when using DC source Keysight E4980A/AL Precision LCR Meter...
  • Page 224 Measurement Procedure and Examples Measurements Using DC source Keysight E4980A/AL Precision LCR Meter...
  • Page 225: Overview Of Remote Control

    “USB Remote Control System” on page 240. You must install Keysight I/O Libraries Suite in the external controller in advance. Use Keysight I/O Libraries Suite 14 or higher. For further information on I/O Libraries Suite, see the Keysight I/O Libraries Suite manual.
  • Page 226: Gpib Remote Control System

    IEEE 488.1, IEC-625, IEEE 488.2, and JIS-C1901. The GPIB interface allows you to control the Keysight E4980A/AL from an external computer. The computer sends commands and instructions to the E4980A/AL and receives data sent from the E4980A/AL via the GPIB.
  • Page 227: Device Selector

    Use a personal computer or workstation equipped with the GPIB interface. You need to install software to control this instrument via the GPIB into the external controller (for example, HTBasic or Keysight VEE). 3. Other devices (other instruments and/or peripherals that serve your purpose) 4.
  • Page 228: Lan Remote Control System

    Required Equipment 1. E4980A/AL 2. External controller (personal computer or workstation that can be connected to a LAN and with Keysight I/O Libraries Suite installed) 3. Other devices (other instruments and/or peripherals that serve your purpose) 4. LAN cable for connecting the E4980A/AL with the external controller...
  • Page 229 Overview of Remote Control LAN remote control system Preparing the E4980A/AL Before controlling the E4980A/AL via a LAN, you need to configure the network Chapter 5, function. For detailed information on the procedure, refer to “System Configurations.”. Keysight E4980A/AL Precision LCR Meter...
  • Page 230: Control Over Sicl-Lan Server

    Windows environment. Step 1. From your PC’s Start menu, click Program - Keysight I/O Libraries Suite - Keysight Connection Expert to open the Keysight Connection Expert setting screen.
  • Page 231 OK. You can change the settings as necessary. For details, refer to the Keysight I/O Libraries Suite manual. Step 4. In the Keysight Connection Expert screen, check that the E4980A/AL has been added. Keysight E4980A/AL Precision LCR Meter...
  • Page 232 The following example shows how to control the E4980A/AL whose IP address is set to 192.168.1.101. When using Keysight VEE for PCs, use Keysight VEE Pro 6 for Windows or higher. Step 1. On the Keysight VEE’s I/O menu, click Instrument Manager..
  • Page 233 Name: SICL_LAN (you can specify any name), Interface: TCPIP, Board Number: 0, and TCPIP Address: TCPIP0::192.168.1.101::inst0::INSTR. Figure 8-3 shows an example of control using the direct I/O interface that has been set in the above procedure. Figure 8-3 Example of control using Keysight VEE Keysight E4980A/AL Precision LCR Meter...
  • Page 234: Control Over Telnet Server

    :FUNC:RANG:AUTO command to set the auto range setting to on, the :FUNC:IMP command to set a measurement item to Cp-D, and the *TRG command to trigger the instrument. The setting check is made with the query after each setting. Keysight E4980A/AL Precision LCR Meter...
  • Page 235 (WinSock1.1 and WinSock2.0), created by porting BSD Sockets to Windows and expanding it, is available. Control using Keysight VEE Keysight VEE allows you to control the E4980A/AL through the connection to Figure 8-5 the socket of port 5025 using the To/From Socket.
  • Page 236 Overview of Remote Control LAN remote control system Figure 8-5 Example of control using Keysight VEE Keysight E4980A/AL Precision LCR Meter...
  • Page 237: Control Via Web Server

    Step 2. Type http://10.116.199.134/ in the address field and press the enter key. Enter CURRENT IP ADDR on the E4980A/AL’s SYSTEM CONFIG page for an IP address. Step 3. The Web server’s start screen appears. Figure 8-6 Web Server Start Screen Keysight E4980A/AL Precision LCR Meter...
  • Page 238 Welcome Page Displays various setting information Control Instrument Virtual front panel and sending/reading SCPI commands Get Image Receives images Transfer Data Receives test results Configure LAN Displays and edit the LAN configuration Displays the Help page Keysight E4980A/AL Precision LCR Meter...
  • Page 239 Password of Configure LAN For the Web Server function, a password must be entered. The default password is “Keysight”. Changing the Password of the Configure LAN You can change the password from the Gear icon on the right upper corner.
  • Page 240: Usb Remote Control System

    Configuration of the USB Remote Control System Required Equipment 1. E4980A/AL (models with USB interface port (type mini-B)). 2. External controller (personal computer with Keysight I/O Libraries Suite and USB host port (type A)) installed. 3. Other USB-compatible devices (instruments and/or peripherals for specific purposes).
  • Page 241 USB cable to a target device, a dialog box will appear for USB device registration. The E4980A/AL will be identified as a new device if its serial number has been changed. Keysight E4980A/AL Precision LCR Meter...
  • Page 242 Changing Alias on Setting Screen The following are steps using Keysight I/O Libraries Suite 14. Step 1. From your PC’s Start menu, click Program - Keysight IO Libraries Suite -Keysight Connection Expert to open the setting screen. Step 2. In the setting screen, select the alias names from USB0 onward in the My Instruments frame, and then select the Create and edit alias from I/O Configuration on the menu bar.
  • Page 243 You can control the E4980A/AL by programming using SICL/VISA with Visual Studio in a Windows environment. For further information on controlling the E4980A/AL, see the manual for SICL or VISA. For Keysight I/O Libraries Suite, use Keysight I/O Libraries Suite 14.
  • Page 244 The following example shows how to control the E4980A/AL, for which alias is given as E4980_USBIF. When using Keysight VEE for PCs, use Keysight VEE Pro 7 for Windows or a later version. Step 1. On the Keysight VEE's I/O menu, click Instrument Manager...
  • Page 245 Overview of Remote Control USB Remote Control System Figure 8-10 Sample Control Using Keysight VEE (USB) Keysight E4980A/AL Precision LCR Meter...
  • Page 246: Sending Scpi Command Messages

    A semicolon does not change the current path but separates two commands in the same message. Figure 8-11 shows an example of how to use colons and semicolons to efficiently access commands in the command tree. Keysight E4980A/AL Precision LCR Meter...
  • Page 247: Grammar Of Messages

    HTBasic automatically sends the message terminator after the last data byte. Parameters A space (ASCII code: 32) is required between a command and its first parameter. When sending several parameters in a single command, separate each parameter with a comma (,). Keysight E4980A/AL Precision LCR Meter...
  • Page 248: Remote Mode

    Send the LOCAL BUS command from the controller to cancel this state, which enables control of the E4980A/AL on the front panel. Keysight E4980A/AL Precision LCR Meter...
  • Page 249: Trigger System

    “Measurement” and “Waiting for Trigger” states. • :ABOR • *RST • Other commands to change settings (:FREQ, etc.) When the DC bias function is turned on, the trigger system switches to the “Idle” state. Keysight E4980A/AL Precision LCR Meter...
  • Page 250 As shown in the table below, the instrument is triggered differently depending on which trigger source is specified. To specify the trigger source, use the :TRIG:SOUR command. Trigger Source How instrument is triggered Internal trigger The instrument is automatically triggered within itself. (Internal) Keysight E4980A/AL Precision LCR Meter...
  • Page 251 E4980A/AL is triggered by sending the :TRIG command. When reading measurement results via a controller in the “Idle” state, you need to use the :FETch? query command. 2. *TRG command or TRIGGER (GET) bus command Keysight E4980A/AL Precision LCR Meter...
  • Page 252 “Waiting for Trigger” state When the E4980A/AL is in remote mode, measurements can be made only on the screen displayed by pressing [Display Format]. A typical flowchart of data transfer using the trigger system is shown below. Keysight E4980A/AL Precision LCR Meter...
  • Page 253 Overview of Remote Control Trigger System Figure 8-14 Trigger system and data transfer (INT/MAN) Keysight E4980A/AL Precision LCR Meter...
  • Page 254 Trigger system and data transfer (EXT/BUS) When the E4980A/AL is set to the external trigger mode, and is triggered via the external trigger input terminal or an optional interface, this trigger signal has the same effect as the :TRIG command. Keysight E4980A/AL Precision LCR Meter...
  • Page 255: Data Buffer Memory

    ASSIGN @Adrs TO 717 REMOTE @Adrs OUTPUT @Adrs;"*RST;*CLS" OUTPUT @Adrs;"FORM ASCII" OUTPUT @Adrs;"TRIG:SOUR BUS" OUTPUT @Adrs;"DISP:PAGE MEAS" OUTPUT @Adrs;"MEM:DIM DBUF,5" OUTPUT @Adrs;"MEM:FILL DBUF" FOR I=1 TO 5 OUTPUT @Adrs;"TRIGGER:IMMEDIATE" NEXT I OUTPUT @Adrs;"MEM:READ? DBUF" ENTER @Adrs;D(*) Keysight E4980A/AL Precision LCR Meter...
  • Page 256: When Data Buffer Memory Is Used

    When the data buffer memory function is used during a list sweep measurement, the measurement result for one sweep point is stored as one set of measurement data. Therefore, when two or more sweep points are the Keysight E4980A/AL Precision LCR Meter...
  • Page 257 When the list sweep measurement’s limit function is set to off at a sweep point, IN/OUT is 0. Also, when the comparator function is set to off, the BIN No. is 0. Keysight E4980A/AL Precision LCR Meter...
  • Page 258: Output Format Of Data Buffer Memory

    The data format and meaning of <DATA A>, <DATA B>, <STATUS>, <BIN No.>, or <IN/OUT> are the same as those of the BINARY data format described in Data Format of “Data Transfer” on page 264 Keysight E4980A/AL Precision LCR Meter...
  • Page 259 Binary format (Buffer memory) NL^END is not included in the number of bytes to transfer. For example, the number of byte of “:MEM:DIM DBUF, 3”, is (8 x 4) x 3 + 1 = 97 bytes. Keysight E4980A/AL Precision LCR Meter...
  • Page 260: Starting A Measurement Cycle (Triggering The Instrument)

    Start measurement whenever you want while the E4980A/AL performs continuous measurement. Follow the steps below. Step 1. When the trigger system is operating (any state other than “Idle”), abort the trigger system using the :ABOR command. Keysight E4980A/AL Precision LCR Meter...
  • Page 261 When the E4980A/AL is triggered once, the device is measured at all sweep points. 2. Step mode The sweep point is incremented each time the E4980A/AL is triggered. Figure 8-19 Sequential mode and step mode Keysight E4980A/AL Precision LCR Meter...
  • Page 262: Waiting For The End Of Measurement

    Step 3. When an SRQ is generated, the program interrupts the measurement cycle. When the sweep averaging function is on, the E4980A/AL generates an SRQ when it has finished taking a number of measurements that is equal to the sweep averaging number. Keysight E4980A/AL Precision LCR Meter...
  • Page 263 Overview of Remote Control Waiting for the End of Measurement Figure 8-20 Sequence generating SRQ (end of measurement) Keysight E4980A/AL Precision LCR Meter...
  • Page 264: Data Transfer

    MEAS DISPLAY, BIN No., or BIN COUNT page is described below. Figure 8-21 ASCII Format 1 The <DATA A>, <DATA B>, <STATUS>, and <BIN No.> formats are as follows. • <DATA A> and <DATA B> format: Keysight E4980A/AL Precision LCR Meter...
  • Page 265 The <BIN No.> data is output with the measurement data only when the comparator function is set to on. The data output formats for <BIN No.> uses a two- or three-ASCII character data-length format as follows. SN or SNN (S:+/-, N: 0 to 9) Keysight E4980A/AL Precision LCR Meter...
  • Page 266 When the comparator function of the list sweep measurement is not used, the <IN/OUT> data output result is 0 (zero). The data output formats for <IN/OUT> use the two-ASCII character fixed-length format as follows. (S:+/-, N: 0 to 1) Keysight E4980A/AL Precision LCR Meter...
  • Page 267 Data formats of the <DATA A>, <DATA B>, <STATUS>, and <BIN No.> are common (8 bytes, IEEE754 floating point format) as follows. The meaning of each data item is the same as that in the ASCII format. Keysight E4980A/AL Precision LCR Meter...
  • Page 268 RN = 0 For example, S = 1 EXP = 0111111 1111 (1023) 1,000,000,000,000,000,000,000,000,000,000,000,000,000,000,000,000,000 (251) RN = (-1)1 x 2(1023 - 1023) x {1 + (251 / 252)} = -1 x 1 x 1.5 = -1.5 Keysight E4980A/AL Precision LCR Meter...
  • Page 269 <DATA A>, <DATA B>, <STATUS>, and <IN/OUT> are repeated as many times as there are sweep points. Each data format is the same as the 8-byte data format described previously. Figure 8-25 Binary Format 2 (List sweep) Keysight E4980A/AL Precision LCR Meter...
  • Page 270: Status Byte

