DTF (Distance to Fault) Measurements
DTF Measurement Settings
1.8 GHz = 0.18 dB/m
2 GHz = 0.2 dB/m
The correction for loss at 5 meters in one direction: 0.18 dB/m * 5m = 0.9 dB.
All DTF measurements correct for loss for travel down the DUT and back, so
double the correction: 0.9 dB * 2 = 1.8 dB.

Window Settings

Window settings provide the ability to choose between optimizing DTF
measurements for resolving closely-spaced faults or for the ability to measure
low-level faults.
How to select Window settings
— Press Meas Setup 4.
— Then Settings.
— Then press
— Then press
— Then press
— Then press
— Again press Done.

DTF Units

The DTF Units setting is available ONLY on the DTF Settings table.
By default, X-axis units for DTF measurement settings are displayed in Meters.
— Select DTF Units.
— Then Choose from:
— M (meters)
— Feet
Keysight N9912-90001 User's Guide
arrows to move to the Window row.
Ed
it.
Window
and choose from the following:
Maximum
(Optimized for dynamic range) the noise floor is lowered
to provide the ability to measure low-level responses. (Default
setting)
Med ium
– Compromise between Min and Max window settings.
Minimum
– Best Response Resolution, providing the ability to resolve
between two closely-spaced responses.
Done Ed
it.
4- 25