Improving Frequency Resolution and Accuracy
Using a Frequency Counter to Improve Frequency Resolution and Accuracy
Using a Frequency Counter to Improve Frequency Resolution and
Accuracy
This procedure uses the signal analyzer internal frequency counter to increase the
resolution and accuracy of the frequency readout.
Step
1 Select the mode.
2 Preset the mode.
3 Enable the internal
reference signal.
4 Set the center frequency
and span.
5 Turn the frequency
counter on.
Figure 4-1
Using Marker Counter
6 Turn off the marker
counter.
Action
•
Press Mode, Spectrum Analyzer.
•
Press Mode Preset.
•
Press Input/Output, RF
Calibrator, 50 MHz.
a. Press FREQ Channel, Center
Freq, 50, MHz
b. Press SPAN X Scale, Span, 80,
MHz.
•
Press Marker, More, Marker
Count, Counter (On).
•
Press Marker, More, Marker
Count, Count (Off).
Or
Press Marker, Off.
56
Notes
The marker counter remains on