    The value of each bit indicates the status of an internal function of the E4980A/AL, and two bits of the status byte are used as summary bits of the registers. Bits are set to “1” and reset to “0.” Figure 8-26 Status Byte Register Keysight E4980A/AL Precision LCR Meter...
  • Page 271 This bit is set to “1” when one or more enabled bits of the operation Summary status register is set to “1.” This bit is cleared when all bits of the operation status register are set to 0. (This bit is not cleared by serial polling.) Keysight E4980A/AL Precision LCR Meter...
  • Page 272: Enabling The Status Byte

    Bit 6 (RQS) is non-maskable, and bits 0 to 3 are always 0 (zero). Thus, it is meaningless to mask these bits. (The *SRE command’s bit pattern for masking bit 6 is ignored, while the *SRE command’s bit pattern for masking bits 0 to 6 is accepted, but is meaningless.) Keysight E4980A/AL Precision LCR Meter...
  • Page 273: Operation Status Register Structure

    The structure of the operation status register group is shown in Figure 8-28 below. This group consists of the following three registers. • Operation status condition register • Operation status event register • Operation status event enable register Figure 8-27 Operation Status Register Structure Keysight E4980A/AL Precision LCR Meter...
  • Page 274 0 (zero) unused always 0 (zero) List Sweep Measurement This bit is set to “1” when a last sweep point Complete Bit measurement of the list sweep measurement is completed. Keysight E4980A/AL Precision LCR Meter...
  • Page 275 The default setting is :STAT:OPER:ENAB 0 (all bits of the operation status event register are disabled). Bit 1, bit 2, and bits 5 to 15 are always 0 (zero). Thus, it is meaningless to mask them. Keysight E4980A/AL Precision LCR Meter...
  • Page 276: Standard Event Status Register

    -100 to -199 When you read the contents of the standard event status register using the *ESR? query command, the standard event status register is cleared, and bit 5 of the status byte is reset to “0.” Keysight E4980A/AL Precision LCR Meter...
  • Page 277 Overview of Remote Control Status Byte Figure 8-28 Standard Event Status Register Keysight E4980A/AL Precision LCR Meter...
  • Page 278 User Request (URQ) Bit always 0 (zero) Power On (PON) Bit This bit is set to “1” when the E4980A/AL’s power supply has been turned off and then on since the last time this register was read. Keysight E4980A/AL Precision LCR Meter...
  • Page 279: Enabling The Event Status Register

    The default setting is *ESE 0 (all bits of the standard event status register are disabled). Bits 1 and 6 of the event status register are always 0 (zero). Thus it is meaningless to mask these bits. Keysight E4980A/AL Precision LCR Meter...
  • Page 280 Overview of Remote Control Status Byte Keysight E4980A/AL Precision LCR Meter...
  • Page 281: Sample Program

    Keysight E4980A/AL Precision LCR Meter User’s Guide Sample Program This chapter describes sample programs of the E4980A/AL. The programs are written in VBA macro. Set Up Measurement Conditions and LCD Display This application program sets the measurement conditions and the LCD display of the E4980A/AL.
  • Page 282 ErrorCheck viVPrintf(AgtE4980A/AL, "DISP:PAGE MSET" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "DISP:LINE ""Control Example""" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "FUNC:IMP ZTD" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, "FREQ 1E6" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "VOLT 1" + vbLf, 0) Keysight E4980A/AL Precision LCR Meter...
  • Page 283 ErrorCheck viVPrintf(AgtE4980A/AL, "FUNC:DEV2:MODE ABS" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "FUNC:DEV1:REF 10000" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "FUNC:DEV2:REF 1" + vbLf, 0) Cells(12, 2).Value = "End to set configuration." '=================================== ' Setup End '=================================== Call viClose(defrm) '=================================== ' ErrorProc '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 284 MsgBox "*** Error : " + strVisaErr End If End Sub '=================================== ' SelectLCR '=================================== Sub selectLCR(defrm As Long, AgtE4980A/AL As Long) Dim Lcr As String Lcr = Worksheets("ControlPanel").Range("B3").Value If Lcr = "GPIB" Then ErrorCheck viOpen(defrm, "GPIB0::17::INSTR", 0, 0, AgtE4980A/AL) Keysight E4980A/AL Precision LCR Meter...
  • Page 285 Sample Program Set Up Measurement Conditions and LCD Display End If If Lcr = "USB" Then ErrorCheck viOpen(defrm, "USB0::2391::2313::MY12345678::0::INSTR", 0, 0, AgtE4980A/AL) End If End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 286: Detecting End Of Measurement

    Sets the trigger and status byte. STB Check Waits until the status byte becomes 192. Displays the status byte while waiting and again after processing. Setup End Closes the I/O bus. ErrorProc Displays an error number when an error occurs. Keysight E4980A/AL Precision LCR Meter...
  • Page 287 '=================================== ErrorCheck viOpenDefaultRM(defrm) Call selectLCR(defrm, AgtE4980A/AL) ErrorCheck viSetAttribute(AgtE4980A/AL, VI_ATTR_TMO_VALUE, TimeOutTime) '=================================== ' Setup Start '=================================== ErrorCheck viVPrintf(AgtE4980A/AL, "*RST;*CLS" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "*SRE 0" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "STAT:OPER:ENAB 0" + vbLf, 0) Keysight E4980A/AL Precision LCR Meter...
  • Page 288 ' STB Check '=================================== ErrorCheck viReadSTB(AgtE4980A/AL, StbStatus) Range("B12").Value = StbStatus Loop Until StbStatus = 192 ErrorCheck viVPrintf(AgtE4980A/AL, "*CLS" + vbLf, 0) ErrorCheck viReadSTB(AgtE4980A/AL, StbStatus) Range("D12").Value = StbStatus '=================================== ' Setup End '=================================== Call viClose(defrm) '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 289 Sample Program Detecting End of Measurement ' ErrorProc '=================================== ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 290: Read Measurement Results

    'Session to Default Resource Manager Dim AgtE4980A/AL As Long 'Session to instrument Dim Result As String * 500 Dim Res As Variant Dim Nop As Integer, i As Integer Const TimeOutTime = 30000 On Error GoTo ErrorHandler Keysight E4980A/AL Precision LCR Meter...
  • Page 291 ErrorCheck viVPrintf(AgtE4980A/AL, "COMP ON" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "INIT:CONT OFF" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, "INIT:IMM" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "TRIG:IMM" + vbLf, 0) '=================================== ' Meas Read '===================================` ErrorCheck viVPrintf(AgtE4980A/AL, "FETCh?" + vbLf, 0) ErrorCheck viVScanf(AgtE4980A/AL, "%t", Result) Keysight E4980A/AL Precision LCR Meter...
  • Page 292 Cells(13, i + 2).Value = Val(Res(i)) Next i '=================================== ' Setup End '===================================` Call viClose(defrm) '=================================== ' ErrorProc '===================================` ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 293: Read Measurement Data In Ascii Format (Data Buffer Memory)

    Dim Result As String * 500 Dim Res As Variant Dim NoofMeas As Integer, i As Integer, j As Integer, k As Integer Dim outEventType As Long, outEventContext As Long Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 294 ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:CLE DBUF" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:FILL DBUF" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, "APER LONG,5" + vbLf, 0) For i = 1 To NoofMeas ErrorCheck viVPrintf(AgtE4980A/AL, ":INIT:IMM;:TRIG:IMM" + vbLf, 0) Next i '=================================== ' Mem Buf Read '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 295 Next i ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:CLE DBUF" + vbLf, '=================================== ' Setup End '===================================` Call viClose(defrm) '=================================== ' ErrorProc '===================================` ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 296: Read Measurement Data In Ascii Format (List Sweep)

    Dim Result As String * 500 Dim Res As Variant Dim Nop As Long, i As Integer, j As Integer, k As Integer Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== ' Open Instrument '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 297 + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND4 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND5 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND6 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND7 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "INIT:CONT ON" + vbLf, Keysight E4980A/AL Precision LCR Meter...
  • Page 298 Cells(k + 39, j + 2).Value = Val(Res(i)) Next i '=================================== ' Setup End '=================================== Call viClose(defrm) '=================================== ' ErrorProc '=================================== ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 299: Read Measurement Data In Binary (Comparator)

    Dim AgtE4980A/AL As Long 'Session to instrument Dim Res() As Double Dim Nop As Long, i As Integer, j As Integer, k As Integer Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== ' Open Instrument Keysight E4980A/AL Precision LCR Meter...
  • Page 300 ErrorCheck viVPrintf(AgtE4980A/AL, "TRIG:IMM" + vbLf, 0) '=================================== ' Meas Read '=================================== ErrorCheck viVPrintf(AgtE4980A/AL, ":FETC?" + vbLf, 0) Call Scpi_read_binary_double_array(AgtE4980A/AL, Res, Nop) For i = 0 To 3 Cells(13, i + 9).Value = Res(i) Next i Keysight E4980A/AL Precision LCR Meter...
  • Page 301 Double, Nop As Long) Dim dblArray(10000) As Double Dim paramsArray(3) As Long Dim err As Long Dim i As Long Dim lf_eoi As String * 1 Nop = UBound(dblArray) - LBound(dblArray) + 1 paramsArray(0) = VarPtr(Nop) Keysight E4980A/AL Precision LCR Meter...
  • Page 302 Read Measurement Results paramsArray(1) = VarPtr(dblArray(0)) err = viVScanf(vi, "%#Zb%1t", paramsArray(0)) If err <> 0 Then MsgBox "Binary Error" ReDim data(Nop - 1) For i = 0 To Nop - 1 data(i) = dblArray(i) Next End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 303: Read Measurement Data In Binary Format (Data Buffer Memory)

    Dim Res() As Double Dim Nop As Long, i As Integer, j As Integer, k As Integer Dim outEventType As Long, outEventContext As Long Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== ' Open Instrument '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 304 ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:CLE DBUF" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:FILL DBUF" + vbLf, For i = 1 To NofMeas ErrorCheck viVPrintf(AgtE4980A/AL, ":INIT:IMM;:TRIG:IMM" + vbLf, 0) Next i '=================================== ' Mem Buf Read '=================================== ErrorCheck viVPrintf(AgtE4980A/AL, "MEM:READ? DBUF" + vbLf, 0) Keysight E4980A/AL Precision LCR Meter...
  • Page 305 Cells(k + 24, j + 9).Value = Res(i) Next i '=================================== ' Setup End '===================================` Call viClose(defrm) '=================================== ' ErrorProc '=================================== ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 306: Read Measurement Data In Binary Format (List Sweep)

    Dim Nop As Long, i As Integer, j As Integer, k As Integer Dim outEventType As Long, outEventContext As Long Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== ' Open Instrument '=================================== ErrorCheck viOpenDefaultRM(defrm) Keysight E4980A/AL Precision LCR Meter...
  • Page 307 + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND5 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND6 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "LIST:BAND7 A,100,200" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "INIT:CONT ON" + vbLf, ErrorCheck viVPrintf(AgtE4980A/AL, ":TRIG:IMM" + vbLf, 0) '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 308 Cells(k + 39, j + 9).Value = Res(i) Next i '=================================== ' Setup End '===================================` Call viClose(defrm) '=================================== ' ErrorProc '=================================== ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 309: Save/Recall

    'Session to Default Resource Manager Dim AgtE4980A/AL As Long 'Session to instrument Dim Result As String * 200 Const TimeOutTime = 30000 On Error GoTo ErrorHandler '=================================== ' Open Instrument '=================================== ErrorCheck viOpenDefaultRM(defrm) Call selectLCR(defrm, AgtE4980A/AL) Keysight E4980A/AL Precision LCR Meter...
  • Page 310 ErrorCheck viVQueryf(AgtE4980A/AL, "*OPC?" + vbLf, "%t", Result) If Cells(9, 2).Value = "Save" Then MsgBox ("State Save End") Else MsgBox ("State Recall End") End If Cells(15, 2).Value = Trim(Result) '=================================== ' Setup End '=================================== Call viClose(defrm) '=================================== ' ErrorProc Keysight E4980A/AL Precision LCR Meter...
  • Page 311 Sample Program Save/Recall '=================================== ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 312: Read/Write Correction Data At Specified Frequency Points

    Displays an error number when an error occurs. It is not possible to read/write the correction data in the correction based on all frequency points. Example 9-10 Read/Write correction data Sub Example_5() '=================================== ' Configuration '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 313 On Error GoTo ErrorHandler '=================================== ' Open Instrument '=================================== ErrorCheck viOpenDefaultRM(defrm) Call selectLCR(defrm, AgtE4980A/AL) ErrorCheck viSetAttribute(AgtE4980A/AL, VI_ATTR_TMO_VALUE, TimeOutTime) '=================================== ' Setup Start '=================================== ErrorCheck viVPrintf(AgtE4980A/AL, "*RST;*CLS" + vbLf, 0) ErrorCheck viVPrintf(AgtE4980A/AL, "DISP:PAGE CSET" + vbLf, 0) Keysight E4980A/AL Precision LCR Meter...
  • Page 314 For Spot_Lop = 1 To 3 ErrorCheck viVPrintf(AgtE4980A/AL, "CORR:SPOT" & Trim(Str(Spot_Lop)) & ":SHOR" + vbLf, 0) ErrorCheck viVQueryf(AgtE4980A/AL, "*OPC?" + vbLf, "%t", Result) Next Spot_Lop MsgBox ("Connect Load Standard.") For Spot_Lop = 1 To 3 Keysight E4980A/AL Precision LCR Meter...
  • Page 315 = i \ 6 Cells(k + 13, j + 2).Value = Res1(i) i = i + 1 Next Para Next Std Next FreqPoint MsgBox ("Reading correction data completed.") '=================================== ' Correction data uploading '=================================== SendCorr = "" Keysight E4980A/AL Precision LCR Meter...
  • Page 316 Next Std Next FreqPoint ErrorCheck viVPrintf(AgtE4980A/AL, "CORR:USE:DATA:SING " + SendCorr + vbLf, 0) ErrorCheck viVQueryf(AgtE4980A/AL, "*OPC?" + vbLf, "%t", Result) MsgBox ("Sending correction data completed.") '=================================== ' Setup End '=================================== Call viClose(defrm) '=================================== ' ErrorProc '=================================== Keysight E4980A/AL Precision LCR Meter...
  • Page 317 Sample Program Read/Write correction data at specified frequency points. ErrorHandler: ' Display the error message MsgBox "*** Error : " + Error$, MB_ICON_EXCLAMATION End Sub Keysight E4980A/AL Precision LCR Meter...
  • Page 318 Sample Program Read/Write correction data at specified frequency points. Keysight E4980A/AL Precision LCR Meter...
  • Page 319: Scpi Command Reference

    Keysight E4980A/AL Precision LCR Meter User’s Guide 10 SCPI Command Reference This chapter describes the SCPI command reference for the Keysight E4980A/AL. It describes the commands using their abbreviated format in alphabetical order. If you want to look up commands using their fully qualified format, refer to the index for the desired SCPI command.
  • Page 320: Notational Conventions In This Command Reference

    When a parameter is a value type or a string type enclosed with <>, its description, allowable setup range, preset (factory-set) value, and so on are given; when a parameter is a selection type enclosed by {}, the description of each selection item is given. Keysight E4980A/AL Precision LCR Meter...
  • Page 321: Equivalent Key

    [Key] - Item Indicates a series of key operations in which you press [Key], select the item called Item (softkey or field name) on the displayed menu using the cursor keys, and then press the softkey. Keysight E4980A/AL Precision LCR Meter...
  • Page 322: E4980A/Al Commands

    Reads out the value of the Standard Event Status Register. Executing this command clears the register value. Refer to the chapter on Remote Control. (Query Only) Equivalent key No equivalent key is available on the front panel. Keysight E4980A/AL Precision LCR Meter...
  • Page 323: Idn

    1 is read out at the completion of all pending operations. (Query Only) Equivalent key No equivalent key is available on the front panel. *OPT? Syntax *OPT? Description Returns the installed option numbers on the E4980A/AL. (Query Only) Equivalent key [System] - SYSTEM INFO *RST Syntax *RST Keysight E4980A/AL Precision LCR Meter...
  • Page 324: Sre

    *TRG is not a query command, but returns the value. An error will occur when sending the next command unless the value is read. *TST? Syntax *TST? Description Does nothing. The self-test is not executed by this command in the case of the E4980A/AL. Always returns 0. (Query Only) Keysight E4980A/AL Precision LCR Meter...
  • Page 325: Wai

    Enables the Automatic Level Control OFF or 0 (Preset value) Disables the Automatic Level Control Equivalent key [Meas Setup] - MEAS SETUP - ALC :APERture Syntax :APERture {SHORt|MEDium|LONG},<numeric> :APERture? Description Sets the measurement time mode and the averaging rate. Keysight E4980A/AL Precision LCR Meter...
  • Page 326: Bias:current[:Level]

    Parameter <Numeric> Range -100 m to 100 m (with option 001) Unit 1. Depending on signal level. Equivalent key [Display Format] - MEAS DISPLAY - BIAS [Meas Setup] > - MEAS SETUP - BIAS Keysight E4980A/AL Precision LCR Meter...
  • Page 327: Bias:polarity:auto

    If the automatic polarity control is turned OFF, the returned value is always the same the as one for :BIAS:VOLTage[:LEVel]. (Query Only) Equivalent key No equivalent key is available on the front panel. :BIAS:RANGe:AUTO Syntax :BIAS:RANGe:AUTO {ON|OFF|1|0} :BIAS:RANGe:AUTO? Keysight E4980A/AL Precision LCR Meter...
  • Page 328: Bias:state

    Sets the DC bias voltage. Setting does not implicitly turn the DC bias ON. If the bias is set by :BIAS:CURRent[:LEVel], the query command returns error No. 230. The parameter resolution depends on the setting voltage value. Keysight E4980A/AL Precision LCR Meter...
  • Page 329: Comparator:abin

    Selects the condition for generating a beep sound: when sorting with the comparator fails (sorts into OUT OF BIN) or passes (sorts into BINs 1 to 9 or AUX BIN). The beep should be turned on by the :SYSTem:BEEPer:STATe. Keysight E4980A/AL Precision LCR Meter...
  • Page 330: Comparator:bin:clear

    <Bin 1>,<Bin 2>,<Bin 3>,<Bin 4>,<Bin 5>,<Bin 6>,<Bin 7>,<Bin 8>,<Bin 9>,<Out of Bin>,<Aux Bin><NL^END> Equivalent key No equivalent key is available on the front panel. :COMParator:BIN:COUNt[:STATe] Syntax :COMParator:BIN:COUNt[:STATe] {ON|OFF|1|0} :COMParator:BIN:COUNt[:STATe]? Description Enables the BIN count function. Keysight E4980A/AL Precision LCR Meter...
  • Page 331: Comparator:mode

    Of course, the lower limit value must be lower than that of the upper limit. If a BIN is not used, this command returns the values of -9.9E37 and 9.9E37 for the lower and upper limits, respectively. Keysight E4980A/AL Precision LCR Meter...
  • Page 332: Comparator:slimit

    Preset value -9.9E+37, 9.9E+37 :FUNCtion:IMPedance[:TYPE] Unit Depends on Equivalent key [Meas Setup] - LIMIT TABLE - 2nd LOW [Meas Setup] - LIMIT TABLE - 2nd HIGH :COMParator[:STATe] Syntax :COMParator[:STATe] {ON|OFF|1|0} :COMParator[:STATe]? Description Enables the comparator function. Keysight E4980A/AL Precision LCR Meter...
  • Page 333: Comparator:swap

    When the lower limit or upper limit is not set, it returns -9.9E37 and 9.9E37, respectively. Parameter <low>,<high> Unit Depends on :FUNCtion:IMPedance[:TYPE] Equivalent key [Meas Setup] - LIMIT TABLE - LOW [Meas Setup] - LIMIT TABLE - HIGH Keysight E4980A/AL Precision LCR Meter...
  • Page 334: Comparator:tolerance:nominal

    Description Enables the current bias I/F. If the 42841A is connected, the preset status is ON. This is available only for serial prefix number below: MY464/SG464 and below for E4980A MY542/SG542 and below for E4980AL Parameter Description ON or 1...
  • Page 335: Control:scanner:state

    [System] - SYSTEM INFO - SCANNER I/F - OFF :CORRection:LENGth Syntax :CORRection:LENGth <numeric> :CORRection:LENGth? Description Selects the cable length correction setting. Parameter <Numeric> Range 0|1|2|4 Preset value Unit Equivalent key [Meas Setup] - CORRECTION - CABLE - 0 m Keysight E4980A/AL Precision LCR Meter...
  • Page 336: Correction:load:state

    Sets the type of LOAD correction reference to “Cp-Rp” Sets the type of LOAD correction reference to “Cs-D” Sets the type of LOAD correction reference to “Cs-Q” CSRS Sets the type of LOAD correction reference to “Cs-Rs” Keysight E4980A/AL Precision LCR Meter...
  • Page 337: Correction:method

    Description SINGle (Preset value) Sets the correction mode to “SINGLE” MULTiple Sets the correction mode to “MULTI” Equivalent key [Meas Setup] - CORRECTION - MODE - SINGLE [Meas Setup] - CORRECTION - MODE - MULTI Keysight E4980A/AL Precision LCR Meter...
  • Page 338: Correction:open[:Execute]

    Executes SHORT correction based on all frequency points. (No Query) Equivalent key [Meas Setup] - CORRECTION - SHORT - MEAS SHORT :CORRection:SHORt:STATe Syntax :CORRection:SHORt:STATe {ON|OFF|1|0} :CORRection:SHORt:STATe? Description Enables SHORT correction. Parameter Description ON or 1 Enables SHORT correction Keysight E4980A/AL Precision LCR Meter...
  • Page 339: Correction:spot[1-201]:Frequency

    :CORRection:SPOT[1-201]:LOAD:STANdard Syntax :CORRection:SPOT[1-201]:LOAD:STANdard <reference value for primary>, <reference value for secondary> :CORRection:SPOT[1-201]:LOAD:STANdard? Description Sets the standard reference values at the specified measurement point. The measurement function for the reference values can be selected using :CORRection:LOAD:TYPE. Keysight E4980A/AL Precision LCR Meter...
  • Page 340: Correction:spot[1-201]:Open[:Execute]

    ON or 1 Enables the specified measurement point OFF or 0 (Preset value) Disables the specified measurement point Equivalent key [Meas Setup] - CORRECTION - FREQ - ON [Meas Setup] - CORRECTION - FREQ - OFF Keysight E4980A/AL Precision LCR Meter...
  • Page 341: Correction:use[:Channel]

    Range 0 to 127 <open(n) A>, <open(n) B>, <short(n) A>, <short(n) B>, <load(n)A>, <load(n) B> Preset value Units OPEN: S; SHORT: ohms; LOAD: Depends on :CORRection:LOAD:TYPE Example of Use Dim Result As String * 50000 Keysight E4980A/AL Precision LCR Meter...
  • Page 342: Correction:use:data:single

    Dim Result As String * 50000 viVQueryf(AgtE4980A/AL, "CORR:USE:DATA:SING?" + vbLf, "%t", Result) Equivalent key [Meas Setup] - CORRECTION - MODE - SINGLE [Meas Setup] - CORRECTION - OPEN A [Meas Setup] - CORRECTION - OPEN B Keysight E4980A/AL Precision LCR Meter...
  • Page 343: Current[:Level]

    [Meas Setup] - MEAS SETUP - LEVEL :DISPlay:CCLear Syntax :DISPlay:CCLear Description Clears errors or caution messages from the display. (No Query) Equivalent key No equivalent key is available on the front panel. :DISPlay:ENABle Syntax :DISPlay:ENABle {ON|OFF|1|0} :DISPlay:ENABle? Description Enables display updates Keysight E4980A/AL Precision LCR Meter...
  • Page 344: Display:line

    [Meas Setup] - MEAS SETUP - USER COMMENT - PREV :DISPlay:PAGE Syntax :DISPlay:PAGE {MEASurement|BNUMber|BCOunt|LIST|MSETup|CSETup|LTABle|LSETup|CATAlo g|SYSTem|SELF|MLARge|SCONfig|SERVice} :DISPlay:PAGE? Description Selects the page to be displayed. Parameter Description MEASurement (Preset value) Sets displayed page to <MEAS DISPLAY> BNUMber Sets displayed page to <BIN No. DISPLAY> Keysight E4980A/AL Precision LCR Meter...
  • Page 345: Display[:Window]:Text[1-2][:Data]:Fmsd:data

    [Meas Setup] - CORRECTION [Meas Setup] - LIMIT TABLE [Meas Setup] - LIST SETUP [Save/Recall] [System] - SYSTEM INFO [System] - SYSTEM CONFIG [System] - SELF TEST [System] - SERVICE :DISPlay[:WINDow]:TEXT[1-2][:DATA]:FMSD:DATA Syntax :DISPlay[:WINDow]:TEXT[1-2][:DATA]:FMSD:DATA <most significant digit> :DISPlay[:WINDow]:TEXT[1-2][:DATA]:FMSD:DATA? Keysight E4980A/AL Precision LCR Meter...
  • Page 346: Display[:Window]:Text[1-2][:Data]:Fmsd[:State]

    [Display Format] - Fixed Decimal Point Menu - D.P. FIX :FETCh[:IMPedance]:CORRected? Syntax :FETCh[:IMPedance]:CORRected? Description Returns a complex measurement result (R-X format) after correction. (Query Only) Equivalent key No equivalent key is available on the front panel. :FETCh[:IMPedance][:FORMatted]? Syntax :FETCh[:IMPedance][:FORMatted]? Keysight E4980A/AL Precision LCR Meter...
  • Page 347: Fetch:smonitor:iac

    OFF, 0.0 is returned. (Query Only) Equivalent key No equivalent key is available on the front panel. :FORMat:ASCii:LONG Syntax :FORMat:ASCii:LONG {ON|OFF|1|0} :FORMat:ASCii:LONG? Description Enables the long format (+0.0000000000E+00). This is applied to the returned values of :FETCh[:IMPedance][:FORMatted]?/:MEMory:READ?/*TRG. Keysight E4980A/AL Precision LCR Meter...
  • Page 348: Format:border

    In the case of the binary mode, any value other than [,64] is acceptable but ignored. A query returns REAL, 64 in case of using the binary format. See the chapter on program examples for data transfer. Parameter Description ASCii (Preset value) Sets the transfer mode to ASCII Keysight E4980A/AL Precision LCR Meter...
  • Page 349: Format:exponent:digit

    Syntax :FREQuency[:CW] <numeric> :FREQuency[:CW]? Description Sets the frequency for normal measurement. Parameter <Numeric> Range 20 to 2M Preset value Unit Resolution Depends on the setting frequency. 1. The range varys depending on the frequency option. Keysight E4980A/AL Precision LCR Meter...
  • Page 350: Function:dcresistance:range:auto

    Selects the DCR measurement range. This command turns the auto range function OFF. Parameter <Numeric> Range 10|100|1k|10k|100k Preset value Unit ohms Equivalent key [Meas Setup] - MEAS SETUP - DCR RNG :FUNCtion:DEV[1-2]:MODE Syntax :FUNCtion:DEV[1-2]:MODE {ABSolute|PERCent|OFF} :FUNCtion:DEV[1-2]:MODE? Keysight E4980A/AL Precision LCR Meter...
  • Page 351: Function:dev[1-2]:Reference:fill

    [Meas Setup] - MEAS SETUP - REF A - MEASURE [Meas Setup] - MEAS SETUP - REF B - MEASURE :FUNCtion:DEV[1-2]:REFerence[:VALue] Syntax :FUNCtion:DEV[1-2]:REFerence[:VALue] <numeric> :FUNCtion:DEV[1-2]:REFerence[:VALue]? Description Sets the reference value for deviation measurement. Parameter <Numeric> Preset value Keysight E4980A/AL Precision LCR Meter...
  • Page 352: Function:impedance:range:auto

    [Meas Setup] - MEAS SETUP - RANGE - HOLD :FUNCtion:IMPedance:RANGe[:VALue] Syntax :FUNCtion:IMPedance:RANGe[:VALue] <numeric> :FUNCtion:IMPedance:RANGe[:VALue]? Description Selects the impedance measurement range. This command turns the auto range function OFF. Parameter <Numeric> Range 100m|1|10|100|300|1k|3k|10k|30k|100k Preset value Unit ohms Keysight E4980A/AL Precision LCR Meter...
  • Page 353: Function:impedance[:Type]

    Sets the Impedance Parameter Type to “Ls-Rs” Sets the Impedance Parameter Type to “Ls-Rdc” LSRD Sets the Impedance Parameter Type to “R-X” Sets the Impedance Parameter Type to “Z-thd” Sets the Impedance Parameter Type to “Z-thr” Keysight E4980A/AL Precision LCR Meter...
  • Page 354: Function:smonitor:iac[:State]

    OFF or 0 Does nothing Equivalent key No equivalent key is available on the front panel. :FUNCtion:SMONitor:IDC[:STATe] Syntax :FUNCtion:SMONitor:IDC[:STATe] {ON|OFF|1|0} :FUNCtion:SMONitor:IDC[:STATe]? Description Enables the DC current-level monitor function. This is available when option 001 is installed. Keysight E4980A/AL Precision LCR Meter...
  • Page 355: Function:smonitor:vac[:State]

    No equivalent key is available on the front panel. :FUNCtion:SMONitor:VDC[:STATe] Syntax :FUNCtion:SMONitor:VDC[:STATe] {ON|OFF|1|0} :FUNCtion:SMONitor:VDC[:STATe]? Description Enables the DC voltage-level monitor function. This is available when option 001 is installed. Parameter Description ON or 1 DC voltage monitor function ON Keysight E4980A/AL Precision LCR Meter...
  • Page 356: Hcopy:sdump:data

    Description Enables the automatic trigger to change state from the “Idle” state to the “Wait for Trigger” state. Refer to the chapter on Remote Control. Parameter Description ON or 1 Enables automatic trigger state change Keysight E4980A/AL Precision LCR Meter...
  • Page 357: Initiate[:Immediate]

    OFF. Unit Depends on :FUNCtion:IMPedance[:TYPE] Equivalent key [Meas Setup] - LIST SETUP - LMT - A [Meas Setup] - LIST SETUP - LMT - B [Meas Setup] - LIST SETUP - LMT -  Keysight E4980A/AL Precision LCR Meter...
  • Page 358: List:bias:current

    DC voltage, the query command returns error No. 230. The value 9.9E37 should be set for any empty points. Parameter <point n> Range of n 1 to 201 Unit Equivalent key [Meas Setup] - LIST SETUP - BIAS[V] :LIST:CLEar:ALL Syntax :LIST:CLEar:ALL Keysight E4980A/AL Precision LCR Meter...
  • Page 359: List:current

    No. 230. The value 9.9E37 should be set for any empty points. Parameter <point n> Range of n 1 to 201 Unit Equivalent key [Meas Setup] - LIST SETUP - DC SRC[V] Keysight E4980A/AL Precision LCR Meter...
  • Page 360: List:frequency

    Sets the list sweep mode to sequence mode STEPped Sets the list sweep mode to step mode Equivalent key [Meas Setup] - LIST SETUP - MODE - SEQ [Meas Setup] - LIST SETUP - MODE - STEP :LIST:SEQuence:TSTamp:CLEar Syntax :LIST:SEQuence:TSTamp:CLEar Keysight E4980A/AL Precision LCR Meter...
  • Page 361: List:sequence:tstamp:data

    Range of n 1 to 201 Unit Depends on :LIST:STIMulus:TYPE :LIST:STIMulus:TYPE 1. When the second parameter of command is set to NONE, this parameter is not needed. Example of Use viVPrintf(AgtE4980A/AL, "LIST:STIM:TYPE FREQ, VOLT" + vbLf, 0) Keysight E4980A/AL Precision LCR Meter...
  • Page 362: List:stimulus:mdata

    Second Parameter NONE (Preset value) Selects no second stimulus parameter. The same operation is performed as the :LIST:FREQuency, :LIST:CURRent so on according to the first stimulus parameter. IRANge Selects impedance range as the second stimulus parameter. Keysight E4980A/AL Precision LCR Meter...
  • Page 363: List:voltage

    <point n> Range of n 1 to 201 Unit Example of Use viVPrintf(AgtE4980A/AL, ":LIST:VOLT 0.5, 0.8, 1, 9.9e37, 1.5" + vbLf, 0) Equivalent key [Meas Setup] - LIST SETUP - LEVEL[V] :MEMory:CLEar Syntax :MEMory:CLEar DBUF Keysight E4980A/AL Precision LCR Meter...
  • Page 364: Memory:dim

    “9.90000E+37,+9.90000E+37,-1,+0” is returned the number of times specified by :MEMory:DIM. When the data buffer memory is not filled to the specified size (specified by the :MEMory:DIM command), “9.90000E+37,+9.90000E+37,-1,+0” remains for the rest of the data locations. (Query Only) Keysight E4980A/AL Precision LCR Meter...
  • Page 365: Mmemory:delete[:Register]

    Stores the state to the memories. Numbers 0 to 9 are located in the internal memory, while Nos. 10 to 19 are situated in the external USB memory. Number 10 is automatically recalled. (No Query) Keysight E4980A/AL Precision LCR Meter...
  • Page 366: Output:dc:isolation:level:auto

    Selects DC isolation range. This command turns the auto range function OFF. Parameter <Numeric> Range 20 |200 |2 m|20 m|100 m Preset value 20 m Unit Equivalent key [Meas Setup] - MEAS SETUP - DCI RNG Keysight E4980A/AL Precision LCR Meter...
  • Page 367: Output:dc:isolation[:State]

    Does nothing. This function is provided because of its command compatibility with 4284A. Parameter Description ON or 1 Does nothing OFF or 0 (Preset value) Does nothing Equivalent key No equivalent key is available on the front panel. :SOURce:DCSource:STATe Syntax :SOURce:DCSource:STATe {ON|OFF|1|0} :SOURce:DCSource:STATe? Keysight E4980A/AL Precision LCR Meter...
  • Page 368: Source:dcsource:voltage[:Level]

    Equivalent key [Meas Setup] - MEAS SETUP - DC SRC :STATus:OPERation:CONDition? Syntax :STATus:OPERation:CONDition? Description Returns the value of the Operation Status Condition register. (Query Only) Equivalent key No equivalent key is available on the front panel. Keysight E4980A/AL Precision LCR Meter...
  • Page 369: Status:operation:enable

    (No Query) Equivalent key No equivalent key is available on the front panel. :SYSTem:BEEPer:STATe Syntax :SYSTem:BEEPer:STATe {ON|OFF|1|0} :SYSTem:BEEPer:STATe? Description Enables a beep sound. Parameter Description ON or 1 Enables a beep sound Keysight E4980A/AL Precision LCR Meter...
  • Page 370: System:beeper:tone

    [System] - SYSTEM CONFIG - BEEPER TONE - TONE 4 [System] - SYSTEM CONFIG - BEEPER TONE - TONE 5 :SYSTem:COMMunicate:GPIB[:SELF]:ADDRess Syntax :SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric> :SYSTem:COMMunicate:GPIB[:SELF]:ADDRess? Description Sets the GPIB address. Parameter <Numeric> Range 0 to 30 Preset value Keysight E4980A/AL Precision LCR Meter...
  • Page 371: System:communicate:lan[:Self]:Address

    ON or 1 (Preset value) Enables automatic IP address setup OFF or 0 Disables automatic IP address setup Equivalent key [System] - SYSTEM CONFIG - AUTO-IP ENABLED - ON [System] - SYSTEM CONFIG - AUTO-IP ENABLED - OFF Keysight E4980A/AL Precision LCR Meter...
  • Page 372: System:communicate:lan[:Self]:Configure

    Returns the current IP address. (Query Only) Equivalent key No equivalent key is available on the front panel. :SYSTem:COMMunicate:LAN[:SELF]:CURRent:DGATeway? Syntax :SYSTem:COMMunicate:LAN[:SELF]:CURRent:DGATeway? Description Returns the current Gateway address. (Query Only) Equivalent key No equivalent key is available on the front panel. Keysight E4980A/AL Precision LCR Meter...
  • Page 373: System:communicate:lan[:Self]:Current:smask

    ON or 1 (Preset value) Enables the DHCP function OFF or 0 Disables the DHCP function Equivalent key [System] - SYSTEM CONFIG - DHCP ENABLED - ON [System] - SYSTEM CONFIG - DHCP ENABLED - OFF Keysight E4980A/AL Precision LCR Meter...
  • Page 374: System:communicate:lan[:Self]:Mac

    Restarts the network (No Query) This is available only for serial prefix number below: MY464/SG464 and below for E4980A MY542/SG542 and below for E4980AL Equivalent key [System] - SYSTEM CONFIG - DHCP ENABLED - RESTART NETWORK [System] - SYSTEM CONFIG - AUTO-IP ENABLED - RESTART NETWORK...
  • Page 375: System:date

    Unit days Resolution Equivalent key [System] - SYSTEM CONFIG - DATE/TIME - DATE - YEAR [System] - SYSTEM CONFIG - DATE/TIME - DATE - MONTH [System] - SYSTEM CONFIG - DATE/TIME - DATE - DAY Keysight E4980A/AL Precision LCR Meter...
  • Page 376: System:error[:Next]

    Technologies” and “Keysight Technologies”. A power cycle or reboot is required for the changes in the instrument’s manufacturer string to take effect and to be recognized in the Keysight Connection Expert. The string will remain for the subsequent power cycle or reboot.
  • Page 377: System:preset

    :SYSTem:RESTart Description Reboots the instrument immediately. (No Query) This is available only for serial prefix number below: MY464/SG464 and below for E4980A MY542/SG542 and below for E4980AL Equivalent key No equivalent key is available on the front panel. :SYSTem:TIME Syntax :SYSTem:TIME <hour>,<minute>,<second>...
  • Page 378: Trigger:delay

    100  Resolution Equivalent key [Meas Setup] - MEAS SETUP - STEP DLY :TRIGger[:IMMediate] Syntax :TRIGger[:IMMediate] Description Triggers the unit. (No Query) Equivalent key [Trigger] :TRIGger:SOURce Syntax :TRIGger:SOURce {INTernal|HOLD|EXTernal|BUS} :TRIGger:SOURce? Description Selects the trigger mode. Keysight E4980A/AL Precision LCR Meter...
  • Page 379: Trigger:tdel

    [Meas Setup] - TRIG SETUP - TRIG DLY :VOLTage[:LEVel] Syntax :VOLTage[:LEVel] <numeric> :VOLTage[:LEVel]? Description Sets the voltage level for the measurement signal. If the signal is set by :CURRent[:LEVel], the query command returns error No. 230. Keysight E4980A/AL Precision LCR Meter...
  • Page 380 Parameter <Numeric> Range 0 to 20 Preset value Unit 100  Resolution 1. Depending on the DC bias level. Equivalent key [Display Format] - MEAS DISPLAY - LEVEL [Meas Setup] - MEAS SETUP - LEVEL Keysight E4980A/AL Precision LCR Meter...
  • Page 381: Command List

    Comparator function On/Off :COMParator[:STATe] Swap parameter function On/Off :COMParator:SWAP BIN limit values setup for the :COMParator:TOLerance:BIN[1-9] tolerance mode Nominal value setup for the :COMParator:TOLerance:NOMinal tolerance mode Control Handler I/F enable :CONTrol:HANDler:STATe Scanner I/F enable :CONTrol:SCANner:STATe Keysight E4980A/AL Precision LCR Meter...
  • Page 382 SINGLE mode correction data :CORRection:USE:DATA:SINGle readout Current Current level setup :CURRent[:LEVel] Data buffer Data buffer memory clear and :MEMory:CLEar memory disable Data buffer memory size setup :MEMory:DIM Data buffer memory enable :MEMory:FILL Data buffer memory readout :MEMory:READ Keysight E4980A/AL Precision LCR Meter...
  • Page 383 :FETCh[:IMPedance]:CORRected readout Measurement result return with the :FETCh[:IMPedance][:FORMatted] selected measurement function AC current monitor data readout :FETCh:SMONitor:IAC DC current monitor data readout :FETCh:SMONitor:IDC AC voltage monitor data readout :FETCh:SMONitor:VAC DC voltage monitor data readout :FETCh:SMONitor:VDC Keysight E4980A/AL Precision LCR Meter...
  • Page 384 Measurement function selection :FUNCtion:IMPedance[:TYPE] Dummy command (for 4284A :FUNCtion:SMONitor:IAC[:STATe] compatibility) DC current level monitor enable :FUNCtion:SMONitor:IDC[:STATe] Dummy command (for 4284A :FUNCtion:SMONitor:VAC[:STATe] compatibility) DC voltage level monitor enable :FUNCtion:SMONitor:VDC[:STATe] Hcopy Screen image output to the :HCOPy:SDUMp:DATA controller Keysight E4980A/AL Precision LCR Meter...
  • Page 385 Instrument setting reset *RST Service Request Enable register *SRE setup Status Byte register readout *STB Trigger *TRG Dummy command (for 4284A *TST compatibility) Waits *WAI Initiate Auto trigger state initiate enable :INITiate:CONTinuous WAIT FOR TRIGGER initiate :INITiate[:IMMediate] Keysight E4980A/AL Precision LCR Meter...
  • Page 386 State delete from the memories :MMEMory:DELete[:REGister] State Load from the memories :MMEMory:LOAD:STATe[:REGister] State Store to the memories :MMEMory:STORe:STATe[:REGister] Status Operation Status Condition register :STATus:OPERation:CONDition register readout Operation Status Enable register :STATus:OPERation:ENABle setup Operation Status Event register :STATus:OPERation[:EVENt] return Keysight E4980A/AL Precision LCR Meter...
  • Page 387 :SYSTem:KLOCk Instrument setting and correction :SYSTem:PRESet data reset Clock date setup :SYSTem:TIME Trigger Step delay time setup :TRIGger:DELay Trigger :TRIGger[:IMMediate] Selects the trigger mode :TRIGger:SOURce Trigger delay time setup :TRIGger:TDEL Voltage Voltage level setup :VOLTage[:LEVel] Keysight E4980A/AL Precision LCR Meter...
  • Page 388: Command Tree

    [No Query] :ABORt [No Query] :AMPLitude :ALC {ON|OFF|1|0} :APERture {SHORt|MEDium|LONG },< numeric> :BIAS :CURRent [:LEVel] < numeric> :POLarity :AUTO {ON|OFF|1|0} :CURRent [:LEVel] < numeric> [Query only] :VOLTage [:LEVel] < numeric> [Query only] :STATe {ON|OFF|1|0} :VOLTage Keysight E4980A/AL Precision LCR Meter...
  • Page 389 {ON|OFF|1|0} :MODE {ATOLerance|PTOLerance|SEQuence} :SEQuence :BIN < array> :SLIMit < numeric>,< numeric> [:STATe] {ON|OFF|1|0} :SWAP {ON|OFF|1|0} :TOLerance :BIN[1-9] < numeric>,< numeric> :NOMinal < numeric> :CONTrol :HANDler :STATe {ON|OFF|1|0} :SCANner :STATe {ON|OFF|1|0} :CORRection :LENGth < numeric> :LOAD Keysight E4980A/AL Precision LCR Meter...
  • Page 390 [No Query] :STANdard < numeric>,< numeric> :OPEN [:EXECute] [No Query] :SHORt [:EXECute] [No Query] :STATe {ON|OFF|1|0} :USE [:CHANnel] < numeric> :DATA [:MULTi] < numeric>,< array> :SINGle < array> :CURRent [:LEVel] < numeric> :DISPlay :CCLear [No Query] Keysight E4980A/AL Precision LCR Meter...
  • Page 391 [Query only] :IDC < numeric> [Query only] :VAC < numeric> [Query only] :VDC < numeric> [Query only] :FORMat :ASCii :LONG {ON|OFF|1|0} :BORDer {NORMal|SWAPped} [:DATA] {ASCii|REAL },< numeric> :EXPonent:DIGit {2|3} :FREQuency [:CW] < numeric> :FUNCtion :DCResistance Keysight E4980A/AL Precision LCR Meter...
  • Page 392 :AUTO {ON|OFF|1|0} [:VALue] < numeric> [:TYPE] {CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS|LP D|LPQ|LPG|LPRP|LPRD|LSD|LSQ|LSRS|LS RD|RX|ZTD|ZTR|GB|YTD|YTR|VDID} :SMONitor :IAC [:STATe] {ON|OFF|1|0} :IDC [:STATe] {ON|OFF|1|0} :VAC [:STATe] {ON|OFF|1|0} :VDC [:STATe] {ON|OFF|1|0} :HCOPy :SDUMp :DATA < block> [Query only] :INITiate :CONTinuous {ON|OFF|1|0} [:IMMediate] [No Query] Keysight E4980A/AL Precision LCR Meter...
  • Page 393 < array> :MDATa < array> [Query only] :TYPE {FREQuency|VOLTage|CURRent|BVOLtag e|BCURrent|DCSVoltage },{NONE|IRANge|FREQuency|VOLTage|CU RRent|BVOLtage|BCURrent|DCSVoltage} :VOLTage < array> :MEMory :CLEar {DBUF} [No Query] :DIM {DBUF },< numeric> :FILL {DBUF} [No Query] :READ {DBUF },< block> [Query only] :MMEMory Keysight E4980A/AL Precision LCR Meter...
  • Page 394 :VALue < numeric> [:STATe] {ON|OFF|1|0} :HPOWer {ON|OFF|1|0} :SOURce :DCSource :STATe {ON|OFF|1|0} :VOLTage [:LEVel] < numeric> :STATus :OPERation :CONDition < numeric> [Query only] :ENABle < numeric> [:EVENt] < numeric> [Query only] :SYSTem :BEEPer [:IMMediate] [No Query] Keysight E4980A/AL Precision LCR Meter...
  • Page 395 < string> [Query only] :SMASk < string> :DATE < numeric>,< numeric>,< numeric> :ERRor [:NEXT] < numeric>,< string> [Query only] :KLOCk {ON|OFF|1|0} :PERSona :MANufacture [:NAME] < string> :DEFault :PRESet [No Query] :TIME < numeric>,< numeric>,< numeric> :TRIGger Keysight E4980A/AL Precision LCR Meter...
  • Page 396 SCPI Command Reference Command list Command Parameter Note :DELay < numeric> [:IMMediate] [No Query] :SOURce {INTernal|HOLD|EXTernal|BUS} :TDEL < numeric> :VOLTage [:LEVel] < numeric> Keysight E4980A/AL Precision LCR Meter...
  • Page 397: Softkey Functions

    Displays next page PAGE PREV Displays previous page PAGE MODE Selects the sweep mode for the list sweep measurement :LIST:MODE SEQ function STEP Selects the sweep mode for the list sweep measurement :LIST:MODE STEP function Keysight E4980A/AL Precision LCR Meter...
  • Page 398 Cs- ... Cs-D Selects the measurement function :FUNC:IMP:TYPE CSD Cs-Q Selects the measurement function :FUNC:IMP:TYPE CSQ Cs-Rs Selects the measurement function :FUNC:IMP:TYPE CSRS RETUR Returns to the softkey display screen one level higher Lp- ... Keysight E4980A/AL Precision LCR Meter...
  • Page 399 Z- ... Z-d Selects the measurement function :FUNC:IMP:TYPE ZTD Z-r Selects the measurement function :FUNC:IMP:TYPE ZTR RETUR Returns to the softkey display screen one level higher LEVEL Returns the actual applied bias output level :VOLT:LEV :CURR:LEV Keysight E4980A/AL Precision LCR Meter...
  • Page 400 Executes the OPEN correction at the specified :CORR:SPOT[1-201]:OPEN OPEN measurement point :EXEC MEAS Executes the SHORT correction at the specified :CORR:SPOT[1-201]:SHOR SHORT measurement point :EXEC Enables the specified measurement point :CORR:SPOT[1-201]:STAT Enables the specified measurement point :CORR:SPOT[1-201]:STAT FUNC Cp- ... Keysight E4980A/AL Precision LCR Meter...
  • Page 401 Returns to the softkey display screen one level higher Ls- ... Ls-D Selects the measurement function of the reference values :CORR:LOAD:TYPE LSD for the LOAD correction Ls-Q Selects the measurement function of the reference values :CORR:LOAD:TYPE LSQ for the LOAD correction Keysight E4980A/AL Precision LCR Meter...
  • Page 402 Executes the LOAD corection at the specified :CORR:SPOT[1-201]:LOAD: measurement point STAN LOAD B Executes the LOAD corection at the specified :CORR:SPOT[1-201]:LOAD: measurement point STAN MODE MULTI Selects the correction mode :CORR:METH MULT SINGLE Selects the correction mode :CORR:METH SING OPEN Keysight E4980A/AL Precision LCR Meter...
  • Page 403 Turns OFF the auxiliary BIN counting function :COMP:ABIN OFF Turns ON the auxiliary BIN counting function :COMP:ABIN ON BEEP FAIL Selects the condition for generating a beep sound :COMP:BEEP FAIL PASS Selects the condition for generating a beep sound :COMP:BEEP PASS Keysight E4980A/AL Precision LCR Meter...
  • Page 404 Sets the nominal value for the tolerance mode of the :COMP:TOL:NOM comparator function LIST SETUP FREQ[Hz] FREQ Clears the previous list sweep table, sets frequency :LIST:FREQ [Hz] sweep as the list sweep parameter, and sets frequency points for the frequency list sweep Keysight E4980A/AL Precision LCR Meter...
  • Page 405 NEXT Displays the next page PAGE PREV Displays the previous page PAGE HIGH CLEAR Clears the sweep point that has been selected LINE NEXT Displays the next page PAGE PREV Displays the previous page PAGE Keysight E4980A/AL Precision LCR Meter...
  • Page 406 MEAS SETUP Turns OFF the Automatic Level Control (ALC) :AMPL:ALC OFF Turns ON the Automatic Level Control (ALC) :AMPL:ALC ON Sets the integration time and the averaging rate :APER BIAS Sets the DC bias. :BIAS:VOLT:LEV :BIAS:CURR:LEV Keysight E4980A/AL Precision LCR Meter...
  • Page 407 Selects the deviation measurement mode :FUNC:DEV2:MODE OFF FREQ Sets the measurement frequency :FREQ:CW Selects the deviation measurement mode :FUNC:DEV1:MODE PERC Selects the deviation measurement mode :FUNC:DEV1:MODE ABS Selects the deviation measurement mode :FUNC:DEV1:MODE OFF FUNC Keysight E4980A/AL Precision LCR Meter...
  • Page 408 Selects the measurement function :FUNC:IMP:TYPE LSQ Ls-Rdc Selects the measurement function :FUNC:IMP:TYPE LSRD Ls-Rs Selects the measurement function :FUNC:IMP:TYPE LSRS RETUR Returns to the softkey display screen one level higher MORE Selects the measurement function :FUNC:IMP:TYPE GB Keysight E4980A/AL Precision LCR Meter...
  • Page 409 Selects impedance measurement range :FUNC:IMP:RANG:VAL AUTO Enables the auto range function for the impedance :FUNC:IMP:RANG:AUTO measurement HOLD Enables the auto range function for the impedance :FUNC:IMP:RANG:AUTO measurement REF A Sets the reference value for deviation measurement :FUNC:DEV1:REF:VAL Keysight E4980A/AL Precision LCR Meter...
  • Page 410 Enters arbitrary comments into the comment line :DISP:LINE NEXT Displays the next page PREV Displays the previous page VDC MON Disables the DC voltage level monitor function :FUNC:SMON:VDC:STAT Enables the DC voltage level monitor function :FUNC:SMON:VDC:STAT [Preset] CLEAR SET&CORR Keysight E4980A/AL Precision LCR Meter...
  • Page 411 SAVE DATA RETURN Returns to the softkey display screen one level higher SAVE & Stores the state from the memories :MMEM:STOR STOP :MEM:CLE START LOG Enables the data buffer memory to store the :MEM:FILL measurement data Keysight E4980A/AL Precision LCR Meter...
  • Page 412 Disables a beep sound. :SYST:BEEP:STAT OFF Enables a beep sound. :SYST:BEEP:STAT ON BEEPER TONE TONE 1 Selects beep tone 1 :SYST:BEEP:TONE 1 TONE 2 Selects beep tone 2 :SYST:BEEP:TONE 2 TONE 3 Selects beep tone 3 :SYST:BEEP:TONE 3 Keysight E4980A/AL Precision LCR Meter...
  • Page 413 Sets the IP address to MANUAL :SYST:COMM:LAN:SELF:C ONF MAN FACTORY Sets the IP address at the default DEFAULT MANUAL GATEWAY ENTER Sets the static Gateway address :SYST:COMM:LAN:SELF:D FACTORY Sets the IP address at the default DEFAULT Keysight E4980A/AL Precision LCR Meter...
  • Page 414 Disables the scanner I/F :CONT:SCAN:STAT OFF Enables the scanner I/F :CONT:SCAN:STAT ON [Trigger] Triggers the unit :TRIG:IMM 1. The field name changes depending on the list sweep parameter. 2. The field name changes when entering a comment. Keysight E4980A/AL Precision LCR Meter...
  • Page 415: Data Processing

    SCPI Command Reference Data Processing Data Processing The internal data processing flowchart for the E4980A/AL is shown in Figure 10-1 Figure 10-1 Data Processing Flowchart Keysight E4980A/AL Precision LCR Meter...
  • Page 416 SCPI Command Reference Data Processing Keysight E4980A/AL Precision LCR Meter...
  • Page 417: Specifications And Supplemental Information

    Keysight E4980A/AL Precision LCR Meter User’s Guide 11 Specifications and Supplemental Information For the specifications and supplemental information of the E4980A/AL Precision LCR meter, refer to the datasheet at http://literature.cdn.keysight.com/litweb/pdf/5989-4435EN.pdf...
  • Page 418 Specifications and Supplemental Information Keysight E4980A/AL Precision LCR Meter...
  • Page 419: Precautions For Use And Daily Checks

    Keysight E4980A/AL Precision LCR Meter User’s Guide 12 Precautions for Use and Daily Checks This chapter describes precautions to take when using the E4980A/AL and explains how to perform regular maintenance on the device. Precautions for Use This section describes the precautions to take in using the E4980A/AL.
  • Page 420 Precautions for Use Setup procedure Step 1. Press the [Local/Lock] key on the front panel. Step 2. Confirm that LOCK is displayed in the status display area in the lower-right corner of the LCD display. Keysight E4980A/AL Precision LCR Meter...
  • Page 421: Daily Checks (Executing The Self-Test)

    INCR+ and DECR- softkeys. Step 6. Press EXECUTE - TEST START (2 in Figure 12-2) to execute the selected test item. Figure 12-2 Self-test screen and procedure The following test items can be checked by using the self-test. Keysight E4980A/AL Precision LCR Meter...
  • Page 422 6 HANDLER INTERFACE Checks the handler interface. (Visual confirmation only; no pass/fail result shown on screen.) 7 SCANNER INTERFACE Checks the scanner interface. (Visual confirmation only; no pass/fail result shown on screen.) Keysight E4980A/AL Precision LCR Meter...
  • Page 423: Cleaning This Instrument

    To remove stains on parts other than the unknown terminals and DC source ports, wipe them gently with a soft cloth that is dry or one that is wetted with a small amount of water and wrung tightly. Keysight E4980A/AL Precision LCR Meter...
  • Page 424: Cautions Applicable To Requesting Repair, Replacement, Regular Calibration, Etc

    Cautions Applicable to Requesting Repair, Replacement, Regular Calibration, etc. Caution when Sending the Unit If it is necessary to send the unit to a Service Center of Keysight Technologies, please follow the instructions below. Equipment to be Sent When requesting repair or regular calibration of the unit at our Service Center, send only the E4980A/AL main unit without any installed option.
  • Page 425: Troubleshooting

    Office or the company from which you bought the device. Either or them may solve your trouble. When you contact to Keysight Technologies’ Sales and Services Office or the company from which you bought the device, provide the screen shot which you...
  • Page 426 Error messages or Follow the suggestions on Error Warning messages Messages / Warning Messages solves the trouble? Contact Keysight Technology’ s Sales Restart from Factory and Service Office or the company from Reset solves the which you bought the device.
  • Page 427: Check Items When Trouble Occurs

    If normal operation does not resume after taking the above measures, there is the possibility of a failure. Unplug the power cable immediately and contact Keysight Technologies’s Customer Contact, listed at the end of this guide, or the company from which you bought the device.
  • Page 428: Beeping Persists When Turning On The Comparator Function

    For those cases, connect OPEN/SHORT/LOAD properly or fixture/cable properly and then perform the correction with new conditions. If the measured value is still abnormal even after taking the above measure, turn off the correction function. Keysight E4980A/AL Precision LCR Meter...
  • Page 429: Saving To Usb Memory Fails

    Some USB memory does not work on the E4980A/AL with firmware A.xx.xx. See the section on “USB interface port” in data sheet. An error message or warning message is displayed For information on error messages and warning messages, refer to “Error Messages” on page 435. Keysight E4980A/AL Precision LCR Meter...
  • Page 430: Check Items When Trouble Occurs During Remote Control

    You cannot read out the measured value. Confirm that the data transfer format is set up correctly. An error message is displayed Check that the program is correct. “Error Messages” on page 435. For details on error messages, refer to Keysight E4980A/AL Precision LCR Meter...
  • Page 431: A.manual Changes

    Manual Changes by Serial Number Serial Number Make Manual Changes MY464/SG464 and below for E4980A Change 5 MY542/SG542 and below for E4980AL The ten-character serial number is stamped on the serial number plate (Figure A-1) on the rear panel. Figure A-1...
  • Page 432: Change 5

    Compliant with LXI (Lan eXtensions for Instrumentation) standard Class C DHCP ON/OFF function and AUTO IP address ON/OFF function integrated (IP CONFIG field). “Status” which displays the LAN connection status changed (CURRENT LAN STATUS monitor). -432 Keysight E4980A/AL User’s Guide...
  • Page 433 Please delete the descriptions about this command from this manual. — “:SYSTem:COMMunicate:LAN[:SELF]:CONFigure” on page 372 The firmware revision A.02.00 or later does not support the following SCPI commands. “:SYSTem:COMMunicate:LAN[:SELF]:AIP[:STATe]” on page 371 — “:SYSTem:COMMunicate:LAN[:SELF]:DHCP[:STATe]” on page 373 — Keysight E4980A/AL User’s Guide -433...
  • Page 434 Manual Changes Manual Changes -434 Keysight E4980A/AL User’s Guide...
  • Page 435: B.error Messages

    Keysight E4980A/AL Precision LCR Meter User’s Guide Error Messages The Keysight E4980A/AL provides error messages to indicate its operating status. This appendix describes the error messages of the E4980A/AL in numeric order. Error Messages An error message is displayed in the instrument status display area in the lower-left part of the screen.
  • Page 436: Gpib Devices In Ieee488.2

    In this instrument, this error occurs when the number of characters exceeds 254 in a character-string parameter. -436 Keysight E4980A/AL User’s Guide...
  • Page 437 Expression is improper. -168 Block data not allowed Block data is not allowed. -161 Invalid block data Invalid block data was received (for example, END received before length satisfied). Block data error -160 Block data is improper. Keysight E4980A/AL User’s Guide -437...
  • Page 438 The header suffix is out of range. -113 Undefined header A command not defined in this instrument, though not illegal in the syntactic structure, has been received. For example, when the message “:OUTP:DC:ISOL:FREQ:AUTO ON” is sent instead of the correct program -438 Keysight E4980A/AL User’s Guide...
  • Page 439 Invalid separator -103 The message unit separator (for example, “;”, “,”) is improper. Syntax error -102 Unrecognized command or data type was received. Invalid character -101 Invalid character was received. Command Error -100 Improper command. Keysight E4980A/AL User’s Guide -439...
  • Page 440: Setup

    A fault has occurred in a circuit to absorb back-emf of the 42842A/B. Suggestion: Confirm the connection to the 42842A/B. Fixture over temperature The temperature inside the 42842A/B bias test fixture exceeds the limit. Suggestion: Turn off the DC bias and cool down the equipment. -440 Keysight E4980A/AL User’s Guide...
  • Page 441 The opening and closing detection sensor of the 42842A/B protective cover detects an error. Suggestion: When this error occurs, contact Keysight Technology’s Sales and Service Office or the company from which you bought the device. Fixture cover open The protective cover of the 42842A/B is open.
  • Page 442 Select the sequential mode using COMP:MODE. No values in sweep list List sweep measurement is performed when the measurement point is not set on the LIST SWEEP SETUP screen. Suggestion: Set the measurement point on the LIST SWEEP SETUP screen. -442 Keysight E4980A/AL User’s Guide...
  • Page 443 USB memory into the E4980A/AL. You need to wait for a few seconds after inserting the USB. Suggestion: Prepare the USB memory device or save data into number 0 to 9. Keysight E4980A/AL User’s Guide -443...
  • Page 444: Hardware

    Suggestion: Check the connection with a cable when you change the setting of the interface board. When the interface board connected properly, contact Keysight Technology’s Sales and Service Office or the company from which you bought the device. 1070 Fan failed Cooling fan hardware failure is detected.
  • Page 445 1080 Power failed Power unit hardware failure is detected. Suggestion: When this error occurs, contact Keysight Technology’s Sales and Service Office or the company from which you bought the device. 1103 A1 EEPROM write error An error is generated while writing data to A1 EEPROM.
  • Page 446: Warning Messages

    This message appears when attempting to change or delete the list sweep parameter from the E4980A/AL’s screen after setting the sweep point for the second parameter with the GPIB command. Suggestions: Use the GPIB command to change the list sweep parameter. -446 Keysight E4980A/AL User’s Guide...
  • Page 447 The measurement result data status of the measurement result output is set to 3. Suggestions: Reduce the test signal level so that the test signal current doesn’t exceed a value shown in Table 11-6. Keysight E4980A/AL User’s Guide -447...
  • Page 448 Error Messages Warning Messages Signal source overvoltage The signal source voltage exceeds its limit. Suggestions: Lower the voltage setting of the signal source. -448 Keysight E4980A/AL User’s Guide...
  • Page 449: C.list Of Default Values

    Keysight E4980A/AL Precision LCR Meter User’s Guide List of Default Values This appendix gives the default values, settings for Save/Recall of an object, and settings for backing up an object when using the Keysight E4980A/AL. -449 Keysight E4980A/AL User’s Guide...
  • Page 450: List Of Default Values, Save/Recall Settings, And Backup Settings

    List of Default Values, Save/Recall Settings, and Backup Settings List of Default Values, Save/Recall Settings, and Backup Settings The table below shows the following settings for the Keysight E4980A/AL. For “List of LAN Factory Default the factory default settings of the LAN, refer to Settings”...
  • Page 451: List Of Default Values

    FUNC Cp-D << << LEVEL << << MEAS TIME << << RANGE AUTO(ON) << << [Local/Lock] << << [Meas Setup] CORRECTION CABLE << << << FREQ << FUNC Cp-D << LOAD << LOAD A << Keysight E4980A/AL User’s Guide -451...
  • Page 452 << << LIST SETUP MODE << << MEAS SETUP << << << << BIAS << << BIAS POL FIX(OFF) << << DC SRC << << DCI ISO << << DCI RNG 20 mA << << -452 Keysight E4980A/AL User’s Guide...
  • Page 453 STEP DLY << << TRIG << << TRIG DLY << << USER ““ << << COMMENT VDC MON << << [Recall A] [Recall B] [Save/Recall] MEDIA << << [System] SYSTEM CONFIG BEEPER ENABLE BEEPER TONE Keysight E4980A/AL User’s Guide -453...
  • Page 454 Factory-shipped Setting Default Value Save/ Backup Available Recall Means of Defining a Preset *RST Setting DATE/TIME GPIB ADDR IP CONFIG AUTO MANUAL “0.0.0.0” GATEWAY MANUAL IP “192.168.1.101” ADDR MANUAL “255.255.255.0” SUBNET MASK SYSTEM INFO HANDLER SCANNER -454 Keysight E4980A/AL User’s Guide...
  • Page 455: List Of Lan Factory Default Settings

    1. If you change the item which requires reconnection of the network, a message “Click ‘Renew LAN Settings’ to make saved chantes effective” will appear. 2. xxxxx indicates the serial number of the E4980A/AL. Keysight E4980A/AL User’s Guide -455...
  • Page 456 List of Default Values List of LAN Factory Default Settings -456 Keysight E4980A/AL User’s Guide...
  • Page 457: D.bias Current Interface

    Bias Current Interface This chapter explains how to use Option 002 Bias Current Interface, which adds a digital interface to allow the E4980A to control the Keysight E4980A Precision LCR Meter’s bias current source. This information is only for serial prefix number below:...
  • Page 458: Product Overview

    Cable length setting applies only to the 1 m setting. Turn off all the power switches of the 42841A and E4980A when connecting or disconnecting any connection involving 42841A and E4980A. The 42842C bias current test fixture cannot be used with the E4980A/AL. -458 Keysight E4980A/AL User’s Guide...
  • Page 459: Specifications When Measuring With The 42841A

    10 mA (0 A to 1 A) 42841A 100 mA (1 A to 20 A) (connect 42841A x 1 unit) 20 mA (0 A to 2 A) 200 mA (2 A to 40 A) (connect 42841A x 2 units) Keysight E4980A/AL User’s Guide -459...
  • Page 460: Preparation

    Current Control Input A connector of the 42841A using the cable and the adapter. The cable is furnished with the 42841A and the adapter is furnished with the E4980A. 3. Connect the 42842A test fixture directly to the front panel of the 42841A. -460 Keysight E4980A/AL User’s Guide...
  • Page 461 Bias Current Interface Preparation Figure D-1 Cable connection for an E4980A (20 A) system Keysight E4980A/AL User’s Guide -461...
  • Page 462 42841A and the adapter is furnished with the E4980A. 5. Connect the 42841A Bias Current Control Input B connector connected to the 42842B to the Bias Current Control Input A connector of the other 42841A using the furnished Bias Current Interface Cable (2). -462 Keysight E4980A/AL User’s Guide...
  • Page 463 Bias Current Interface Preparation Figure D-2 Cable connection for an E4980A(40 A) system Keysight E4980A/AL User’s Guide -463...
  • Page 464: Measurement Procedures

    For measurement procedures when using the 42841A Bias Current Source, see “Basic Measurement Procedure” of the Chapter, “Operation Procedure and Examples” in the 42841A manual. The measurement procedure is the same as that of the 4284A. -464 Keysight E4980A/AL User’s Guide...
  • Page 465: E.handler Interface

    Interface for the Keysight E4980A/AL Precision LCR Meter. Overview The Keysight E4980A/AL Option 201 Handler Interface outputs signals that indicate the end of a measurement cycle, the result of bin sorting by the comparator, and the result of Go/No-Go test by the list sweep comparator. In addition, the instrument accepts input of external trigger and key-lock signals.
  • Page 466: Specifications

    /READY_FOR_TRIGGER “Ready for trigger” signal Input Signals Photo isolation Table E-2 Input Signals Signal Overview /KEY_LOCK Key lock (locks out the front panel keyboard) External trigger signal, pulse width ≥ 1 s /EXT_TRIG -466 Keysight E4980A/AL User’s Guide...
  • Page 467: Signal Line Definitions

    These signals indicate the result of sorting. They show into which bin the DUT has /BIN2 been sorted. /BIN3 All of these output signals are open-collector based. /BIN4 /BIN5 /BIN6 /BIN7 /BIN8 /BIN9 /OUT_OF_BINS /AUX_BIN Keysight E4980A/AL User’s Guide -467...
  • Page 468 DUT can be connected to the UNKNOWN terminal. However, measurement data are not available until the /EOM signal is output. /EOM Output “End of measurement cycle” signal. When this signal is output, the measurement data and sorting results are available. -468 Keysight E4980A/AL User’s Guide...
  • Page 469 Signal Line Definitions Table E-3 Pin layouts for the comparator Signal name Signal Description number direction COM2 Common signals for external DC current EXT.DCV2. COM2 COM1 Common signals for external DC current EXT.DCV1. COM1 COM1 Keysight E4980A/AL User’s Guide -469...
  • Page 470 Handler Interface Signal Line Definitions Figure E-1 Example of /PHI, /PLO, and /SREJ signal areas (for the comparator) -470 Keysight E4980A/AL User’s Guide...
  • Page 471 Handler Interface Signal Line Definitions Figure E-2 Handler interface connector pin layout Keysight E4980A/AL User’s Guide -471...
  • Page 472 Analog measurement time Depends on settings Digital math time 0 Sec Trigger wait time after the output of /EOM, 0 Sec /READY_FOR_TRIGGER The measurement time is affected by the GPIB/USB/Network and Front panel processes. -472 Keysight E4980A/AL User’s Guide...
  • Page 473: Signal Line Definitions For The List Sweep Comparator

    Figure E-5 comparator). Also, shows a signal timing chart. A signal prefixed with a slash (/) indicates that the signal is negative logic (active low). Keysight E4980A/AL User’s Guide -473...
  • Page 474 (including the comparator test time) and the corresponding measurement data are available. The comparator test result is available after the /EOM signal for the last sweep point is output. 12-18 Same assignments as for the comparator. 22-29 32-36 -474 Keysight E4980A/AL User’s Guide...
  • Page 475 Handler Interface Signal Line Definitions Figure E-4 Example of signal areas (for the list sweep comparator) Keysight E4980A/AL User’s Guide -475...
  • Page 476 Timing chart (for the list sweep comparator) Table E-6 Time segments in the time chart Time segment Minimum value Typical 1 sec Trigger pulse width /INDEX and /EOM response time of measurement time 0 Sec 350 msec -476 Keysight E4980A/AL User’s Guide...
  • Page 477 Trigger wait time after the output of /EOM 0 Sec 1. The measurement time is affected by the GPIB/USB/Network and Front panel processes. For settling and display times, refer to Chapter 11, “Specifications and Supplemental Information”, on page 417. Keysight E4980A/AL User’s Guide -477...
  • Page 478: Electrical Characteristics

    Circuit common of /EOM E4980A/AL /ALARM /READY_FOR_TRIGGER When external voltage (EXT.DCV2) is used: COM1 +5 V or +12 V when internal power supply is used. +5 V to +24 V (EXT.DCV1) when external power supply is used. -478 Keysight E4980A/AL User’s Guide...
  • Page 479 +5 V to +24 V (EXT.DCV2) when external power supply is used. Figure E-6 shows the output circuit configuration for the test result output signals, and Figure E-7 shows that for the control output signals. Figure E-6 Output circuit configuration for the test result output signals Keysight E4980A/AL User’s Guide -479...
  • Page 480: Dc Isolation Input Signals (Photo-Coupled)

    (+5 V,+12 V) or external voltage (EXT.DCV2). The trigger input limiting resistor should be set via jumper J6 depending on the magnitude of the photocoupler anode voltage. For more information, see “Setting Up the Handler Interface Board” on page 483. -480 Keysight E4980A/AL User’s Guide...
  • Page 481 5 V to 24 V 2.4 mA 6.6 mA 9.1 mA 14.6 mA When internal pull-up voltage is used: Circuit common of E4980A/AL When external voltage is used: COM2 Figure E-8 shows the circuit configuration for the input signals. Keysight E4980A/AL User’s Guide -481...
  • Page 482 Handler Interface Electrical Characteristics Figure E-8 Circuit configuration for the input signals -482 Keysight E4980A/AL User’s Guide...
  • Page 483: Setting Up The Handler Interface Board

    1. The factory default setting is position (N) shown in the table above. The handler interface board has labels “JP1” and “JP2” that indicate where jumpers JP1 and JP2 are located. Both jumpers JP1 and JP2 have a circle mark indicating their #1 side. Keysight E4980A/AL User’s Guide -483...
  • Page 484 When this switch is OFF, the DC isolation output for COM2 is /READY_FOR_TRIGGER isolated. Not used. Not used. 1. The bit numbers referenced in this table are the numbers printed on the bit switch (S1) of the handler interface board. -484 Keysight E4980A/AL User’s Guide...
  • Page 485 Step 3. Draw out the handler interface board and unplug the flat cable that is connected to it. Step 4. Remove the handler interface board. Step 5. Go through the following flow chart to determine the jumper and bit switch settings for the test result output signals: Keysight E4980A/AL User’s Guide -485...
  • Page 486 Setting 1 Instrument’s circuit Required +5 V, internal common Setting 2 Instrument’s circuit Required +12 V, internal common Setting 3 COM1 Required +5 V through +24 V, external (EXT.DCV1) Setting 4 COM1 Not required -486 Keysight E4980A/AL User’s Guide...
  • Page 487 The pull-up resistor for DCV2 uses J3, whose location is indicated as J3 on the handler interface board. The pull-up resistor must be a resistor array of which the characteristics satisfy the following formula:  Equation E-2 ------ - Keysight E4980A/AL User’s Guide -487...
  • Page 488 (0.98 N-Em, 0.1 kgf-m Max). The typical pull-up resistor values are: Table E-14 The typical pull-up resistor vaules Pull-Up Voltage Pull-Up Resistor Keysight Part Number 1.7 k ( = 5 V / 3 mA) 1810-0276 (1.5 k) -488 Keysight E4980A/AL User’s Guide...
  • Page 489 4.0 k ( = 12 V / 3 mA) 1810-0279 (4.7 k) 15 V 5.0 k ( = 15 V / 3 mA) 1810-0279 (4.7 k) 24 V 8.0 k ( = 24 V / 3 mA) 1810-0280 (10.0 k) Keysight E4980A/AL User’s Guide -489...
  • Page 490: Using The Handler Interface

    Step 8. Press the [Display Format] key to open the MEAS DISPLAY page. Prepare for measurement by pressing one of the following softkeys: MEAS DISPLAY, BIN No., and BIN COUNT. You can also use the BIN No. DISPLAY page to enable or disable the comparator. -490 Keysight E4980A/AL User’s Guide...
  • Page 491: Setting Up The Handler Interface Board For Use With The List Sweep Comparator

    Step 5. Press the ON softkey. The handler interface is enabled for signal input/output. Step 6. Press the [Display Format] key to open the MEAS DISPLAY page. Press the LIST SWEEP softkey so that you can begin list sweep measurement. Keysight E4980A/AL User’s Guide -491...
  • Page 492: Signals Activated When Overload Occurs

    Signals are not output when the limit parameter (LMT field) is OFF during the list sweep measurement. Signals are not output when the comparator function (COMP field) is OFF during the measurement other than list sweep. -492 Keysight E4980A/AL User’s Guide...
  • Page 493: F.scanner Interface

    Keysight E4980A/AL Precision LCR Meter User’s Guide Scanner Interface This chapter provides electrical characteristics of the interface signal line necessary to use the Keysight E4980A/AL Precision LCR meter option 301. -493 Keysight E4980A/AL User’s Guide...
  • Page 494: Overview

    Overview Overview When the Keysight E4980A/AL is used with the option 301 Scanner Interface, it is possible to store and use up to 128 sets of correction data (OPEN, SHORT, and LOAD) for up to 201 user-defined test frequencies. These 128 sets of correction data can be used for each measurement using the multi-channel correction function.
  • Page 495: Specifications

     1 V EXT_TRIG Pull-up 5 to 15 V 6.3 to15 mA Voltage Table F-3 Output Signal Characteristics: Opto-isolated, negative true Signal Name External Pull-Up Voltage Maximum Current /INDEX Max. 15 V 6 mA /EOM Keysight E4980A/AL User’s Guide -495...
  • Page 496: Signal Input/Output Connector

    End of measurement. Since this signal is asserted when a series of measurement is completed, the E4980A/AL moves to the next measurement. When this signal is asserted, the measurement data and the comparison results are valid. -496 Keysight E4980A/AL User’s Guide...
  • Page 497 External trigger. This signal triggers the E4980A/AL on the rising edge of a pulse when the trigger mode is set to EXT_TRIG. The / (slash) in the signal name means that the signal is asserted when low. Keysight E4980A/AL User’s Guide -497...
  • Page 498: Scanner Interface I/O Connector Pin Assignments

    /CH4 /CH6 /CH_VALID Input Channel Valid Signal /INDEX Output Analog Measurement Complete Signal. EXT.DCV Input External DC Voltage. Used as pull-up voltage of input/output signals. The setting range is from 5 V to 15 V. -498 Keysight E4980A/AL User’s Guide...
  • Page 499 Step 5. Set the scanner channel, channel selection signals, and the channel valid signal to the next channel to be measured when /INDEX is asserted. Step 6. After /EOM is asserted, the E4980A/AL is ready to make the next measurement. Step 7. Repeat Step 4 through Step Keysight E4980A/AL User’s Guide -499...
  • Page 500: Timing Chart

    Each time of the timing chart Time Minimum Typical 1 sec Trigger Pulse Width 350 sec Measurement Start Delay Time /CH_VALID Wait Time after Measurement 0 nsec Start 0 sec Wait Time after /EOM Output -500 Keysight E4980A/AL User’s Guide...
  • Page 501 Scanner Interface Signal Input/Output Connector Table F-8 Each time of the timing chart Time Minimum Typical 0 sec /CH_VALID Setup Time The measurement time is affected by the GPIB/USB/Network and Front panel processes. Keysight E4980A/AL User’s Guide -501...
  • Page 502 Settling time includes switching time for correction data. Comparison and display time is approx. 4.5 ms. Refer to Table F-8 for information on T1, T2, T3, T4, and T5. Figure F-5 shows a sample of the system configuration using the input/output signals described above. -502 Keysight E4980A/AL User’s Guide...
  • Page 503 Scanner Interface Signal Input/Output Connector Figure F-5 Sample of system configuration of scanner interface Keysight E4980A/AL User’s Guide -503...
  • Page 504: Electrical Characteristics Of The Input/Output Signals

    16.2 mA /CH_VALID EXT_TRIG 0 to 1 V EXT.DCV 11.1 mA 10.5 mA 13.5 mA Figure F-6 shows a diagram of channel selection input signals, and Figure F-7 shows a diagram of external trigger input signals. -504 Keysight E4980A/AL User’s Guide...
  • Page 505 Scanner Interface Electrical Characteristics of the Input/Output Signals Figure F-6 Diagram of channel selection input signal circuits Keysight E4980A/AL User’s Guide -505...
  • Page 506 Scanner Interface Electrical Characteristics of the Input/Output Signals Figure F-7 Diagram of external trigger input signal circuits -506 Keysight E4980A/AL User’s Guide...
  • Page 507: Control Output Signals

    /INDEX and /EOM output circuits. Table F-10 Output signal electrical characteristics Signal Name External Pull-up Voltage Maximum Current /INDEX 15 V max. 6 mA /EOM Figure F-8 Simplified diagram of the output circuits Keysight E4980A/AL User’s Guide -507...
  • Page 508: Setting Up The Scanner Interface Board

    (when on) input resistance 820  8 V < EXT.DCV  15 V (when off) 1. For bits 1 to 3 of bit switch S1, only one bit can be turned on at one time. -508 Keysight E4980A/AL User’s Guide...
  • Page 509 Step 6. Connect the flat cable to the scanner interface board, replace the scanner interface board, and fix it onto place with the two screws. Applying excessive force to the screws may break the tapped hole (less than 0.98 N-m, 0.1 kgf-m). Keysight E4980A/AL User’s Guide -509...
  • Page 510: How To Use The Scanner Interface Board

    This section provides information necessary for the E4980A/AL to operate using the scanner interface. Basic Procedure Figure F-9 shows a sample procedure for using the scanner interface. Follow this flow chart while referring to the following detailed descriptions. -510 Keysight E4980A/AL User’s Guide...
  • Page 511 Scanner Interface How to Use the Scanner Interface Board Figure F-9 Basic Procedure Keysight E4980A/AL User’s Guide -511...
  • Page 512: Activating Scanner Interface

    On this page, the reference values (REF A and REF B) of the standards used for LOAD correction are defined at 201 test frequencies. The reference values can be stored with an appropriate function (FUNC). For example, to use the -512 Keysight E4980A/AL User’s Guide...
  • Page 513 Step 11. Use the cursor arrow keys on the front panel to move the cursor to the REF B field. Step 12. Use the numeric entry keys and softkeys on the front panel to enter the reference value of the secondary parameter at the specified measurement point. Keysight E4980A/AL User’s Guide -513...
  • Page 514: Correction Data Selection

    8-bit binary signals. /CH0 is the LSB (least significant bit) and /CH7 is the MSB Table F-13 (most significant bit). An example is given below in Table F-13 Example of Channel Selection /CH0 /CH1 /CH2 /CH3 /CH4 /CH5 /CH6 /CH7 Channel Number -514 Keysight E4980A/AL User’s Guide...
  • Page 515 Monitoring Current Channel You can monitor the current channel number selected for correction on the CORRECTION page and the MEAS DISPLAY page. The “CH” field on these pages indicates the current channel selected for correction. Keysight E4980A/AL User’s Guide -515...
  • Page 516: Correction Data Measurements

    Step 15. If necessary, repeat steps 11 to 14 for other measurement points (other frequencies). Step 16. [LOAD correction] Use the cursor arrow keys on the front panel to move the cursor to the SPOT No. field. -516 Keysight E4980A/AL User’s Guide...
  • Page 517: Activating The Correction Function

    Step 9. Press the “ON” softkey to set the LOAD correction function to ON. Correction Data Confirmation There are two ways to confirm the measurement data for each channel correction: Use the front panel, and read the data using a query via GPIB. Keysight E4980A/AL User’s Guide -517...
  • Page 518 <open1 A>,<open1 B>,<short1 A>,<short1 B>,<load1 A>,<load1 B>, <open2 A>,<open2 B>,<short2 A>,<short2 B>,<load2 A>,<load2 B>, <open3 A>,<open3 B>,<short3 A>,<short3 B>,<load3 A>,<load3 B>, <open200 A>,<open200 B>,<short200 A>,<short200 B>,<load200 A>,<load200 B>, <open201 A>,<open201 B>,<short201 A>,<short201 B>,<load201 A>,<load201 B>, <NL^END> -518 Keysight E4980A/AL User’s Guide...
  • Page 519 <NL^END> Terminators (the EOI line is asserted while the New Line or Line Feed character is being sent on the bus). XXX in the above table indicates the measurement point number entered in the field. Keysight E4980A/AL User’s Guide -519...
  • Page 520 Scanner Interface How to Use the Scanner Interface Board -520 Keysight E4980A/AL User’s Guide...
  • Page 521: G.functional Comparison Between 4284A, 4279A And E4980A/Al

    This appendix provides a functional comparison between 4284A, 4279A, and E4980A/AL. Functional Comparison between 4284A, 4279A, and E4980A/AL The Migration Guide has been provided for a functional comparison between 4284A, 4279A, and E4980A/AL. Refer to the following website. http://www.keysight.com/find/E4980A/AL/ -521 Keysight E4980A/AL User’s Guide...
  • Page 522 Functional Comparison between 4284A, 4279A and E4980A/AL Functional Comparison between 4284A, 4279A, and E4980A/AL -522 Keysight E4980A/AL User’s Guide...
  • Page 523 Circuit Mode, 198 Four Terminal Pair, 203 Four Terminal Pair, Cleaning, Cleaning, 455 Front Panel, 39 Front Panel, Command Reference, Command Reference, 316 Fuse, 26 Fuse, Comment, Comment, 93 Comment Line, Comment Line, 46 Comparator, Comparator, 136 Keysight E4980A/AL User’s Guide...
  • Page 524 LIMIT TABLE SETUP, with two parameters, LIST SWEEP SETUP, Local Lockout, MEAS SETUP, SELF TEST, SERVICE, STSTEM CONFIG, SYSTEM INFO, Parallel Circuit Mode, MAC Address, Parameters, MAN, Parasitics Impedance, Manual Change, Parasitics Incident, Manual Printing History, Keysight E4980A/AL User’s Guide...
  • Page 525 Measurement Result, EXECute], Screenshot, :CORRection:SPOT[1-201]:SHORt:MULTiple[ System Information, :EXECute], Save/Recall, :CORRection:SPOT[1-201]:STATe, Instrument Configuration State, :CORRection:USE:DATA:SINGle, Scanner Interface, 159, 522 :CORRection:USE:DATA[:MULTi], SCPI Command, :CORRection:USE[:CHANnel], :ABORt, :CURRent[:LEVel], :AMPLitude:ALC, :DISPlay:CCLear, :APERture, :DISPlay:ENABle, :BIAS:CURRent[:LEVel], :DISPlay:LINE, :BIAS:POLarity:AUTO, :DISPlay:PAGE, :BIAS:POLarity:CURRent[:LEVel], :DISPlay[:WINDow]:TEXT[1-2][:DATA]:FMSD: :BIAS:POLarity:VOLTage[:LEVel], DATA, Keysight E4980A/AL User’s Guide...
  • Page 526 :SYSTem:COMMunicate:LAN[:SELF]:CURRe :INITiate:CONTinuous, nt:DGATeway, :INITiate[:IMMediate], :SYSTem:COMMunicate:LAN[:SELF]:CURRe :LIST:BAND[1-201], nt:SMASk, :LIST:BIAS:CURRent, :SYSTem:COMMunicate:LAN[:SELF]:DGATe :LIST:BIAS:VOLTage, way, :LIST:CLEar:ALL, :SYSTem:COMMunicate:LAN[:SELF]:DHCP[: :LIST:CURRent, STATe], :LIST:DCSource:VOLTage, :SYSTem:COMMunicate:LAN[:SELF]:MAC, :LIST:FREQuency, :LIST:MODE, :SYSTem:COMMunicate:LAN[:SELF]:PRESet, :LIST:SEQuence:TSTamp:CLEar, :LIST:SEQuence:TSTamp:DATA, :SYSTem:COMMunicate:LAN[:SELF]:RESTart :LIST:STIMulus:DATA, :LIST:STIMulus:MDATa, :SYSTem:COMMunicate:LAN[:SELF]:SMASk, :LIST:STIMulus:TYPE, :LIST:VOLTage, :SYSTem:DATE, :MEMory:CLEar, :SYSTem:ERRor[:NEXT], :MEMory:DIM, :SYSTem:KLOCk, :MEMory:FILL, :SYSTem:PERSona:MANufacturer[:NAME], :MEMory:READ, Keysight E4980A/AL User’s Guide...
  • Page 527 Users Guide, Shield, SHORT Correction, 121, 212 SICL-LAN, Single Correction Mode, Varactor Diode, Skip Key, Specification, Bias Current Interface, Handler Interface, Warning Message, Scanner Interface, Warranty Period, Specified Frequency Points, Web Server, SRQ, Status Byte, Status Register, Keysight E4980A/AL User’s Guide...
  • Page 528 Index Keysight E4980A/AL User’s Guide...
  • Page 529 This information is subject to change without notice. © Keysight Technologies 2006 - 2022 Edition 17, April 2022 *E4980-90230* E4980-90230 www.keysight.com...

This manual is also suitable for:

E4980aE4980al/052E4980al/032E4980al/102

Table of Contents